A Study to Evaluate the Safety and Performance of the Dynalink®-E, Everolimus Eluting Peripheral Stent System for Treating Atherosclerotic de Novo or Restenotic Native Superficial Femoral and Proximal Popliteal Artery Lesions
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 104
- 试验地点
- 19
- 主要终点
- In-stent binary restenosis (>50% stenosis), as measured by duplex ultrasound
研究概览
简要总结
The purpose of this first-in-man study is to evaluate the safety and performance of the Dynalink®-E everolimus eluting peripheral stent system for the treatment of patients with atherosclerotic de novo or restenotic native superficial femoral and proximal popliteal lesions.
Abbott Vascular is ceasing data analysis of the STRIDES Clinical Trial after 2 years. The decision to discontinue the study is not related to any safety concern. The rationale for this proposal is based on the following considerations:
The performance of DYNALINK-E from STRIDES shows no device- or procedure-related deaths and no stent fractures, and the rate of additional revascularizations has been stable since approximately 14 months after the procedure.
Evaluations of the bare metal nitinol DYNALINK and ABSOLUTE stents in the clinical literature show low rates of death, reintervention and stent fracture, which are consistent with STRIDES and demonstrate the safety of the nitinol stent platform of the DYNALINK-E.
Long-term animal studies show no concerns with the drug or polymer coating of DYNALINK-E - everolimus tissue concentration drops below the quantifiable limit by approximately 17 months after implant, and vascular response to the coating is normal with widely patent lumens and struts incorporated into vessel tissue.
The safety and performance of the DYNALINK-E has been substantiated by its clinical and pre-clinical data, and by the clinical data of similar products. Given the demonstrated mechanical integrity of the stent along with the evidence of a healthy long-term vascular response, there is a reasonable expectation of continued low event rates.
详细描述
A prospective, Non-Randomized, Single-arm, Multi-center, Clinical Study to Evaluate the Safety and Performance of the Dynalink®-E, Everolimus Eluting Peripheral Stent System for the Treatment of Atherosclerotic de Novo or Restenotic Native Superficial Femoral and Proximal Popliteal Artery Lesions.
研究设计
- 研究类型
- Interventional
- 分配方式
- Non Randomized
- 干预模型
- Single Group
- 主要目的
- Treatment
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Rutherford Becker Category 2-5
- •Single de novo or restenotic lesion in the superficial femoral or proximal popliteal artery
- •Disease segment length 30-170 mm
- •>50% diameter stenosis or total occlusion
- •Target reference vessel diameter 4.3-7.3 mm
排除标准
- •Target lesion previously treated with stent or surgery
- •Rutherford Becker Category 0, 1, or 6
- •Immunosuppressive disorder or currently receiving immunosuppressive agents
- •Serum creatinine >2.5 mg/dl
结局指标
主要结局
In-stent binary restenosis (>50% stenosis), as measured by duplex ultrasound
时间窗: 6 months
次要结局
- Angiographic in-stent binary restenosis rate (>50%)(12 months)
- Angiographic mean in-stent late loss, in-stent mean minimum lumen diameter (MLD), percent diameter stenosis(12 months)
- Primary, and secondary patency(1, 6, 12, 18 months, 2, 3, 4, 5 years)