Safety and Efficacy of a Novel Hydrophobic Acrylic IOL - iPure: a Randomised Study

Vienna Institute for Research in Ocular Surgery2 个研究点 分布在 1 个国家目标入组 55 人开始时间: 2012年4月1日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
55
试验地点
2
主要终点
Rotational Stability

研究概览

简要总结

In this study the rotational stability and performance - with respect to glistening and PCO formation - of a new hydrophobic acrylic IOL (IPure, PhysIOL, Belgium) will be compared to a gold standard IOL (AMO ZCB00) in a randomized controlled fashion. Fifty eyes of 50 patients will be included. 25 patients will receive the iPure and 25 patients will receive the standard IOL. Study hypothesis: The iPure IOL shows better rotational stability compared to the standard IOL. A clinically relevant difference for rotational stability is defined as 2°.

详细描述

Although there may be no direct benefit to the subjects under study, the investigation will assess the stability which determines the efficacy for future toric optics and the amount of glistening and PCO formation of the IPure and ZCB00 hydrophobic acrylic IOLs.

25 eyes of 25 patients will be recruited for each IOL, that is 50 eyes of 50 patients. The randomization will be done with a binomial law. The IOL have a priori the same IOL constants which eliminates the necessity to know which IOL is going to be implanted at the preoperative time. A clinically relevant difference for rotational stability will be 2°. The reasonable standard deviation of the position of an IOL in the eye is 2.4°. A total of 48 patients is necessary for a probability of 80% to measure a clinically relevant difference concerning the IOL stability.

In the week before surgery, the eye to be operated is examined at the slit-lamp and by indirect fundoscopy and measurement of intraocular pressure (IOP) will be performed routinely. Additionally, routine biometry is performed using the IOLMaster (Zeiss Meditec AG, Jena, Germany) for axial length measurement, K-readings and anterior chamber depth (ACD) measurements.

Randomisation will be performed using online randomisation and patients will be allocated 1:1 either to the study group (iPure group), or to the control group (ZCB00).

A self sealing incision, injection of viscoelastic substance, capsulorhexis, phacoemulsification, irrigation/aspiration of cortical material and injection of viscoelastic substance into the capsular bag are performed in a standardised fashion. The IOL is implanted via injector into the capsular bag. Following the implantation of the IOL, the viscoelastic substance is aspirated thoroughly from the anterior chamber, as well as retro-lentally, to assure complete removal of the OVD.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Parallel
主要目的
Treatment
盲法
Double (Participant, Investigator)

入排标准

年龄范围
21 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Age-related cataract
  • Age 21 and older
  • Visual Acuity > 0.05
  • Normal findings in the medical history and physical examination unless the investigator considers an abnormality to be clinically irrelevant.

排除标准

  • Relevant other ophthalmic diseases such as pseudoexfoliation syndrome, floppy iris syndrome, corneal pathologies
  • Previous ocular surgery or trauma.
  • Pregnancy (pregnancy test will be taken pre-operatively in women of reproductive age)

结局指标

主要结局

Rotational Stability

时间窗: 2 years

measured using the Purkinje meter (Murcia University) and retroillumination images (degrees of rotation)

次要结局

  • Tilt in degrees(2 years)
  • visual acuity(2 years)
  • Decentration in mm(2 years)
  • Anterior chamber depth(2 years)
  • Glistening(2 years)
  • Capsulorhexis area(2 years)
  • PCO(2 years)