Physician-modified Endovascular Graft for Repair of Complex Thoracoabdominal and Abdominal Aortic Aneurysms
- Conditions
- Juxtarenal Aortic AneurysmSuprarenal Aortic AneurysmThoracoabdominal Aortic Aneurysm
- Interventions
- Device: Physician-Modified Cook Zenith Alpha Thoracic Endovascular Graft
- Registration Number
- NCT04526938
- Lead Sponsor
- Medstar Health Research Institute
- Brief Summary
The primary objective of this study is to assess the use of a physician-modified Cook Alpha Thoracic Endovascular Graft in the repair of complex aortic aneurysms and thoracoabdominal aortic aneurysms and aneurysms secondary to aortic dissections in high-risk patients having appropriate anatomy. The primary intent of the study is to assess the safety and preliminary effectiveness of the device. Additionally, the study will assess renal function, radiation exposure, and quality of life.
- Detailed Description
This study is a prospective, single-center, non-randomized, single-arm study. Each enrolled subject will undergo periodic follow-up evaluations involving physical exams, computed tomography angiography (CTA) of chest, abdomen and pelvis, abdominal duplex ultrasound, creatinine measurement, quality of life questionnaire at specific, protocol-defined intervals for a period of five years following the physician-modified Cook Zenith® Alpha Endovascular Graft implant.
Recruitment & Eligibility
- Status
- ACTIVE_NOT_RECRUITING
- Sex
- All
- Target Recruitment
- 200
-
Age ≥ 18 years old
-
Life expectancy > 2 year(s)
-
Patient is considered to be at high risk of morbidity and mortality with open surgical repair
-
A patient may be entered into the study if the patient has at least one of the following:
- Juxtarenal, suprarenal or a thoracoabdominal aortic aneurysm with a diameter ≥ 5.5 cm or 2 times the normal aortic diameter
- Aneurysm with a history of growth ≥ 0.5 cm per year
- Saccular aneurysms deemed at significant risk for rupture based upon physician interpretation
- Symptomatic juxtarenal, suprarenal or a thoracoabdominal aortic aneurysm
- Thoracoabdominal aortic aneurysms secondary to aortic dissection meeting above criteria
-
Patient is considered to be at high risk of morbidity and mortality with open surgical repair
-
Iliac or femoral access vessel morphology that is compatible with vascular access techniques, devices or accessories, with or without use of a surgical conduit
-
Nonaneurysmal aortic segment proximal to the aneurysm (neck) with a:
- Minimum neck length of 20mm
- Diameter in the range of 21-38mm
- Angle less than 60° relative to the axis of the aneurysm
- Angle less than 60° relative to the axis of the suprarenal aorta
-
Minimum branch vessel diameter greater than 5 mm
-
Iliac artery distal fixation site greater than 10mm in length and diameter in the range of 8-21mm.
- Under the age of 18 years
- Can be treated in accordance with the instructions for use with a legally marketed endovascular prosthesis
- Is eligible for enrollment in a manufacturer-sponsored IDE at the investigational site
- Willing to participate in a sponsor-investigator IDE with access to a non-physician modified endovascular prostheses at the investigational site
- Willing to travel to an investigational site with access to a non-physician modified endovascular prosthesis
- Unwilling or unable to comply with the follow-up schedule
- Inability or refusal to give informed consent by patient or legal representative
- Patient is pregnant or breastfeeding
- Life expectancy < 2-years despite successful aneurysm exclusion
- Free rupture with hemodynamic instability
- Ongoing infection
- Diagnosis of, or suspected connective tissue disease
- Known sensitivities or allergies to the materials of construction of the devices
- Known hypersensitivity or contraindication to anticoagulation or contrast media that cannot be adequately medically managed
- Uncorrectable coagulopathy
- Body habitus that would inhibit x-ray visualization of the aorta or exceeds the safe capacity of the equipment
- Patient has had a major surgical or interventional procedure unrelated to the treatment of the aneurysm planned ≤ 30 days of the endovascular repair
- Unstable angina (defined as angina with a progressive increase in symptoms, new onset at rest or nocturnal angina)
- Systemic or local infection that may increase the risk of endovascular graft infection
- Leaking or free rupture of aneurysm associated with hemodynamic instability.
- Baseline creatinine greater than 3.0 mg/dL
- Known history of, or suspected connective tissue disorders (e.g., Marfan Syndrome, Ehler's Danlos Syndrome)
Anatomical exclusion criteria:
-
Inadequate femoral or iliac access compatible with the required delivery systems, not amenable to open surgical or endovascular conduit placement
-
Absence of a non-aneurysmal aortic segment for proximal seal zone with:
- A diameter measured outer wall to outer wall of no greater than 38mm and no less than 21 mm;
- Parallel aortic wall with <20% diameter change and without significant calcification and/or thrombus in the selected area of seal zone
-
Visceral vessel anatomy not compatible with physician-modified Zenith® Alpha Endovascular Graft due to excessive occlusive disease or small size not amenable to stent graft placement
-
Unsuitable distal iliac artery fixation site and anatomy:
- Common iliac artery fixation site diameter measured outer wall to outer wall on a sectional image (CT) <8.0 mm with inability to perform open surgical conduit
- Iliac artery diameter measured outer wall to outer wall on a sectional image (CT) >21 mm at distal fixation site, with inability to perform open internal iliac artery revascularization or iliac branch stent graft
- Iliac artery distal fixation site <10 mm in length
- Inability to preserve at least one hypogastric artery
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description Endovascular repair Physician-Modified Cook Zenith Alpha Thoracic Endovascular Graft Endovascular repair of complex aortic aneurysms and thoracoabdominal aortic aneurysms including those secondary to aortic dissection using a physician-modified endovascular graft.
- Primary Outcome Measures
Name Time Method Proportion of study subjects with treatment success at 1 year 1 year Treatment success is defined as a composite of technical success and freedom from the following:
* Aneurysm enlargement \[i.e., \>5mm) as compared to a preoperative CT measure using orthogonal (i.e, perpendicular to the centerline) measurements\]
* Aneurysm rupture
* Aneurysm-related mortality
* Conversion to open repair
* Secondary intervention for migration, Type I and III endoleaks, device integrity failure (e.g., fracture), and patency-related events (i.e., device component stenosis or occlusion and embolic events)Freedom from Major Adverse Events (MAE) at 30 days 30 days Major Adverse Events include: Death, Bowel Ischemia, myocardial infarction, paraplegia/paraparesis, renal failure, respiratory failure, and stroke
Technical success 1 year Technical success is defined as:
* Successful delivery (i.e., ability to deliver the implant to the intended implantation site, without the need for unanticipated corrective intervention related to delivery) and deployment of the device at the intended location,
* Patency of all endovascular graft and branch stent components
* Absence of device deformations requiring unplanned placement of an additional device
* Absence of inadvertent covering of aortic branch vessels
* Successful withdrawal (i.e., successful withdrawal of the delivery system, without need for unanticipated corrective intervention related to withdrawal)
- Secondary Outcome Measures
Name Time Method Quality of Life measures 30 days, 6 months; 1, 2, 3, 4 and 5 years Questionnaire using SF - 36™ Health Survey
Trial Locations
- Locations (1)
MedStar Washington Hospital Center
🇺🇸Washington, District of Columbia, United States