Prospective, ranDomized, Controlled, multicEnter Study for the Treatment of CALCIFied Coronary Artery Lesions With Rotational Atherectomy vs Intravascular LithotripsY
试验速览
- 阶段
- 不适用
- 发起方
- 入组人数
- 100
- 主要终点
- Percent of stent expansion evaluated by Optical Coherence Tomography assessed by core lab.
研究概览
简要总结
Prospective, 1:1 randomized, controlled, multicenter trial to assess effectiveness and safety of Intravascular Lithotripsy (IVL) compared to Rotational Atherectomy (RA) treatment in calcified coronary lesions
详细描述
- Introduction
Calcified coronary lesions are frequently observed and with increasing age and comorbidities of the patients they become even more prevalent (1,2). Calcified lesions are often difficult to dilate by conventional angioplasty leading to vessel perforation, dissection, or stent under-expansion, which impact survival, myocardial infarction rates, and target lesion revascularization (3 - 6). Lesion preparation before stent implantation has been strongly recommended using high-pressure dilatation, scoring/cutting balloons, or rotational atherectomy (RA) devices (7-12).
RA can effectively modify calcified plaques by differential cutting, facilitating balloon dilatation, plaque fracture, stent delivery, and expansion (11). Patients undergoing RA experienced various clinical and angiographic complications, including vasospasm, perforation, short term vessel closure, side branch loss, and slow-flow/no-reflow resulting in stroke, myocardial infarction, and death (12). In the PREPARE-CALC trial (The Comparison of Strategies to Prepare Severely Calcified Coronary Lesions Trial) comparing RA and scoring or cutting balloons before drug-eluting stent implantation achieved a higher procedural success in RA (98% versus 81%), complications including larger dissection, perforation, and pericardial effusion were seen in 3%, 4%, and 3% of cases, respectively, in-hospital MACE occurred in 2% of cases (13). In the ORBIT II trial (14) an alternative atherectomy device was used which is based on the same mechanism of action as RA. In this trial the primary effectiveness performance goal, defined as successful stent delivery with residual stenosis < 50% and without in-hospital MACE, was not achieved in 11.1 % and the in-hospital MACCE rate was at 9.8 %, the rate of periprocedural myocardial infarction (MI), defined as CK-MB level > 3x ULN at discharge was 9.3% (non-Q-wave MI of 8.6 % and Q-wave MI of 0.7 %). Peri-procedural MI based on the 4th Universal Definition (15) after RA or ORBIT atherectomy have not been reported so far.
Intravascular lithotripsy (IVL) disrupts subendothelial calcification by electrohydraulic-generated sonic pressure waves (16,17). In a recently published multicenter registry 78 patients were assigned to primary IVL therapy for patients with calcified de-novo lesions (n=39 lesions), secondary IVL therapy for patients with calcified lesions in which non-compliant balloon dilatation failed (n=22 lesions), and tertiary IVL therapy in patients with stent under expansion after previous stenting (n=17 lesions). The primary endpoint of strategy success (stent expansion with <20% in-stent residual stenosis) was reached in 84.6% with de-novo lesions with only 3 Type B dissections and no in-hospital MACCE (18). Recently Intravascular Lithotripsy for Treatment of Severely Calcified Coronary Artery Disease (Disrupt CAD III) Study (19) has been presented using the same in- and exclusion criteria as the ORBIT II study. The performance of IVL was compared with the performance of the ORBIT device, as being reported in the literature (14). The primary safety endpoint (cardiac death, or myocardial infarction, or target vessel revascularization) as well as the primary effectiveness endpoint (successful stent delivery with residual stenosis < 50 % and without in-hospital MACE) were 92.2 %, which were non-inferior to those reported in the ORBIT II study. Stent expansion was measured by QCA, data on OCT-based measurements of stent expansion in comparison to RA is not available. The rate of in-hospital peri-procedural MI, defined as CK-MB level > 3x ULN at discharge, was 6.8 %. Peri-procedural myocardial infarction and injury are associated with an increased rate of cardiovascular events at 30 days in patients undergoing elective coronary stenting (20).
In order to evaluate the effectiveness of IVL in comparison to RA based on OCT- measurements of stent expansion and to study the incidence of in-hospital MACCE following either RA or IVL, we aim to perform a randomized trial. 2. Device Description and Intended Use
研究设计
- 研究类型
- Interventional
- 分配方式
- Randomized
- 干预模型
- Parallel
- 主要目的
- Treatment
- 盲法
- Single (Outcomes Assessor)
盲法说明
Core Lab is blinded
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Subject age ≥ 18 years
- •Subject has been informed of the nature of the study, agrees to participate and has signed a Medical Ethics Committee approved consent form, understands the duration of the study, agrees to attend follow-up visits
- •LVEF >25%
- •Single de-novo target lesion with stenosis ≥ 70% and < 100% or ≥ 50% and < 70% with evidence of ischemia, or FFR ≤ 0.80, or lumen area ≤ 4.0 mm2 (≤ by 6.0mm2 in left main) IVUS or OCT
- •Target vessel RVD ≥ 2.5 mm and ≤4.0 mm
- •Lesion length ≤ 60 mm
- •Lesion site severe calcification: Angiographic radio-opacities prior to contrast involving both sides of arterial wall with total calcium length ≥10 mm, or presence of ≥270° of calcium on at least one cross section by IVUS or OCT or incomplete balloon expansion at 12 atm
- •Target lesion was traversed by a guide wire
排除标准
- •Failure to successfully cross the target lesion with the guidewire
- •Target lesion in a coronary artery bypass graft
- •In-stent-Restenosis
- •Thrombus in the target vessel
- •Chronic total occlusion in the target vessel
- •ST-elevation myocardial infarction (STEMI) within the last 4 weeks prior to enrollment
- •Stroke within the last 3 months prior to enrollment
- •Decompensated heart failure
- •Life expectancy of less than one year
- •Chronic kidney disease (serum creatinine > 2.5mg/dl)
- •Pregnant or lactating females
- •Receiving dialysis or immunosuppressant therapy
- •Platelet Count < 100.000mm3 or > 600.000mm3
- •Significant gastrointestinal bleeding or any coagulopathy that would contraindicate the use of antiplatelet therapy
- •Known allergies or sensitivity to heparin, aspirin, other anticoagulant/antiplatelet therapies or contrast media that cannot be adequately pre-treated prior to index procedure
- •Surgical procedure within the last 30 days prior to enrollment
结局指标
主要结局
Percent of stent expansion evaluated by Optical Coherence Tomography assessed by core lab.
时间窗: at the end of the intervention
Primary Effectiveness endpoint
Rate of in-hospital MACCE
时间窗: 72 hours
primary Safety endpoint
次要结局
- Rate of peri-procedural myocardial injury ,Peri-procedural myocardial infarction, Slow flow/no-reflow(72 hours)
- Percentage of mean stent expansion evaluated by optical coherence tomography as assessed by core lab.(at the end of the intervention)