临床试验/NCT03528174
NCT03528174已完成不适用

Assessment of a Novel Sensing Catheter During Automated Insulin Delivery in Patients With Type 1 Diabetes

Oregon Health and Science University2 个研究点 分布在 1 个国家目标入组 10 人开始时间: 2018年4月16日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
10
试验地点
2
主要终点
Mean Absolute Relative Difference of Sensed Glucose Values

研究概览

简要总结

The study described within this protocol is designed to test a single hormone closed-loop algorithm for managing blood glucose in type 1 diabetes using insulin only and a new device for measuring glucose continuously in the immediate vicinity of subcutaneous insulin delivery. Subjects will undergo one all day closed loop study using insulin only. The closed loop insulin only system includes a Dexcom G5 transmitter, a Nexus 5 smart phone and two Tandem tslim pumps connected to two Pacific Diabetes Technologies glucose sensing cannulas. The closed loop system will be started after G5 2 hour sensor start-up with an IV catheter for frequent blood withdrawal for 10 hours.

详细描述

Subjects will undergo one 12 hour inpatient study. During this intervention visit, subjects will wear the single hormone closed loop system which includes two t:slim insulin pumps to deliver insulin or saline through two investigational PDT glucose sensing cannulas, and a Dexcom G5 CGM to measure glucose. The subject will come to the research center at approximately 7am for the inpatient visit. The Dexcom G5 sensor and PDT glucose sensing cannulas will be placed after arrival. An 18-22 gauge IV catheter will be placed for blood sampling every 15 minutes after sensor warm-up is complete for measuring blood glucose concentration. After a 2 hour warm-up period for the G5 sensor, glucose will be controlled using the FMPD single hormone mode. The single hormone mode of the controller determines insulin delivery rates based on proportional and derivative error through one t:slim insulin pump. The second t:slim insulin pump will deliver normal saline at the same delivery rates as the insulin. Subjects will eat breakfast and lunch at approximately 11 am and 3 pm respectively. Subjects will have the ability to complete light exercise on a treadmill. The closed loop system will be stopped at approximately 6pm with removal of all devices.

During the study, the subject will wear one subcutaneous DexcomTM G5 continuous glucose monitoring (CGM) system. The CGM system will provide sensed glucose data every 5 minutes. The accuracy of the sensed data will be obtained by reference measurements of two YSI venous blood glucose values to calibrate the sensor at the beginning of the closed loop study. Sensed glucose data will be wirelessly transmitted via Bluetooth Low Energy (BTLE) from the Dexcom G5 transmitter to the Nexus 5 master controller every five minutes. The controller is a Google Nexus 5 phone. The smart phone will wirelessly communicate via BTLE to two Tandem t:slim insulin pumps, one for automated insulin delivery and one of automated saline delivery.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Device Feasibility
盲法
None

入排标准

年龄范围
21 Years 至 65 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Diagnosis of type 1 diabetes mellitus for at least 1 year.
  • Male or female subjects 21 to 65 years of age.
  • Current use of an insulin pump for at least 3 months with stable insulin pump settings for > 2 weeks.
  • HbA1c ≤ 10% at screening.
  • Body Mass Index ≥
  • Total daily insulin requirement is less than 200 units/day.
  • Willingness to follow all study procedures, including attending all clinic visits.
  • Willingness to sign informed consent and HIPAA documents.

排除标准

  • Female of childbearing potential who is pregnant or intending to become pregnant or breast-feeding, or is not using adequate contraceptive methods. Acceptable contraception includes birth control pill / patch / vaginal ring, Depo-Provera, Norplant, an IUD, the double barrier method (the woman uses a diaphragm and spermicide and the man uses a condom), or abstinence.
  • Any cardiovascular disease, defined as a clinically significant EKG abnormality at the time of screening or any history of: stroke, heart failure, myocardial infarction, angina pectoris, or coronary arterial bypass graft or angioplasty. Diagnosis of 2nd or 3rd degree heart block or any non-physiological arrhythmia judged by the investigator to be exclusionary.
  • Renal insufficiency (GFR < 60 ml/min, using the MDRD equation as reported by the OHSU laboratory).
  • Liver failure, cirrhosis, or any other liver disease that compromises liver function as determined by the investigator.
  • Hematocrit of less than 36% for men, less than 32% for women.
  • History of severe hypoglycemia during the past 12 months prior to screening visit or hypoglycemia unawareness as judged by the investigator.
  • History of Diabetic Ketoacidosis during the prior 6 months prior to screening visit, as diagnosed on hospital admission or as judged by the investigator.
  • Adrenal insufficiency.
  • Any active infection.
  • Known or suspected abuse of alcohol, narcotics, or illicit drugs.
  • Seizure disorder.
  • Active foot ulceration.
  • Severe peripheral arterial disease characterized by ischemic rest pain or severe claudication.
  • Major surgical operation within 30 days prior to screening.
  • Use of an investigational drug within 30 days prior to screening.
  • Chronic usage of any immunosuppressive medication (such as cyclosporine, azathioprine, sirolimus, or tacrolimus).
  • Bleeding disorder, treatment with warfarin, or platelet count below 50,
  • Allergy to lispro insulin.
  • Allergy to acrylate-based skin adhesives.
  • Need for uninterrupted treatment of acetaminophen.
  • Current administration of oral or parenteral corticosteroids.
  • Any life threatening disease, including malignant neoplasms and medical history of malignant neoplasms within the past 5 years prior to screening (except basal and squamous cell skin cancer).
  • Beta blockers or non-dihydropyridine calcium channel blockers.
  • Current use of any medication intended to lower glucose other than insulin (ex. use of liraglutide).
  • Any clinically significant disease or disorder which in the opinion of the Investigator may jeopardize the subject's safety or compliance with the protocol.

研究组 & 干预措施

Single Hormone closed loop

Experimental

Subjects will have glucose managed using the Artificial Pancreas Control system (APC) using insulin only. Insulin will be infused through the Pacific Diabetes Technologies CGM Insulin Infusion system.

干预措施: Artificial Pancreas Control system (APC) (Device)

Single Hormone closed loop

Experimental

Subjects will have glucose managed using the Artificial Pancreas Control system (APC) using insulin only. Insulin will be infused through the Pacific Diabetes Technologies CGM Insulin Infusion system.

干预措施: Pacific Diabetes Technologies CGM Insulin Infusion system (Device)

结局指标

主要结局

Mean Absolute Relative Difference of Sensed Glucose Values

时间窗: up to 10 hours

Assess the mean absolute relative difference for the PDT CGM/insulin infusion system based on all reference YSI venous blood glucose values.

次要结局

  • Percent of Time With Sensed Glucose Between 70-180 mg/dl(up to 10 hours)
  • Mean Sensed Glucose(up to 10 hours)
  • Mean Score of Visual Analog Scale at 15 Minutes(15 minutes post insertion)
  • Mean Draize Scale for Edema for the Left Sensing Cannula at End of Study for All Subjects(end of 10 hour study)
  • Percent of Time With Sensed Glucose Less Than 54 mg/dl(up to 10 hours)
  • Percent of Time With Sensed Glucose Greater Than 250 mg/dl(up to 10 hours)
  • Mean Draize Scale for Erythema for the Right Sensing Cannula at End of Study for All Subjects(End of 10 hour study)
  • Percent of Time With Sensed Glucose Less Than 70 mg/dl(up to 10 hours)
  • Mean Draize Scale for Erythema for the Left Sensing Cannula at End of Study for All Subjects(End of 10 hour study)
  • Mean Score of Visual Analog Scale at End of Study(End of 10 hour study)
  • Number of Carbohydrate Treatments(up to 10 hours)
  • Mean Amount of Insulin Delivered(up to 10 hours)
  • Mean Score of Visual Analog Scale at 120 Minutes(120 minutes post insertion)