Evaluation of the Accuracy of the Hemochron Signature Elite Point of Care Micro Coagulation Instrument Compared With Laboratory Assay

NHS Lothian1 个研究点 分布在 1 个国家目标入组 24 人开始时间: 2013年10月1日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
NHS Lothian
入组人数
24
试验地点
1
主要终点
The International Normalised Ratio (INR) Will be Measured by the Laboratory Assay and the Hemochron Signature Elite Device.

研究概览

简要总结

Central Neuraxial Blockade (CNB) is commonly used in anaesthetics. It has several advantages over a general anaesthetic however contraindications include coagulopathy. Guidelines suggest a CNB should not be undertaken with an International Normalised Ratio (INR) of 1.5 or greater. A common cause of an elevated INR is warfarin therapy and regular monitoring of its effects is required.

The Hemochron Signature Elite is a portable micro coagulation system that can be used as a point of care (POC) anticoagulation monitoring device. Marketing is focussed on its accuracy, convenience and efficiency in delivering a coagulation result.

This study will assess the accuracy of this device in calculating the International Normalised Ratio (INR) result. This will be compared to another INR result obtained on the same sample via the current gold standard testing equipment in the Royal Infirmary Edinburgh (RIE) laboratories.

The study will include patients presenting for surgery who have had their warfarin therapy stopped within the previous 7 days. Patients on warfarin therapy requiring surgery are currently identified in preoperative assessment clinics, on average 12 weeks prior to their date of surgery.

Anecdotally there appears to be a difference between the 'expected' and actual INR when a POC device is used. Concern exists that patients may be inappropriately denied the potential benefits of central neural blockade (CNB) if a POC device provides a falsely high result, or conversely the investigators may be proceeding with CNB in patients with an unacceptably high INR if POC provides a falsely low result. There is also the possibility of improving hospital efficiency using this device as it provides a result much faster than the laboratory.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

性别
All
接受健康志愿者

入选标准

  • Adult patients presenting for elective surgery who have stopped their warfarin therapy within the last 7 days.

排除标准

  • Adults with incapacity

研究组 & 干预措施

Surgical Patient

These are the results from the blood samples taken from the study participants.

结局指标

主要结局

The International Normalised Ratio (INR) Will be Measured by the Laboratory Assay and the Hemochron Signature Elite Device.

时间窗: 18 months

The correlation between the international normalised ratio measured by the laboratory assay and the Hemochron Signature Elite device was assessed.Number of participants with a higher INR measurement by the Hemochron Signature Elite device than laboratory assay