MedPath

Validation of Belun Ring Gen3 Deep Learning Algorithms With Subxiphoid Body Sensor

Not Applicable
Recruiting
Conditions
Sleep-Disordered Breathing
Sleep Architecture
Arrhythmia
Interventions
Device: Belun Ring
Device: Belun Cor
Registration Number
NCT06255613
Lead Sponsor
Belun Technology Company Limited
Brief Summary

Hypothesis: BR's Gen3 DL algorithms, combined with its subxiphoid body sensor, can accurately diagnose OSA, categorize its severity, identify REM OSA and supine OSA, and detect central sleep apnea (CSA).

Primary Objective:

To rigorously evaluate the overall performance of the BR with Gen3 DL Algorithms and Subxiphoid Body Sensor in assessing SDB in individuals referred to the sleep labs with clinical suspicion of sleep apnea and a STOP-Bang score \> 3, by comparing to the attended in-lab PSG, the gold standard.

Secondary Objectives:

To determine the accuracy of BR sleep stage parameters using the Gen3 DL algorithms by comparing to the in-lab PSG;

To assess the accuracy of the BR arrhythmia detection algorithm;

To assess the impact of CPAP on HRV (both time- and frequency-domain), delta HR, hypoxic burden, and PWADI during split night studies;

To assess if any of the baseline HRV parameters (both time- and frequency-domain), delta heart rate (referred to as Delta HR), hypoxic burden, and pulse wave amplitude drop index (PWADI) or the change of these parameters may predict CPAP compliance;

To evaluate the minimum duration of quality data necessary for BR to achieve OSA diagnosis;

To examine the performance of OSA screening tools using OSA predictive AI models formulated by National Taiwan University Hospital (NTUH) and Northeast Ohio Medical University (NEOMED).

Detailed Description

Not available

Recruitment & Eligibility

Status
RECRUITING
Sex
All
Target Recruitment
79
Inclusion Criteria
  • Provision of signed informed consent form.
  • Clinically assessed and suspicious for OSA with a STOP-Bang score ≥ 3.
Exclusion Criteria
  • Full night PAP titration study.
  • On home O2, noninvasive ventilator, diaphragmatic pacing, or any form of a nerve stimulator.
  • Having atrial fibrillation-flutter, pacemaker/defibrillator, left ventricular assist device (LVAD), or status post cardiac transplantation.
  • Recent hospitalization or recent surgery in the past 30 days.
  • Unstable cardiopulmonary status on the night of the study judged to be unsafe for sleep study by the sleep tech and/or the on-call sleep physician.

If a participant did not sleep for at least 4 hours of technically valid sleep based on the Belun Ring method for diagnostic assessments, or a minimum of 3 hours of technically valid sleep during the diagnostic phase of a split-night study, the patient will be excluded from statistical analysis.

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
UH-ParticipantBelun CorPotential participants will be identified from patients scheduled for in-lab PSG at the two medical centers of University Hospitals
UH-ParticipantBelun RingPotential participants will be identified from patients scheduled for in-lab PSG at the two medical centers of University Hospitals
Primary Outcome Measures
NameTimeMethod
Performance of diagnosing sleep disordered breathing by BR's Gen3 DL algorithms2 years

To establish mean, standard deviation, and target clinical agreement limit values for the differences in sleep-disordered breathing parameters (including AHI3%, AHI4%, REM AHI 3%, REM AHI4%, Supine AHI3%, Supine AHI4%, and CAI) from the BR against PSG. To calculate the sensitivity and specificity values, along with 95% confidence intervals (CIs), for sleep-disordered breathing parameters from the BR against PSG at PSG cutoffs of 5, 15, and 30 events/h. Pearson correlation and regression analysis will be employed to assess the association between sleep-disordered breathing parameters obtained from the BR and PSG.

Secondary Outcome Measures
NameTimeMethod
Accuracy of the NTUH and NEOMED models3 years

The investigators will compute sensitivity and specificity values, accompanied by their respective 95% CIs, for AHI3% and AHI4% from the NTUH/NEOMED models in comparison to PSG. Furthermore, the investigators will determine the ROC curve at PSG-defined cutoffs of 5, 15, and 30 events/h for both AHI3% and AHI4%.

Performance of sleep stage classification by BR's Gen3 DL algorithms2 years

Sensitivity and specificity of the Epoch-by-Epoch (EBE) 4-stage classification (wake, deep sleep, light sleep, REM) of the BR will be compared to EBE PSG data for each stage against the combination of the other three classifications. For instance, sensitivity for detecting wake will measure the proportion of correctly classified wake epochs against the combined classifications of deep sleep, light sleep, and REM sleep. Sensitivity and specificity values, accompanied by 95% CIs, of EBE body position (supine vs. non-supine) against PSG will also be computed.

To evaluate the minimum duration of quality data necessary for OSA diagnosis2 years

To explore the minimum number of hours needed by the BR for an accurate OSA diagnosis, sensitivity, specificity, and Matthews Correlation Coefficient (MCC) values, along with 95% CIs, will be computed for sleep-disordered breathing parameters of Belun total record time (bTRT) and belun total sleep time (bTST) with different numbers of hours (e.g., 2, 3, 4, 5, 6, 7, and 8 hours) against PSG data. In cases where varying numbers of subjects meet different bTRT and bTST criteria at different durations, resampling methods like oversampling, undersampling, bootstrapping, etc., will be employed to address data imbalances.

Accuracy of the BR arrhythmia-detecting algorithm3 years

To calculate the performance metrics, including sensitivity and specificity, along with 95% CIs, for the detection of arrhythmias by BR device in comparison to assessments made by a board-certified sleep physician. The physician will review the PSG ECG signals of the subjects suspected of having significant arrhythmias during the sleep study.

Biomarker relationship analysis3 years

Pearson correlation and regression analysis will be employed: (1) to explore the relationships between autonomic nervous system (ANS) metrics and delta HR, hypoxic burden, and pulse wave amplitude drop index (PWADI); (2) to evaluate the association between HRV, Delta HR, hypoxic burden, PWADI, CPAP adherence, sleep-disordered breathing parameters, and demographic data (including age, gender, and BMI). If data distribution assumptions are violated, the Spearman correlation will be used instead of the Pearson correlation.

Trial Locations

Locations (2)

UH Geauga Health Center Services

🇺🇸

Chardon, Ohio, United States

University Hospitals Cleveland Medical Center

🇺🇸

Cleveland, Ohio, United States