Testing the Effectiveness and Safety of a New Prototype of a Passive Industrial Exoskeleton for Back Support
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 9
- 试验地点
- 2
- 主要终点
- Metabolic cost of ergospirometric testing
研究概览
简要总结
This research will provide data on evaluation of the effectiveness and safety of using a passive exoskeleton designed to reduce the negative impact of static and dynamic loads associated with body tilts, lifting and holding weights on the musculoskeletal system of workers. To evaluate a passive industrial exoskeleton for back support, a protocol was developed in which individual production operations will be modeled.
This research will recruit 9 valunteers. During a 4-hour laboratory visit, each subject will perform three identical experimental tests: one test without an exoskeleton and two tests using two different modifications of the exoskeleton prototype. To avoid sequence bias, the test order will be randomized (balanced). Two types of labor operations will be simulated. One operation will consist in lifting the load, which is 20% of the volunteer's body weight, from a height of 14 cm to a surface with a height of 70 cm, the frequency of lifting is 1 time in 30 seconds. The second operation is holding a cargo weighing 5 kg in a forced working position (tilt of the body more than 30 °) for 1 minute, after which rest will follow for 1 minute, and then the work will be repeated. Both types of work will be performed for 105 minutes, then a break of 15 minutes, then 105 minutes of work. Heart rate, blood pressure, ECG (Custo) will be monitored. An ergospirometric device (Metamax) will be used to collect muscle activity data (Trust-M, Myoton), inertial sensors (Trust-M) will be used to evaluate the biomechanics of movements. In addition, questionnaires will be filled out for a subjective assessment of the use of exoskeleton prototypes (local discomfort scale, exoskeleton properties assessment questionnaire).
研究设计
- 研究类型
- Interventional
- 分配方式
- Randomized
- 干预模型
- Crossover
- 主要目的
- Prevention
- 盲法
- Single (Participant)
盲法说明
Blinding is not possible in this testing set-up, participants however are not informed about the exact purpose of the tests and which exoskeleton is a prototype and which exoskeleton is a commercially available exoskeleton
入排标准
- 年龄范围
- 18 Years 至 45 Years(Adult)
- 性别
- Male
- 接受健康志愿者
- 是
入选标准
- •healthy individuals (based on medical screening)
排除标准
- •musculoskeletal system diseases
- •musculoskeletal injuries
- •health deviations at the time of the study
结局指标
主要结局
Metabolic cost of ergospirometric testing
时间窗: 4 hours
A Metamax 3B instrument (Cortex, Germany) will be used to measure a metabolic cost during all labor activities, 4 hours.
Blood pressure
时间窗: 6 hours
A Custo med blood pressure monitor will be used by all subjects throughout the study
Heart rate
时间窗: 6 hours
A Polar H10 heart rate monitor (Polar Electro, Finland) will carry all items during the entire research
Myotonometry
时间窗: 1.5 hours
Myotonometry will be performed with a Myoton Pro device. Parameters of the certain muscles, which get the maximum load, will be measured at rest standing before and after work, as well as outside work in a stress position without an exoskeleton and with it.
Range of motion in large joints of the upper, lower limbs and spine
时间窗: 1 hour
Inertial sensors will be mounted on the body of the volunteer. The amplitude of the moving in the shoulder, elbow, hip, knee joints and in the thoracolumbar spine is determined. Flexion and extension, abduction and adduction, rotation are performed.
Surface electromyography
时间窗: 1.5 hours
Trust-M sensors with electrodes will be used to collect electromyographic data in accordance with the SENIAM recommendations. The muscles that will be monitored, the muscles of the arms, legs and back, which account for the maximum load. The locations of the sensors will be prepared by shaving, abrasion with sandpaper, and alcohol cleaning. The average amplitude of the EMG spectrum will be collected. The recording will be carried out at rest, standing before and after work, as well as outside work in the voltage position without exoskeleton and with it.
Local Comfort Rating Scale
时间窗: 1 hour
This questionnaire is for assessing the duration of maintaining a working posture without discomfort, fixing their appearance, as well as the nature of these sensations. Points can vary from 1 to 10, where 1 is the absence of pain, 10 is unbearable pain.
次要结局
- Exoskeleton parameter evaluation form(30 minutes)
研究者
Andrey Mikhailovich Geregey
Head of the Laboratory of Personal Protective Equipment and Industrial Exoskeletons Andrey Geregey
Federal State Budgetary Scientific Institution "Izmerov Research Institute of Occupational Health"