Effect of Cardiac Resynchronization Therapy (CRT) on the Defibrillation Threshold (DFT) Estimates

Abbott Medical Devices2 个研究点 分布在 1 个国家目标入组 77 人开始时间: 2008年1月1日最近更新:
适应症
干预措施

试验速览

阶段
4 期
状态
已完成
入组人数
77
试验地点
2
主要终点
Defibrillation Threshold Difference Obtained in Volts (V) Between Implant and 6 Months

研究概览

简要总结

The purpose of this prospective study is to evaluate the effect of cardiac resynchronization therapy (CRT) on the defibrillation threshold (DFT) estimates in cardiac resynchronization therapy defibrillators (CRT-D) patients. The hypothesis of the study is that defibrillation threshold (DFT) will decrease with 6 months of cardiac resynchronization therapy (CRT).

详细描述

Cardiac Resynchronization Therapy (CRT) has emerged as a promising therapeutic addition in patients with drug refractory heart failure (HF). Along with providing relief of symptoms of HF, cardiac resynchronization therapy defibrillators (CRT-D) are used for the prevention of sudden cardiac death (SCD). Although there are concerns that the defibrillation threshold (DFT) estimates are elevated in the heart failure (HF) patient population due to lower LV ejection fraction (EF) and left ventricular (LV) dysfunction, there is paucity of data available to evaluate this theory.

Recently, two different studies in a retrospective manner evaluated the energy requirements in patients receiving cardiac resynchronization therapy defibrillators (CRT-D). Burke et al analyzed DFTs in 50 patients each implanted with a cardiac resynchronization therapy defibrillators (CRT-D) device and an implantable cardioverter defibrillator (ICD). Although the ejection fraction (EF) in cardiac resynchronization therapy defibrillators (CRT- D) group was lower than the implantable cardioverter defibrillator (ICD) group, the mean defibrillation thresholds (DFTs) between the two groups were not significantly different (10.2 ± 6.1 J for the cardiac resynchronization therapy (CRT) group vs. 9.5 ± 5.0 J for the control group)14. In the ASSURE study, Doshi et al. showed that patients receiving cardiac resynchronization therapy defibrillators (CRT-D) devices do not have higher energy requirements when compared to patients receiving modern single or dual chamber implantable cardioverter defibrillators (ICDs). Although, there was a trend toward higher energy requirements found among patients with higher degrees of heart failure.

Major cardiac resynchronization therapy (CRT) trials have showed that cardiac resynchronization therapy (CRT) therapy has positive effects on ejection fraction (EF) and the heart failure (HF) condition of the patient over time. But, there is no study that has evaluated the defibrillation thresholds (DFTs) in the heart failure (HF) patients over time of receiving cardiac resynchronization therapy (CRT). This information will help the clinicians decide if they need to perform more invasive procedures during device implant to lower DFTs in patients who do not meet the defibrillation safety margin or they should just wait over time for the CRT to reduce the defibrillation threshold (DFT). Also, there is no published data about the stability of defibrillation thresholds (DFTs) in heart failure (HF) patients over time. The results from this study will also help to clarify whether it is safe to never test the defibrillation thresholds (DFTs) post-implant in this patient population.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Treatment
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Patient meets standard indications for a Cardiac Resynchronization Therapy - Defibrillator (CRT-D).
  • Patient will be implanted with an FDA approved St Jude Medical (SJM) Cardiac Resynchronization Therapy - Defibrillator (CRT-D) and compatible defibrillation lead system.
  • Patient is able to tolerate defibrillation threshold (DFT) testing.
  • Patient is geographically stable and willing to comply with the required follow-up schedule.
  • Patient has a life expectancy of greater than 6 months from the time of implant.
  • Patient has stable heart failure (HF) medications at least one month prior to enrollment.

排除标准

  • Inability to successfully implant an intravascular lead Cardiac Resynchronization Therapy - Defibrillator (CRT-D) device. (i.e. exclude epicardial leads).
  • Patient is getting his Cardiac Resynchronization Therapy - Defibrillator (CRT-D) device replaced.
  • Inability to successfully obtain the defibrillation threshold (DFT) at implant.
  • Currently participating in a clinical trial that includes an active treatment arm or another data collection registry.
  • Recent (within 24 hours) administration of Nesiritide™.
  • Patient is on amiodarone (other antiarrhythmic agents known to affect defibrillation thresholds - DFTs) at the time of enrollment.
  • Patient is pregnant.
  • Patient is less than 18 years old.

研究组 & 干预措施

Cardiac Resynchronization Therapy - Defibrillator (CRT-D)

Other

Patients in the study who received a Cardiac Resynchronization Therapy - Defibrillator (CRT-D) are indicated for it. It's a single arm study in which patients underwent defibrillation threshold (DFT) testing at implant and 6 months.

干预措施: Cardiac Resynchronization Therapy - Defibrillator (CRT-D) (Device)

结局指标

主要结局

Defibrillation Threshold Difference Obtained in Volts (V) Between Implant and 6 Months

时间窗: Baseline and 6 months

All patients underwent defibrillation threshold testing at cardiac resynchronization therapy-defibrillator (CRT-D) implant and then at 6 months. The outcome measure is the difference in DFT (defibrillation threshold) in volts between implant and 6 months.

次要结局

  • Defibrillation Threshold Difference Obtained in Joules (J)(Baseline and 6 months)