临床试验/NCT01295528
NCT01295528已完成不适用

Verify the functionS of BPM TMB-986 to Comply With ANSI/AAMI SP10

BTS International1 个研究点 分布在 1 个国家目标入组 95 人开始时间: 2010年5月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
95
试验地点
1
主要终点
Verify the accuracy of measure functions of device

研究概览

简要总结

The clinical protocol of the clinical testing of this device:

  1. Objective of the test: To verify the function of device.
  2. Test methods and procedures: Performance test in two positions: Seated and Supine.
  3. DUT: Transtek Blood Pressure Monitor, Model: TMB-986. Cuff size: 22-32cm and 22-42cm.
  4. Comparison device: Yuyue medical BP meter, accuracy: ±1mmHg and range: 0-300mmHg.
  5. Study endpoints: Comply with ANSI/AAMI SP10-2002 standard.
  6. Statistical methodology used: Description of statistical methods.
  7. Result: Meet the requirements of SP10.

详细描述

Hospital Information The data was collected by West China Medical College Clinical Investigator Team at Wuhou District Hospital, No. 9 Tongzilin Road, Wuhou District, Chengdu 610041, P.R. China.

Investigator: Catharine Zhang, Nurse A; Vivian Luo, Nurse B; Dr. James Zhu, Sponsor.

Contact Dr. James Zhu Tel: +8613308036568 Used Equipments DUT (Device Under Test): Transtek Blood Pressure Monitor, TMB-986 Reference Device: Yuyue medical BP meter, BP212, accuracy: ±1mmHg and range: 0-300mmHg.

Test Protocol

  1. Test Purpose:

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Retrospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • male,female,

排除标准

  • below 18 years old

结局指标

主要结局

Verify the accuracy of measure functions of device

时间窗: 10 days

Verify the accuracy of device if meet the requirements of ANSI/AAMI SP10 when it measure Systolic \& Diastolic blood pressure.

次要结局

未报告次要终点