Comparison of Intraocular Pressure Measurements Between Reichert Tono-Vera Tonometer and Goldmann Tonometry
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 61
- 试验地点
- 2
- 主要终点
- Demonstrate compliance with ANSI Z80.10-2014-Ophthalmics-Ophthalmic Instruments-Tonometers.
研究概览
简要总结
Determine if the Tono-Vera Tonometer accurately measures intraocular pressure (IOP). The hypothesis of this test is to confirm the Tono-Vera Tonometer is equivalent to the Goldmann Applanation Tonometer (within +/- 5.0 mmHg).
详细描述
The Tono-Vera Tonometer (Reichert, Inc.) is a portable rebound tonometer intended to measure IOP. Subjects will undergo a standard ophthalmologic examination. Subjects will then be measured with four tonometers: Goldmann Applanation Tonometer (C.S.O. SRL), Ocular Response Analyzer G3 (Reichert, Inc.), iCare ic100 tonometer (iCare Finland Oy), and Tono-Vera Tonometer.
研究设计
- 研究类型
- Interventional
- 分配方式
- Non Randomized
- 干预模型
- Sequential
- 主要目的
- Diagnostic
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 90 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Subjects must be male or female, between the ages of 18 and 90 years old;
- •Be able and willing to provide signed informed consent
- •Be able to follow study instructions
排除标准
- •Subjects with only one functional eye;
- •Subjects with one eye having poor or eccentric fixation;
- •Subjects with central corneal thickness greater than 600 µm or less than 500 µm (about 2 standard deviations the human mean);
- •Subjects with corneal scarring or who have had corneal surgery, including corneal laser surgery;
- •Subjects with concomitant ocular diseases such as: microphthalmos, buphthalmos, nystagmus, keratoconus, severe dry eye syndrome, blepharospasm, any other corneal or conjunctival pathology or infection;
- •Contact lens wearers;
- •Known allergy to proparacaine or fluorescein as these are used to anesthetize the eye and allow IOP measurement, respectively, when used with the GAT.
结局指标
主要结局
Demonstrate compliance with ANSI Z80.10-2014-Ophthalmics-Ophthalmic Instruments-Tonometers.
时间窗: Through study completion, approximately 4 months.
Subjects will be distributed into Low IOP, Medium IOP, and High IOP groups based on Goldmann Applanation Tonometer measurements. The Tono-Vera Tonometer measurements will be within +/-5.0 mmHg of the Goldmann Applanation Tonometer measurements in all groups.
次要结局
- Data collection for two measurement modes.(Through study completion, approximately 4 months.)
- Data collection for device calibration.(Through study completion, approximately 4 months.)