A Prospective Canadian Multi-Center Randomized Study; Spontaneous AtrioVEntricular Conduction PReservation
试验速览
- 阶段
- 3 期
- 状态
- 已完成
- 发起方
- 入组人数
- 450
- 试验地点
- 14
- 主要终点
- Percentage of ventricular pacing
研究概览
简要总结
1 Purpose
In patients with sinus node dysfunction (SND), the use of AAI mode is rare. In paroxysmal high-degree or complete atrio-ventricular (AV) block, paroxysmal vagally-mediated bradycardia, single chamber atrial pacemakers are contraindicated although ventricular support is only rarely needed. As a result, these patients are usually exposed to an unnecessary high proportion of ventricular pacing due to the programming of the system to DDD mode for preventive safety reasons. This drains the pulse generator battery, and more importantly has long-term adverse effects on ventricular function in a subset of vulnerable patients, leading to left ventricular (LV) asynchrony, mitral regurgitation and increased left atrium (LA) size. The deleterious effects of right ventricular (RV) pacing, which increased heart failure (HF) morbidity/mortality, atrial fibrillation (AF) and related adverse events have been highlighted in several large clinical studies. Limiting RV pacing in non-dependant patients is therefore desirable.
The Symphony DR 2550 cardiac pacemakers (SORIN Group, Montrouge, France) offer a new pacing mode that automatically switches from AAI(R) mode to DDD(R) or DDI(R) in event of severe atrioventricular conduction disorder, returning spontaneously to AAI(R) mode as soon as the spontaneous AV conduction has resumed. This particular mode is called the AAIsafeR2(R).
The objective of this study is three-fold: to demonstrate that AAIsafeR2 mode (i) reduces the percentage of ventricular pacing in patients with a spontaneous AV conduction and (ii) prevents atrial arrhythmias by improving atrial hemodynamics and (iii) improves the long-term hemodynamic status of implanted patients in comparison to the "standard" DDD mode, set with long AV Delays (DDD Long AVD).
The scientific soundness lying in this study is a proposal of an optimal programming of dual-chamber pacemakers (PM) with regard to the preservation of spontaneous AV conduction.
详细描述
In patients with sinus node dysfunction (SND), the use of AAI mode is rare. In paroxysmal high-degree or complete atrio-ventricular (AV) block, paroxysmal vagally-mediated bradycardia, single chamber atrial pacemakers are contraindicated although ventricular support is only rarely needed. As a result, these patients are usually exposed to an unnecessary high proportion of ventricular pacing due to the programming of the system to DDD mode for preventive safety reasons. This drains the pulse generator battery, and more importantly has long-term adverse effects on ventricular function in a subset of vulnerable patients, leading to left ventricular (LV) asynchrony, mitral regurgitation and increased left atrium (LA) size. The deleterious effects of right ventricular (RV) pacing, which increased heart failure (HF) morbidity/mortality, atrial fibrillation (AF) and related adverse events have been highlighted in several large clinical studies. Limiting RV pacing in non-dependant patients is therefore desirable.
The Symphony DR 2550 cardiac pacemakers (SORIN Group, Montrouge, France) offer a new pacing mode that automatically switches from AAI(R) mode to DDD(R) or DDI(R) in event of severe atrioventricular conduction disorder, returning spontaneously to AAI(R) mode as soon as the spontaneous AV conduction has resumed. This particular mode is called the AAIsafeR2(R).
The objective of this study is three-fold: to demonstrate that AAIsafeR2 mode (i) reduces the percentage of ventricular pacing in patients with a spontaneous AV conduction and (ii) prevents atrial arrhythmias by improving atrial hemodynamics and (iii) improves the long-term hemodynamic status of implanted patients in comparison to the "standard" DDD mode, set with long AV Delays (DDD Long AVD).
The scientific soundness lying in this study is a proposal of an optimal programming of dual-chamber pacemakers (PM) with regard to the preservation of spontaneous AV conduction.
2 Study devices
研究设计
- 研究类型
- Interventional
- 分配方式
- Randomized
- 干预模型
- Parallel
- 主要目的
- Treatment
- 盲法
- Single (Participant)
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Inclusion criteria / Indications Any patient who fulfils one or more of the official guidelines to be implanted with a dual-chamber pacemaker (primo-implantation) may be included in the study.
排除标准
- •Exclusion criteria / Contraindications
- •The patients presenting with one or more of the following characteristics cannot be included:
- •Permanent complete AV block
- •Permanent atrial and/or ventricular arrhythmias
- •already implanted with a cardioverter-defibrillator (ICD)
- •Likely to have a cardiac surgery in the next six months, mainly for:
- •severe coronary artery disease
- •severe valvular disease
- •Refuses to sign an consent form after having received the appropriate information
- •Refuses to co-operate
- •Not able to understand the study objectives and protocol
- •Not available for scheduled follow-up
- •With a life expectancy less than one year
- •Already included into another clinical study competing with the objectives of the CAN-SAVE R study.
- •<18 years old.
- •Contraindications Any patient to whom a contraindication from device labeling applies shall not be included in the study.
结局指标
主要结局
Percentage of ventricular pacing
Atrial Fibrillation/Atrial Arrhythmia Evolution
Cardiac Function Evolution
次要结局
- Evolution of conduction Disturbances
- Safety Analysis
- Congestive Hear Failure and Atrial Fibrillation Clinical Events