Evaluating Forced Oscillation Technique (FOT) in Abolishing Flow Limitation in Patients With Stable Chronic Pulmonary Disease (COPD)
Trial Snapshot
- Phase
- Not Applicable
- Status
- Completed
- Sponsor
- Philips Respironics
- Enrollment
- 22
- Locations
- 2
- Primary Endpoint
- Detection of Expiratory Flow Limitation by Mead and Wittenberger (M-W) Technique
Study Overview
Brief Summary
To validate the capability of the Respironics Synchrony ventilator to properly detect Expiratory Flow Limitation (EFL) as compared to the gold standard method of detecting EFL via the Mead and Wittenberger technique.
Detailed Description
Positive end expiratory pressure (PEEP) is used in COPD patients to counteract the intrinsic PEEP (PEEPi), which represents the end expiratory recoil pressure of the total respiratory system due to the presence of dynamic hyperinflation (DH).
DH commonly occurs in COPD, where the presence of expiratory flow-limitation (EFL) requires the patient to breath at higher lung volumes to produce the necessary after appropriate leak correction, showed a sensitivity and specificity in detecting EFL expiratory flow. To be effective, the PEEP level applied to the patient should be equal to PEEPi.
The continuous monitoring of EFL could be a useful tool to select the minimum PEEP level required to abolish it.
EFL can be detected using the forced oscillation technique (FOT) by an index which quantifies, for each breath, the within-breath variations of respiratory reactance (delta Xrs) at 5Hz.
Study Design
- Study Type
- Interventional
- Allocation
- Na
- Intervention Model
- Single Group
- Primary Purpose
- Treatment
- Masking
- None
Eligibility Criteria
- Ages
- — to 80 Years (Child, Adult, Older Adult)
- Sex
- All
- Accepts Healthy Volunteers
- No
Inclusion Criteria
- Not provided
Exclusion Criteria
- Not provided
Arms & Interventions
Respironics Synchrony ventilator (Non Invasive Ventilation)
Non Invasive Ventilation using forced oscillation technique (FOT)
Intervention: Respironics Synchrony ventilator (Non Invasive Ventilation) (Device)
Outcomes
Primary Outcomes
Detection of Expiratory Flow Limitation by Mead and Wittenberger (M-W) Technique
Time Frame: 2 hours
Number of breaths are cumulative across all participants were found to be flow limited, not-flow limited or indeterminate as determined by the Mead and Wittenberger (M-W) technique of esophageal pressure. Each participant's individual breaths was imported into Matlab software program that displayed the flow vs transpleural pressure and breaths were individually analyzed according to the Mead Wittenberger technique.
Detection of Expiratory Flow Limitation by a Commercial Mechanical Ventilator
Time Frame: 2 hours
Number of participants found to have expiratory flow limitation as determined by a commercial mechanical ventilator.
Detection of Expiratory Flow Limitation
Time Frame: within 2 hours
Expiratory flow limitation (EFL) is an increase in transpulmonary pressure (cmH2O) with no change expiratory flow (lpm). Expiratory Flow Limitation or absence thereof will be detected by using a measurement of changes in reactance (DeltaXrs - cmH2O\*s/L) it will be compared to the the Mead and Whittenberger technique (via esophageal balloon measuring transpulmonary pressures). Two measurements will be obtained, DeltaXrs and Transpulmonary pressure. Comparisons of these measurements will be made to determine if the participant exhibits EFL.
Secondary Outcomes
No secondary outcomes reported