Evaluation of Commonly Used Blood Glucose Meters in Vietnam and Effect of Hematocrit on Device's Accuracy
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 135
- 试验地点
- 1
- 主要终点
- Hematocrit
研究概览
简要总结
Rationale: Most BGM (blood glucose monitoring) devices in Vietnam are manufactured for home use only. However, they are commonly used in healthcare facility, primarily at point of care. We want to assess the validity of these devices by using CLSI (Clinical and Laboratory Standards Institute) POCT12-A3 to assess the accuracy.
Objective: To validate the accuracy and effect of hematocrit on these devices, along with the precision profile using CLSI EP15A3
Study design: This is a cross-sectional study.
Study population: Because we need a wide range of blood glucose level. We chose pregnant women who underwent 75g glucose tolerance test.
详细描述
Most blood glucose monitors circulating the market are manufactured after ISO 15197:2013 standard, which is intended for self-measurement. Medical organizations need to verify blood glucose monitors if they are designed for prescription point-of-care use or validate blood glucose monitors if they are designed for over-the-counter use.
Pregnant women are selected after consent, their blood are then drawn by phlebotomist into an EDTA (ethylenediamine tetraacetic acid) with NaF (sodium flouride) blood tube. Samples are run on alinity h hematology analyzer, manufactured by Abbott, to measure hematocrit. The analyzers are verified by laboratory procedure, recommended by CLSI (Clinical and Laboratory Standards Institute) method and quality control is checked before in accordance to the laboratory procedure. After that, blood tube is mixed thoroughly before being pipetted. Each BGM device will measure blood glucose level of the pipetted whole blood. Finally, samples will be run on a cobas c 502 analyzer to measure plasma glucose level, which is a reference method for determining the accuracy of the BGM devices.
Statistical analysis
Meter performance is acceptable for use in hospitals when 95% of the individual results from the POC glucose meter system agree within 0.67 mmol/L (12 mg/dL) of the laboratory analyzer values at glucose concentrations below 5.55 mmol/L (100 mg/dL) and within 12.5% of the laboratory analyzer values at glucose concentrations at or above 5.55 mmol/L (100 mg/dL). In addition, the sum of the number of individual results with 1) errors that exceed 0.83 mmol/L (15 mg/dL) at glucose concentrations below 4.2 mmol/L (75 mg/dL) and 2) errors that exceed 20% at glucose concentrations at or above 4.2 mmol/L (75 mg/dL) should not exceed 2% of all results. An ordinary linear regression is performed to see how well the glucose level of BGM devices correlates with the reference method. Bland-Altman difference plots are used to describe the difference along with the hematocrit of each sample.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Only
- 时间视角
- Cross Sectional
入排标准
- 年龄范围
- 18 Years 至 45 Years(Adult)
- 性别
- Female
- 接受健康志愿者
- 是
入选标准
- •Pregnant women undergo 75g glucose tolerance test
排除标准
- •Usage of anticoagulant drug
- •History of bleeding disorders.
结局指标
主要结局
Hematocrit
时间窗: Each participant's blood samples are run immediately after phlebotomy
Hematocrit in percentage, measured on hematology analyzer alinity h
ONETOUCH VerioVue glucose concentration
时间窗: Each participant's blood samples are run immediately after phlebotomy
Glucose concentration, measured on assessed blood glucose monitoring device ONETOUCH VerioVue, Serial number YEPMT06H
Contour Plus glucose concentration
时间窗: Each participant's blood samples are run immediately after phlebotomy
Glucose concentration, measured on assessed blood glucose monitoring device Contour Plus, Serial number 6128289
ACCU-CHEK Inform II glucose concentration
时间窗: Each participant's blood samples are run immediately after phlebotomy
Glucose concentration, measured on assessed blood glucose monitoring device ACCU-CHEK Inform II, Serial number UU14463019
EasyGlucose glucose concentration
时间窗: Each participant's blood samples are run immediately after phlebotomy
Glucose concentration, measured on assessed blood glucose monitoring device EasyGlucose, Serial number G16D21I0700816
Reference Glucose concentration
时间窗: Each participant's blood samples are run immediately after phlebotomy
Glucose concentration, measured on reference glucose analyzer cobas c 502
VivaChek Ino glucose concentration
时间窗: Each participant's blood samples are run immediately after phlebotomy
Glucose concentration, measured on assessed blood glucose monitoring device VivaChek Ino, Serial number 311Z0021DFA
ACCU-CHEK performa glucose concentration
时间窗: Each participant's blood samples are run immediately after phlebotomy
Glucose concentration, measured on assessed blood glucose monitoring device ACCU-CHEK performa, Serial number 68827245972
次要结局
未报告次要终点
研究者
Huan Nguyen Pham
Technical supervisor
Hung Vuong Hospital
