MedPath

Cataract Lens Hardness Based on Phaco Tip Resistance

Conditions
Cataract Senile
Registration Number
NCT05110222
Lead Sponsor
Ospedale Policlinico San Martino
Brief Summary

Phacoemulsificaton tip elongation depends upon voltage once resonant frequency of piezoelectric crystals has been reached.

Purpose of the study is to record voltage data of 20 consecutive cataract surgeries in order to evaluate correlation between voltage and LOCS classification of lens nuclei using a system that deploys a feedback mechanism capable of keeping invariant elongation regardless of enountered resistence

Detailed Description

The R-Evolution (Optikon 2000 Rome, Italy) phaco machine is equipped with a patented system named "minimal stress" that used the direct piezoelectric effect from the resonant ceramics that drive the phacoemulsificator tip to assess in real time and adjust phaco tip elongation that continuously varies according to encountered resistance.

After priming in balanced salt solution, phaco tip elongation and the amount of potential difference (voltage) needed to obtain a ginev elongation in balanced solution is assessed. Once the phaco tip is used during surgery, the resistence offered by the crystalline lens material to the phaco tip reduces actual elongaton which in turn reduces the voltage produced by the directo piezometric effect. The machine uses voltage produced by actual elongation to adjust the diving voltage and therefore can ensure accurate pahco tip elongation under any condition of lens material resistance.

Purpose of the study is to record driving voltage and its variations and evaluate if there is a correlation between phaco tip resistence and drving voltage.

This is evaluated both in vitro activating the phaco tip in increasingly dense glycerol solutions and in real life phaco surgery recording the voltage requested by phaco tip resistence and correlating it to the classification of lens opacity (LOCS III system) clinically used to grade cataract hardness.

Recruitment & Eligibility

Status
UNKNOWN
Sex
All
Target Recruitment
20
Inclusion Criteria

diagnosis of nuclear or cortical cataract

Exclusion Criteria

previous ocular trauma, previous ocular surgery, connective tissue disease

Study & Design

Study Type
OBSERVATIONAL
Study Design
Not specified
Primary Outcome Measures
NameTimeMethod
cataract lens hardness according to LOCS III classification systemat enrolling

lens opacity and estimated nucleous hardness according to international LOCS classification

peak voltageat the time of surgery

maximum voltage necessary to emulsify cataract nucleous

average voltageat the time of surgery

voltage necessary to emulsify cataratct nuclous

Secondary Outcome Measures
NameTimeMethod

Trial Locations

Locations (1)

IRCCS Policlinico San Martino

🇮🇹

Genoa, Italy

© Copyright 2025. All Rights Reserved by MedPath