临床试验/NCT05482035
NCT05482035已完成不适用

Clinical Trial on the Effectiveness of Blood Pressure Function of Wrist Blood Pressure Monitor

Huawei Device Co., Ltd1 个研究点 分布在 1 个国家目标入组 88 人开始时间: 2022年2月16日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
88
试验地点
1
主要终点
mean difference of systolic blood pressure

研究概览

简要总结

The accuracy and safety of wrist blood pressure (model: MLY-B10) is evaluated according to the requirement of EN ISO 81060-2:2019/A1:2020.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Subjects should voluntarily participate in the clinical trial and sign the informed consent form.
  • Subjects must be willing and able to follow the investigation procedures.
  • Subjects are over 18 years old whose wrist circumference should be within 13.0-20.0 cm.

排除标准

  • Subjects with serious arrhythmias (atrial fibrillation, atrial flutter, ventricular flutter, ventricular fibrillation, ventricular tachycardia, borderline tachycardia, sinus arrest, pathological sinus syndrome, pacemaker rhythm) showed by electrocardiogram during screening period.
  • Subject with prior allergy to nylon, fluororubber, thermoplastic polyurethane elastomer (TPU), or polyvinyl chloride (PVC) materials.
  • Subjects with prior arterial obstructive disease or arteritis.
  • Subjects who had previously undergone mastectomy.
  • Subjects with wound or wrist inflammation were measured.
  • Subjects who are receiving intravenous infusion, blood transfusion, or other catheters at the site of measurement.
  • Subjects who had eaten, drank, smoked or exercised vigorously in the 30 minutes prior to the study.
  • Subjects who have participated in other clinical trials within 30 days that may affect this trial.
  • Subjects who the investigator consider inappropriate to attend the trial.

结局指标

主要结局

mean difference of systolic blood pressure

时间窗: 30 minutes

mean difference of systolic blood pressure between wrist blood pressure monitor and mercury sphygmomanometer

standard deviation of the difference of diastolic blood pressure

时间窗: 30 minutes

standard deviation of the difference of diastolic blood pressure between wrist blood pressure monitor and mercury sphygmomanometer

mean difference of diastolic blood pressure

时间窗: 30 minutes

mean difference of diastolic blood pressure between wrist blood pressure monitor and mercury sphygmomanometer

standard deviation of the difference of systolic blood pressure

时间窗: 30 minutes

standard deviation of the difference of systolic blood pressure between wrist blood pressure monitor and mercury sphygmomanometer