Wearable Wireless Respiratory Monitoring System That Detects and Predicts Opioid Induced Respiratory Depression
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 14
- 试验地点
- 1
- 主要终点
- Accuracy of tidal volume (TV) measurement in a noisy environment when RMS adaptive filtering and active noise cancellation is turned on versus turned off.
研究概览
简要总结
An observational study will be conducted in approximately 14 participants to evaluate the ability of a wearable, wireless acoustic Respiratory Monitoring System (RMS) to accurately measure a participant's respiratory rate, tidal volume, minute ventilation, and duration of apnea in a noisy environment. Sensor accuracy will be measured with adaptive filtering and active noise cancellation turned on versus turned off.
详细描述
The Respiratory Monitoring System (RMS) consists of a miniature acoustic sensor and a soft flexible cradle that is adhered to the skin of the neck over the proximal trachea (within the sternal notch) with medical grade adhesive. The sensor body consists of a miniature bell stethoscope head, electronics, a microphone that faces the trachea and a microphone that faces the external environment, a Bluetooth low energy transmitter/receiver, an antenna, and a rechargeable battery. The sensor is secured by the cradle at the optimal location to measures the sounds of airflow in the proximal trachea during inhalation and exhalation.
Proprietary machine learning/AI algorithms convert the sounds of airflow into the measurements of respiratory rate (RR), tidal volume (TV), minute ventilation (MV), and duration of apnea. Sensor information is transmitted to a bedside PC that displays the vital sign data in real-time. The wearable, wireless RMS is being developed for hospital and outpatient use as a tool to detect and predict respiratory compromise/clinical deterioration in a more-timely and accurately manor (fewer false alerts/alarms) than current methods.
The breathing data from 14 to 20 participants will be recorded during one study session lasting approximately 90 minutes with the sensor/cradle adhered to the neck over the proximal trachea. Reference breathing data will be recorded simultaneously using a hospital ventilator's pneumotach and capnometer attached to a tight-fitting face mask.
Each subject will be instructed to breath the following protocol 3 or 4 times:
Record RMS data and pneumotach/capnometer data for ~400 seconds with the study subject breathing a normal RR and TV.
研究设计
- 研究类型
- Observational
- 观察模型
- Cohort
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 70 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Age 18 to 70 years.
- •Subject understands the English language, understands the risks, benefits, and alternatives to this research study, and is willing and able to give written informed consent.
排除标准
- •Age <18 years>
- •BMI < 20 or >
- •Does not understand written and spoken English.
- •Anxiety or claustrophobia related to wearing a face mask.
- •History of skin irritation or inflammation related to the adhesive, adhesive tape, or materials used in the trachea sound sensor or facemask.
- •Active infection or inflammation of the skin above the proximal trachea.
- •Excessive facial hair that may prevent a tight seal around the facemask.
- •Unstable cardiac, vascular, pulmonary, hepatic, renal, immune function at the discretion of the investigator.
- •Pregnancy or breast feeding.
- •Current participation in an industry sponsored pharmaceutical study or a medical device study.
结局指标
主要结局
Accuracy of tidal volume (TV) measurement in a noisy environment when RMS adaptive filtering and active noise cancellation is turned on versus turned off.
时间窗: 90 minutes
RMS breathing data and reference pneumotach/capnogram data will be recorded with RMS adaptive filtering and active noise cancellation turned on and turned off to calculate the accuracy of TV measurement.
Accuracy of respiratory rate (RR) measurement in a noisy environment when RMS adaptive filtering and active noise cancellation is turned on versus turned off.
时间窗: 90 minutes
RMS breathing data and reference pneumotach/capnogram data will be recorded with RMS adaptive filtering and active noise cancellation turned on and turned off to calculate the accuracy of RR measurement.
Accuracy of duration of apnea measurement in a noisy environment when RMS adaptive filtering and active noise cancellation is turned on versus turned off.
时间窗: 90 minutes
RMS breathing data and reference pneumotach/capnogram data will be recorded with RMS adaptive filtering and active noise cancellation turned on and turned off to calculate the accuracy of duration of apnea measurement.
Accuracy of minute ventilation (MV) measurement in a noisy environment when RMS adaptive filtering and active noise cancellation is turned on versus turned off.
时间窗: 90 minutes
RMS breathing data and reference pneumotach/capnogram data will be recorded with RMS adaptive filtering and active noise cancellation turned on and turned off to calculate the accuracy of MV measurement.
次要结局
- Measure the signal-to-noise ratio of the RMS output signal in a noisy external environment with adaptive filtering and active noise cancellation turned on and off.(90 minutes)
研究者
Jeffrey Joseph
Professor of Anesthesiology
Thomas Jefferson University