NC Thermometer (Gen 3) ISO 80601-2-56 Validation
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 入组人数
- 105
- 试验地点
- 2
- 主要终点
- Clinical Bias
研究概览
简要总结
The purpose of this clinical validation is to evaluate the the clinical bias, the limits of agreement, and the clinical repeatability of measuring oral temperature using a new thermometer, NC° Thermometer (Gen 3), compared to a reference clinical thermometer.
详细描述
The purpose of this clinical study is to validate the clinical accuracy of the NC° Thermometer (Gen 3) according to the requirements of ISO 80601-2-56:2017. The NC° Thermometer (Gen 3) (the device under study (DUT)) is a predictive thermometer that operates in adjusted mode. The Welch Allyn oral thermometer (Welch Allyn Spot Vital Signs 420 SureTemp temperature module, Welch Allyn, San Diego, CA) with a measurement accuracy of ± 0.1 °C in monitor mode will be used the reference clinical thermometer (RCT). During the clinical study, subjects will be asked to perform three consecutive oral measurements with the DUT, and one oral measurement with the RCT.
研究设计
- 研究类型
- Observational
- 观察模型
- Other
- 时间视角
- Other
入排标准
- 年龄范围
- 5 Years 至 —(Child, Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •They are more than 5 years old
- •They have not taken any antipyretic in the preceding 120 minutes
- •They have no medical conditions which might impact the clinical accuracy validation results, such as inflammation at the measuring site (the mouth)
- •They are not taking medications such as barbiturates, thyroid preparations, antipsychotics, and recent immunizations
排除标准
- 未提供
结局指标
主要结局
Clinical Bias
时间窗: Up to 1 hour
Clinical bias is the mean difference between the output temperatures of the NC° Thermometer (Gen 3) (DUT) and the reference clinical thermometer (RCT) for a specific reference body site (the mouth) when measured from a selected group of subjects. The clinical bias of the DUT defines closeness between the DUT output temperature and that of the RCT.
Clinical repeatability
时间窗: Up to 1 hour
Clinical repeatability (sometimes called the perceived accuracy) is a measure of the consistency of repeated measurements with identical operating conditions when temperatures are taken within short time intervals from the same measuring site of the same subject by the same operator with the same intermittent clinical thermometer. Clinical repeatability is a pooled standard deviation of triplicate measurements over the population of subjects.
Limits of agreement
时间窗: Up to 1 hour
Dispersion of output temperatures around the clinical bias can be estimated by the standard deviation of the temperature differences between the DUT and the RCT as measured from multiple subjects.