临床试验/NCT05252702
NCT05252702已完成不适用

Aveir Dual-Chamber Leadless i2i IDE Study

Abbott Medical Devices145 个研究点 分布在 1 个国家目标入组 464 人开始时间: 2022年2月2日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
464
试验地点
145
主要终点
Percentage of Participants Free From Aveir DR System-Related Complications at 3-months

研究概览

简要总结

Prospective, non-randomized, multi-center, international study designed to evaluate the safety and effectiveness of the Aveir™ Dual-Chamber (DR) Leadless Pacemaker system.

详细描述

The purpose of this clinical investigation is to evaluate the clinical safety and effectiveness of the Aveir DR Leadless Pacemaker system in a patient population indicated for a DDD(R) pacemaker.

Subjects participating in the study are followed through at least 12 months with data collected at baseline, implant procedure, pre (hospital) discharge, and follow-up at 1 month, 3 months, 6 months, 12 months, and every 6 months thereafter until study completion

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Treatment
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Subject must have at least one of the clinical indications before device implant in adherence with ACC/AHA/HRS/ESC dual chamber pacing guidelines
  • Subject is ≥ 18 years of age or age of legal consent, whichever age is greater
  • Subject has a life expectancy of at least one year
  • Subject is willing to comply with clinical investigation procedures and agrees to return to clinic for all required follow-up visits, tests, and exams
  • Subject has been informed of the nature of the clinical investigation, agrees to its provisions and has provided a signed written informed consent, approved by the IRB/EC

排除标准

  • Subject is currently participating in another clinical investigation that may confound the results of this study as determined by the Sponsor
  • Subject is pregnant or nursing and those who plan pregnancy during the clinical investigation follow-up period
  • Subject has presence of anatomic or comorbid conditions, or other medical, social, or psychological conditions that, in the investigator's opinion, could confound the assessment of the investigational device and/or implant procedure, limit the subject's ability to participate in the clinical investigation or to comply with follow-up requirements of the clinical investigation results
  • Subject has a known allergy or hypersensitivity to < 1 mg of dexamethasone sodium phosphate or any blood or tissue contacting material listed in the IFU
  • Subject has an implanted vena cava filter or mechanical tricuspid valve prosthesis
  • Subject has pre-existing, permanent endocardial pacing or defibrillation leads (does not include lead fragments)
  • Subject has current implantation of either conventional or subcutaneous implantable cardioverter defibrillator (ICD) or cardiac resynchronization therapy (CRT) device
  • Subject has an implanted leadless cardiac pacemaker (except for an Aveir ventricular leadless pacemaker)
  • Subject is implanted with an electrically-active implantable medical device with stimulation capabilities (such as neurological or cardiac stimulators)
  • Subject is unable to read or write

研究组 & 干预措施

Single Arm

Other

Non randomized arm

干预措施: Aveir DR Leadless Pacemaker System (Device)

结局指标

主要结局

Percentage of Participants Free From Aveir DR System-Related Complications at 3-months

时间窗: 3 months

Aveir DR Leadless Pacemaker system complication-free-rate in de novo subjects. A complication was defined as a device-or-procedure-related serious adverse event. Complications included Atrial LP, Ventricular LP, and implant procedure-related complications. Complications possibly related or related to COVID-19 were excluded.

Percentage of Participants Free From Aveir DR System-Related Complications at 12-months

时间窗: 12 months

Aveir DR Leadless Pacemaker system complication-free-rate in de novo subjects. A complication was defined as a device-or-procedure-related serious adverse event. Complications included Atrial LP, Ventricular LP, and implant procedure-related complications. Complications possibly related or related to COVID-19 were excluded.

Percentage of Subjects With Acceptable Atrial Threshold and Amplitude Measurements at 3-months

时间窗: 3 months

Composite success rate of acceptable atrial pacing thresholds and P-wave amplitudes in de novo subjects. A subject was considered to have met the primary effectiveness endpoint #1 if the pacing threshold voltage was ≤ 3.0 V at 0.4 ms at the 3-month visit and the sensed P-wave amplitude was ≥ 1.0 mV at the 3-month visit.

Percentage of Subjects With Acceptable Atrial Threshold and Amplitude Measurements at 12-months

时间窗: 12 months

Composite success rate of acceptable atrial pacing thresholds and P-wave amplitudes in de novo subjects. A subject was considered to have met the success criteria for the endpoint if the pacing threshold voltage was ≤ 3.0 V at 0.4 ms at the 12-month visit and the sensed P-wave amplitude was ≥ 1.0 mV at the 12-month visit.

Percentage of Participants With AV Synchrony Success Rate at Rest While Seated

时间窗: 3 months

AV synchrony success was defined as subjects with a paced or sensed ventricular beat within 300 ms following a paced or sensed atrial beat (i.e., a synchronous cycle) for at least 70% of evaluable cardiac cycles.

次要结局

  • Percentage of Participants Free From Aveir Atrial Leadless Pacemaker-Related Complications at 12-months(12 months)
  • Slope of the Normalized Increase in the Sensor Indicated Rate Versus Normalized Workload of the Aveir Atrial Leadless Pacemaker During Exercise Testing(3 months)
  • Percentage of Participants Free From Aveir Atrial Leadless Pacemaker-Related Complications at 3-months(3 months)

研究者

申办方类型
Industry
责任方
Sponsor