User Performance of the ONYX PLUS Blood Glucose Monitoring System

Ascensia Diabetes Care4 个研究点 分布在 1 个国家目标入组 134 人开始时间: 2015年2月1日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
134
试验地点
4
主要终点
Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method

研究概览

简要总结

The purpose of this study was to determine if untrained subjects with diabetes can operate the Investigational ONYX PLUS Blood Glucose Monitoring System (BGMS) and obtain valid glucose results.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Males and Females, 18 years of age and older
  • Able to speak, read, and understand English
  • Willing to complete all study procedures

排除标准

  • Hemophilia or other bleeding disorder
  • Pregnancy
  • Physical, visual, or neurological impairments that would make the person unable to perform testing with the BGM
  • Previously participated in a BG monitor study using the ONYX PLUS BGMS
  • Working for a medical laboratory, hospital, or other clinical setting that involves training on or clinical use of blood glucose monitors
  • Working for a competitive medical device company, or having an immediate family member who works for such a company
  • A condition which, in the opinion of the investigator or designee, would put the person or study conduct at risk
  • The enrollment goal for the intended use population:
  • At least 70% of subjects will be younger than age 65
  • At least 20% of subjects will have type 1 diabetes

研究组 & 干预措施

Users of the Monitoring System

Experimental

Untrained subjects with diabetes used the ONYX PLUS Investigational Blood Glucose Monitoring System.

干预措施: ONYX PLUS Investigational Blood Glucose Monitoring System (Device)

结局指标

主要结局

Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method

时间窗: 1 hour

Untrained subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) and +/- 15% (\>= 100mg/dL YSI capillary plasma).

次要结局

  • Percent of Venous Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method(1 hour)
  • Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 12.5mg/dL (<100mg/dL) and Within +/- 12.5% (>=100mg/dL) of Laboratory Glucose Method(1 hour)
  • Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS(1 hour)
  • Percent of Alternate Site Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method(1 hour)
  • Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100mg/dL) of Laboratory Glucose Method When Tested by Study Staff(1 hour)
  • Percent of Responses From Persons With Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose(1 hour)

研究者

发起方
Ascensia Diabetes Care
申办方类型
Industry
责任方
Sponsor