跳至主要内容
临床试验/NCT02073851
NCT02073851已完成不适用

A Prospective, Single Arm Feasibility Pilot Study to Evaluate the Safety and Performance of the TriGuard™HDH Embolic Deflection Device in Patients Undergoing Transcatheter Aortic Valve Replacement (TAVR)

Keystone Heart1 个研究点 分布在 1 个国家目标入组 12 人开始时间: 2014年1月1日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
12
试验地点
1
主要终点
Device performance

研究概览

简要总结

Pilot study enrolling up to 12 patients at a single investigational site in the Netherlands. Patients for TAVR will be enrolled to receive the Embolic Deflection Device throughout the duration of the TAVR procedure.

详细描述

Prospective, single center, single arm pilot study enrolling up to 12 patients at a single investigational site in the netherlands. Patients meeting eligibility criteria for TAVR and none of the exclusion criteria will be enrolled to receive the Embolic Deflection Device throughout the duration of the TAVR procedure.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Treatment
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • The patient must be ≥ 18 years of age.
  • The patient meets indications for transcatheter aortic valve replacement procedure.
  • The patient is willing to comply with specified follow-up evaluations.
  • The patient, or legally authorized representative, has been informed of the nature of the study, agrees to its provisions and has been provided written informed consent, approved by the appropriate Medical Ethics Committee (EC).

排除标准

  • Patients undergoing transcatheter aortic valve replacement via the trans-axillary, subclavian, or direct aortic route
  • Pregnant or nursing subjects and those who plan pregnancy in the period up to 1 year following index procedure. Female subjects of child-bearing potential must have a negative pregnancy test done within 7 days prior to index procedure per site standard test.
  • Patients with known diagnosis of acute myocardial infarction (AMI) within 72 hours preceding the index procedure (according to definition) or AMI > 72 hours preceding the index procedure and creatine kinase and creatine kinase fraction have not returned to normal limits at the time of procedure.
  • Patients who are currently experiencing clinical symptoms consistent with new onset AMI, such as nitrate-unresponsive prolonged chest pain.
  • Patients with impaired renal function (estimated Glomerular Filtration Rate [Estimated Glomerular Filtration Rate] <30, calculated from serum creatinine by the Cockcroft-Gault formula).
  • Patients with torturous/unsuitable anatomy as related to major cerebral arteries in the aortic arch that may interfere with device deployment or remain deployed.
  • Patients with a platelet count <100.000 cells/mm³ or >700.000 cells/mm³ or a white blood cell <3000 cells/mm³ within 7 days prior to index procedure.
  • Patients with a history of bleeding diathesis or coagulopathy or patients in whom anti-platelet and/or anticoagulant therapy is contraindicated, or will refuse transfusion.
  • Patients who have received any organ transplant or are on a waiting list for any organ transplant.
  • Poor fluoroscopic visualization due to obesity or other medical reason
  • Hypotension requiring iv/ia medication or other therapy such as resuscitation and defibrillation
  • Patients with known other medical illness or known history of substance abuse that may cause non-compliance with the protocol, confound the data interpretation or is associated with a life expectancy of less than one year.
  • Patients with a known hypersensitivity or contraindication to aspirin, heparin/bivalirudin, clopidogrel/ticlopidine, nitinol, stainless steel alloy, latex, and/or contrast sensitivity that cannot be adequately pre-medicated.
  • Patients with a history of a stroke or transient ischemic attack (TIA) within the prior 6 months.
  • Patients with an active peptic ulcer or upper gastrointestinal (GI) bleeding within the prior 6 months.
  • Patients presenting with cardiogenic shock.
  • Patients with severe peripheral arterial disease that precludes the delivery sheath vascular access.
  • Patients with severe calcification/atheroma, friable or mobile atherosclerotic plaque in the aortic arch
  • Patients with contraindication to cerebral MRI.
  • Patients going through unprotected (cerebral embolization) cardiovascular procedure prior, during or post transcatheter aortic valve replacement procedure, before the post-procedure diffusion-weighted magnetic resonance imaging evaluation
  • Patients who have a planned treatment with any other investigational device or procedure during the study period.

结局指标

主要结局

Device performance

时间窗: During TAVR procedure

Device performance will be evaluated for the following functions: Access the aortic arch, Deploy the TriGuard™HDH from the delivery catheter, Position the TriGuard™HDH to cover the cerebral/carotid vessels, Retrieve the TriGuard™HDH system intact

次要结局

  • Number of new cerebral lesions(one week post procedure)
  • Volume of new cerebral lesions(One week post procedure)

研究者

发起方
Keystone Heart
申办方类型
Industry
责任方
Sponsor

研究点 (1)

Loading locations...

相似试验

DEFLECT II: A Study to Evaluate the Safety and... | 临床试验