跳至主要内容
临床试验/NCT06191393
NCT06191393已完成不适用

Clinical Validation of the Aptitude Medical Systems Metrix COVID/Flu Test for Detection of SARS-CoV-2 and Influenza A/B in Point-of-Care and Non-Laboratory Settings

Aptitude Medical Systems5 个研究点 分布在 1 个国家目标入组 614 人开始时间: 2023年11月20日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
614
试验地点
5
主要终点
Qualitative Detection of RNA From SARS-CoV-2, Influenza A, and Influenza B ; Positive Percent Agreement Performance Comparison of Candidate Device (Metrix) vs FDA-Cleared Comparator Assay

研究概览

简要总结

The Metrix COVID/Flu Test will be evaluated for use in Point-of-Care and Non-Laboratory settings in a home testing environment utilizing the clinical study design described herein. The study will take place in simulated home environments which will be set up within or near active clinical settings (e.g., urgent care facilities). This will be a prospective study conducted at three or more investigational sites located within the United States for the clinical validation of the Metrix COVID/Flu Test for the differential detection of SARS-CoV-2, Influenza A, and Influenza B in AN swab or saliva samples. Additional sites may be added to the study in order to meet minimum subject/sample enrollment requirements and geographic prevalence of respiratory virus infections. One or more investigational sites outside of the US (at which participants are required to be fluent in English) may also be included in the study depending on enrollment needs and geographical/seasonal prevalence of respiratory virus infections. Comparator testing will be performed to determine the infection status of each sample for comparison to results generated by the candidate test. The primary comparator for the study will be an FDA-cleared assay for the detection of SARS-CoV-2, Flu A, and/or Flu B.

详细描述

The Metrix COVID/Flu Test will be evaluated for OTC use in a home use testing environment utilizing the clinical study design described herein. The study will take place in simulated home environments which will be set up within or near active clinical settings (e.g., urgent care facilities). This will be a prospective study conducted at three or more investigational sites located within the United States for the clinical validation of the Metrix COVID/Flu Test for the differential detection of SARS-CoV-2, Influenza A, and Influenza B in AN swab or saliva samples. Additional sites may be added to the study in order to meet minimum subject/sample enrollment requirements and geographic prevalence of respiratory virus infections. One or more investigational sites outside of the US (at which participants are required to be fluent in English) may also be included in the study depending on enrollment needs and geographical/seasonal prevalence of respiratory virus infections. Comparator testing will be performed to determine the infection status of each sample for comparison to results generated by the candidate test. The primary comparator for the study will be an FDA-cleared assay for detection of SARS-CoV-2, Flu A, and Flu B.

The candidate device, Metrix COVID/Flu Test, will be evaluated with two specimen types - saliva and AN swab. Each Metrix COVID/Flu Test kit contains all supplies needed to test either a saliva or AN swab sample, including a corresponding QRI for each specimen type. It is possible for participants to elect to provide only a single sample type for the study; however, both sample types should be collected when possible. If participants will be testing both saliva and AN swab samples, the AN swab test should be conducted first. If participants elect to provide only a single sample type, the AN swab will be prioritized.

This OTC study will take place in simulated home environments which will be set up within or near clinical settings (e.g., urgent care facilities). The study will enroll symptomatic subjects only. Candidate samples will be self-collected by participants (or collected by a guardian for participants under the age of 14 years) and comparator samples will be collected by a healthcare practitioner with informed consent and Institutional Review Board (IRB) approval. Each sample will be coded for confidentiality.

For each participant, a comparator NP swab sample will be obtained by a healthcare practitioner after testing with the candidate device. This NP swab will be tested with the FDA-cleared SARS-CoV-2, Flu A, and Flu B assay according to the IFU and will serve as the primary comparator for both saliva and AN swab samples. An additional saliva sample will also be obtained for SARS-CoV-2 analysis via a secondary comparator method, which will be used for informational purposes only. An additional AN swab sample will be collected and stored frozen and dry for later analysis to support future research validation studies.

No results from the investigational test or comparator testing will be used for clinical diagnosis, management of study participants, or to make treatment decisions.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

盲法说明

Some participants will know their SOC diagnosis before completing testing on our device.

入排标准

年龄范围
2 Years 至 —(Child, Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Participant or guardian understands and is able and willing to provide written informed consent, and assent where applicable, prior to study enrollment.
  • Male or female aged 2 years or older
  • Participant is currently exhibiting fever, or one or more symptoms associated with COVID-19 and/or influenza (such as, but not limited to, chills, cough, shortness of breath or difficulty breathing, fatigue, new loss of taste or smell, sore throat, congestion or runny nose, nausea or vomiting or diarrhea). Participant must still be exhibiting symptoms on the day of sample collection.
  • Participant has not eaten, consumed a beverage, smoked, brushed their teeth, gargled with mouth wash, or chewed gum for 30 minutes prior to collecting a saliva sample.
  • Participant or guardian agrees to read, and is able to read with understanding, each of the saliva and/or AN swab QRI prior to beginning the operation of each of the Metrix COVID/Flu Test.
  • Participant or guardian is able and willing to contribute the required saliva and/or swab samples for testing and understands and is able and willing to sign the study informed consent.

排除标准

  • Participant does not understand and/or is not able and willing to sign the study informed consent and/or assent.
  • Participant or guardian is not able to comply with saliva or nasal swab collection requirements following the QRI.
  • Participant has previously provided a sample for the study.
  • Participant has had seasonal influenza and/or the SARS-CoV-2 vaccine within the past 5 days.
  • Participant is not able to tolerate sample collection.
  • Participant is currently undergoing antiviral treatment such as baloxavir marboxil (trade name Xofluza®), oseltamivir (Tamiflu®), zanamivir (Relenza®), and peramivir (Rapivab®).
  • Participants currently undergoing treatment and/or within the past thirty (30) days with prescription medication to treat novel Coronavirus SARS-CoV-2 infection, which may include but is not limited to Remdesivir (Veklury), Paxlovid, molnupiravir or receiving convalescent plasma therapy for SARS-CoV-
  • Participants who have had a nasal wash or aspirate as part of their standard of care treatment on day of study visit prior to the study sample collection.
  • Participants who have had recent craniofacial injury or surgery, including to correct deviation of the nasal septum, within the previous six (6) months.
  • Participants who do not understand/read the English language.

结局指标

主要结局

Qualitative Detection of RNA From SARS-CoV-2, Influenza A, and Influenza B ; Positive Percent Agreement Performance Comparison of Candidate Device (Metrix) vs FDA-Cleared Comparator Assay

时间窗: 45 minutes

Qualitative detection of RNA from SARS-CoV-2, Influenza A, and Influenza B in unprocessed saliva and/or AN swab samples. The primary objective is to evaluate the performance (positive percent agreement, negative percent agreement, and 95% confidence intervals) of the Metrix COVID/Flu Test for qualitative, differential detection of SARS-CoV-2, Influenza A, and/or Influenza B RNA using AN swab and saliva samples. Comparator testing will determine the infection status of each sample for comparison to the candidate test. The primary comparator will be an FDA-cleared assay that detects SARS-CoV-2, Flu A, and Flu B. Positive Percent Agreement (PPA) is the number of positive specimens identified by the candidate test divided by the total true positives per the comparator.

Qualitative Detection of RNA From SARS-CoV-2, Influenza A, and Influenza B ; Negative Percent AgreementPerformance Comparison of Candidate Device (Metrix) vs FDA-Cleared Comparator Assay

时间窗: 45 minutes

Qualitative detection of RNA from SARS-CoV-2, Influenza A, and Influenza B in unprocessed saliva and/or AN swab samples. The primary objective is to evaluate the performance (positive percent agreement, negative percent agreement, and 95% confidence intervals) of the Metrix COVID/Flu Test for qualitative, differential detection of SARS-CoV-2, Influenza A, and/or Influenza B RNA using AN swab and saliva samples. Comparator testing will determine the infection status of each sample for comparison to the candidate test. The primary comparator will be an FDA-cleared assay that detects SARS-CoV-2, Flu A, and Flu B. Negative Percent Agreement (NPA) is the number of negative specimens identified by the candidate test divided by the total true negatives per the comparator.

次要结局

未报告次要终点

研究者

发起方
Aptitude Medical Systems
申办方类型
Industry
责任方
Sponsor

研究点 (5)

Loading locations...

相似试验