Refractix DSP and TS Meter-DSP Measurement Comparison

Not yet recruiting
Conditions
Healthy
Interventions
Device: Serum Total Protein Measurement with TS Meter-DSP and Refractix DSP
Registration Number
NCT05717036
Lead Sponsor
Reichert, Inc.
Brief Summary

The goal of this clinical trial is to compare the serum plasma measurement results between the FDA cleared TS Meter-DSP and investigational Refractix DSP. Study participants will be individuals being screened for plasma donation in a plasmaphereris center. The main question the study aims to answer is whether the measurement taken on the investigational Refractix DSP device is equivalent, defined as within +/- 0.3g/100ml, of the measurement taken on the TS Meter-DSP. Participants will be asked to provide a second capillary tube of blood (\<0.085ml) from the fingerstick that is part of their routine pre-screening for plasma donation.

Detailed Description

The TS Meter-DSP and investigational Refractix DSP are digital refractometers intended for the quantitative measurement of total protein in human serum or plasma. The devices are intended for use in plasma collection centers and are not for diagnostic purposes. Serum/plasma samples are taken from a successful microhematocrit assay and dispensed via pipette on to the instruments. The instruments measure the temperature compensated refractive index of the serum or plasma sample and digitally convert refractive index to an accurate total protein value. Potential donors with a total protein level of between 6.0 g/100ml and 9.0 g/100ml and that meet other pre-screening requirements are permitted to donate plasma.

These devices are FDA Class I in vitro diagnostic devices; FDA Regulation Number 862.2800 Refractometer for clinical use; with Product Code PSM, Refractometer for Donor Testing.

Participants will be those presenting for plasma donation. Ordinary pre-screening for plasma donation includes a fingerstick and the collection of one small capillary tube (\<0.085ml) of blood. Participants who have consented to this study will be asked to supply a second capillary tube (\<0.085ml) of blood for the clinical trial. The two samples will be centrifuged and hematocrit measured per the plasma center's standard operating procedures. One sample will be dispensed onto the TS Meter-DSP for total protein measurement. This measurement will be used to determine the individual's eligibility to donate plasma. The second sample will be dispensed on to the investigational Refractix DSP for total protein measurement. The measurements from both devices will be recorded on the Case Report Form. The study requires (36) participants with total protein measurements between 6.0 g/100ml and 9.0 g/100ml.

The primary analysis of data will include:

1. a Grubb's test for outlying paired difference values will be performed. If outliers are detected and a plausible non-device failure mechanism can be identified (eg, operator or data entry error), the values will be excluded from the acceptance criterion calculations;

2. an alpha=0.05 level equivalence test with an equivalence interval of +/- 0.3 g/100ml will be performed using the TOST (Two One Sample Test) procedure. A paired t-test 90% confidence interval for the difference in paired sample measurements will be constructed and reported; and

3. calculate the percentage of measurement pairs falling outside of the +/-0.3 tolerance.

Recruitment & Eligibility

Status
NOT_YET_RECRUITING
Sex
All
Target Recruitment
36
Inclusion Criteria
  • Subjects must be between the ages of 18 and 90 years old
  • Subjects must be able and willing to provide signed informed consent
  • Subjects must be able to follow study instructions
Exclusion Criteria
  • Insufficient amount of plasma sample for use on either device in the study.
  • Plasma samples of with total protein measurements outside of the 6.0 g/100ml to 9.0 g/100ml range will be excluded from the data analysis.

Study & Design

Study Type
OBSERVATIONAL
Study Design
Not specified
Arm && Interventions
GroupInterventionDescription
TS Meter-DSP Measurement and Refractix DSP MeasurementSerum Total Protein Measurement with TS Meter-DSP and Refractix DSPThe study will involve blood serum samples from (36) participants. Participants will be those presenting at a plasma collection center for routine plasma donation. Ordinary pre-screening for plasma donation includes a fingerstick and the collection of one small capillary tube (\<0.085ml) of blood. Participants will be asked to provide a second capillary tube (\<0.085ml) of blood from this fingerstick.
Primary Outcome Measures
NameTimeMethod
Serum Total Protein Measurement ComparisonSamples will be collected over one or two days at the plasma center. Data analysis by sponsor will be completed within 8 weeks.

The purpose of this study is to show the investigational Refractix DSP refractometer yields equivalent measurement results to the predicate device TS Meter-DSP through comparison testing using human plasma samples.

Secondary Outcome Measures
NameTimeMethod

Trial Locations

Locations (1)

Hemarus Plasma

🇺🇸

Lauderhill, Florida, United States

Hemarus Plasma
🇺🇸Lauderhill, Florida, United States
Linton W Davis
Contact
754-336-1503
ldavis@hemarus-plasma.com
Melvyn W Kramer, MD
Principal Investigator