Clinical Evaluation of the RHEA Vital Sign Vigilance System in Hospital Patients Patients
- Conditions
- Vital Sign Monitoring
- Interventions
- Device: Investigational GroupDevice: reference group-Earlysense system
- Registration Number
- NCT03332147
- Lead Sponsor
- Darma Inc.
- Brief Summary
This study compares both of the investigational device and reference device to the gold standard Patient Monitor of the accuracy of heart rate and respiratory rate monitoring. The primary variables for the analysis of heart rate (HR) and respiratory rate (RR) are each of the overall mean roots mean square difference (RMSD). Results of the investigational device and reference device compared with the patient monitor is evaluated to see if there are equivalent. Motion and bed exit accuracy will be evaluated by comparing with the manual observation.
- Detailed Description
The testing will be conducted in three periods. In the Accuracy Tests (periods
1 and 2), one of the contact-free devices will be used to monitor the HR and RR, while being compared with the Patient Monitor. The RHEA investigational device and reference device will be used to monitor the same subject at different times, as the sensors of the two devices are placed at a similar location of the bed. HR and RR will be recorded for all three devices.
The primary hypotheses for HR, are as follows:
Adults HO: μ = 3.25 vs. HA: μ \< 3.25
The primary hypotheses for RR, are as follows:
Adults HO: μ = 2.25 vs. HA: μ \< 2.25 where μ represents the population mean HR or RR. If the upper bound of the confidence interval is less than the hypothesized value, the corresponding null hypothesis will be rejected. Comparison of the performance of the two contract-free devices will be performed on the recorded results via statistical analysis. The hypothesis here is that the two devices have equivalent performance. The third test will assess the ability of the RHEA device to accurately detect motion in or exit from the bed, as compared to the manual observation. Motion and no motion accuracy will each be calculated along with their respective two-sided 95% confidence intervals.
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 80
- Male or female, 18 years old and above.
- Provide written informed consent.
- Weight ranging from 20 to 150 kg inclusive.
- Are located in a non-ICU hospital setting.
- Agree to not eat during the testing period.
- Agree to keep still.
- Are connected to a device which may interfere with the device monitoring in this study.
- Are receiving any bedside care which may be incompatible with the study procedures.
- Sleep apnea.
- Pregnant or breastfeeding.
- A likely need to receive or undergo a procedure during the testing period.
- Cannot accept a nasal cannula, or have a monitor lead placed on the chest.
- A significant medical condition in the judgement of the investigator, which may compromise the study testing procedures.
- Are wearing pacemaker or defibrillator.
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- CROSSOVER
- Arm && Interventions
Group Intervention Description Investigational Group Investigational Group The RHEA device and Patient Monitor Device are used to monitor the heart rate (HR) and respiratory rate (RR) of participants. reference group-Earlysense system reference group-Earlysense system The reference device and Patient Monitor Device are used to monitor the heart rate (HR) and respiratory rate (RR) of participants.
- Primary Outcome Measures
Name Time Method RMSD of Investigational Device's Heart Rate and Gold Standard Device's heart rate 45 days Record 20 valid pair heart rate displayed in the investigational device and patient monitor at the same time every one minutes.Calculate the root mean square difference(RMSD) of the heart rate for the investigational device and the gold standard device.
RMSD of Investigational Device's Respiratory Rate and Gold Standard Device's Respiratory rate 45 days Record 20 valid pair respiratory rate displayed in the investigational device and patient monitor at the same time every one minutes.Calculate the root mean square difference(RMSD) of the respiratory rate for the investigational device and the gold standard device.
- Secondary Outcome Measures
Name Time Method Heart Rate t Test Comparison result between the Investigational Device and the Reference Device 45 days Record 20 valid heart rate displayed in the reference device and patient monitor at the same time every one minute.
Calculate the t test results for heart rate for the reference device and the investigational device.Motion Notification Accuracy(%) 45 days Record the time for each instructed movements from manual observation and time for movements notification shown in the investigational device. Calculate the times matched between these two to get the accuracy.
Bed Exit Notification Accuracy(%) 45 days Record the time for each instructed bed exit from manual observation and time for bed exit notification shown in the investigational device. Calculate the times matched between these two to get the accuracy.
Respiratory Rate t Test Comparison result between the Investigational Device and the reference device 45 days Record 20 valid respiratory rate displayed in the reference device and patient monitor at the same time every one minute. Calculate the t test results for respiratory rate for the reference device and the investigational device.
Trial Locations
- Locations (3)
Tufts Medical Center
🇺🇸Boston, Massachusetts, United States
Virginia Commonwealth University
🇺🇸Richmond, Virginia, United States
Medstar Health Research Institute
🇺🇸Washington, District of Columbia, United States