Evaluation of Micropore's SpiraLith TM Absorbents Into Anesthesia Breathing Circuits

Rutgers, The State University of New Jersey1 个研究点 分布在 1 个国家目标入组 186 人开始时间: 2015年7月最近更新:
适应症
干预措施
相关药物

试验速览

阶段
不适用
状态
已完成
入组人数
186
试验地点
1
主要终点
Duration of Use Under Clinical Anesthesia of Each Absorbent

研究概览

简要总结

This is an observational study investigating the duration of use and cost of the new CO2 absorbent Micropore's Spiralith®, compared to Drägersorb® 800+, Drägersorb® Free.

详细描述

The Principal Investigator will obtain IRB approval under expedited review. No patient written informed consent will be required as this is a process improvement minimal risk study that does not necessitates use of patient Protected Health Information. Participants will not be actually enrolled. There will be no change to the standard of care resulting from data collection on CO2 absorbent use, and no patient identifiers will be exposed. The study will take place in three operating rooms, three rooms equipped with Drager Appollo® anesthesia workstations (Draeger Medical Inc., USA) requiring use of CLIC style absorbents. An unblinded member of the research team will have access to the randomization schedule and will be responsible to place Drägersorb® 800+, Drägersorb® Free, or Micropore SpiralithTM absorbents in the designated operating rooms with three Appollo® (CLIC style absorbent). The randomly selected absorbent will be loaded on the anesthesia workstation prior to the patient's arrival in the operating room and the date, time and randomization coding will be marked on the cassette as it appears on the randomization schedule. The anesthesia providers will use low flow anesthesia with or without Sevoflurane in the 3 selected operating rooms. The practice at this institution is to use a semi-closed breathing circuit and maintain a fresh gas flow rate of approximately 2L/min with a selected tidal volume of 6-8 mL/Kg during the maintenance phase of the procedure. Flow rate would be approximately 10 L/min during the induction of anesthesia. The investigators will document time that is required to achieve 90% of the sevoflurane target concentration. These guidelines will be clearly indicated on the anesthesia workstation to ensure consistency across the multiple providers. Data logger files will be saved at the end of each case using the same coding label as the marked cassette. When the absorbent is replaced, the date and time of removal should be clearly marked on the cassette. This same date and time information should also be recorded on the source document ensuring that operating room numbers and anesthesia workstations are clearly and accurately recorded. In order to maintain the anesthesia provider blinded to the type of absorbent used, a white sheet of paper will be taped around the outside of the cassette compartment. The time that the patient is connected and disconnected from the breathing circuit must be recorded on the source document. In addition the investigators will document The absorbent unit will be replaced when the inspiratory CO2 concentration (obtained from the inspiratory limb of the breathing circuit) reaches 4 mmHg. This is an acceptable maximal inspiratory CO2 concentration which leaves sufficient time to complete a case, thereby avoiding cassette replacement during anesthesia. The anesthesia workstation fresh gas flows will be turned off at the time that the patient is disconnected from the breathing circuit to avoid variability in absorbent water content.

The absorbing capacity of the cassettes will be determined by calculating the duration of absorbent use from the time of first use to the time of cassette removal. Each absorbent will be evaluated through a total of 5 rounds. We expect to have about 60 surgeries for each absorbent. Averaged duration of absorbent use will be compared between different types of absorbents (Drägersorb® 800+, Drägersorb® Free, Micropore SpiralithTM).

研究设计

研究类型
Observational
观察模型
Other
时间视角
Other

入排标准

性别
All
接受健康志愿者

入选标准

  • All comers to the 3 designated Operating Rooms.

排除标准

  • 未提供

研究组 & 干预措施

Randomized to Micropore SpiraLith

lithium hydroxide was studied for use in anesthesia as a possible replacement for calcium absorbents. This agent has been used for CO2 absorption in the military and in aerospace for over 50 years due to its high capacity and efficiency in the removal of CO2. It was however not considered usable by the medical industry due to concerns with its granular form. It has now been demonstrated that LiOH does not interact with commonly used inhalation anesthetic agents and appears to have higher CO2 removal capability.7,8

干预措施: Micropore SpiraLith (Device)

Randomized to Micropore SpiraLith

lithium hydroxide was studied for use in anesthesia as a possible replacement for calcium absorbents. This agent has been used for CO2 absorption in the military and in aerospace for over 50 years due to its high capacity and efficiency in the removal of CO2. It was however not considered usable by the medical industry due to concerns with its granular form. It has now been demonstrated that LiOH does not interact with commonly used inhalation anesthetic agents and appears to have higher CO2 removal capability.7,8

干预措施: Sevoflurane (Drug)

Randomized to Drager 800 Absorbent

Calcium hydroxide lime is one of the newer clinically available carbon dioxide absorbents and probably considered the current standard of care. It is mainly composed of calcium hydroxide and calcium chloride and contains two setting agents: calcium sulfate and polyvinylpyrrolidine which contribute to the increased hardness and porosity of this absorbent. The most significant advantage of calcium hydroxide lime over other agents is that it is produced without sodium and potassium hydroxide which are strong bases.

干预措施: Carbon Dioxide Absorbents (Device)

Randomized to Drager 800 Absorbent

Calcium hydroxide lime is one of the newer clinically available carbon dioxide absorbents and probably considered the current standard of care. It is mainly composed of calcium hydroxide and calcium chloride and contains two setting agents: calcium sulfate and polyvinylpyrrolidine which contribute to the increased hardness and porosity of this absorbent. The most significant advantage of calcium hydroxide lime over other agents is that it is produced without sodium and potassium hydroxide which are strong bases.

干预措施: Sevoflurane (Drug)

Randomized to Drager Free

Calcium hydroxide lime is one of the newer clinically available carbon dioxide absorbents and probably considered the current standard of care. It is mainly composed of calcium hydroxide and calcium chloride and contains two setting agents: calcium sulfate and polyvinylpyrrolidine which contribute to the increased hardness and porosity of this absorbent. The most significant advantage of calcium hydroxide lime over other agents is that it is produced without sodium and potassium hydroxide which are strong bases.

干预措施: Carbon Dioxide Absorbents (Device)

Randomized to Drager Free

Calcium hydroxide lime is one of the newer clinically available carbon dioxide absorbents and probably considered the current standard of care. It is mainly composed of calcium hydroxide and calcium chloride and contains two setting agents: calcium sulfate and polyvinylpyrrolidine which contribute to the increased hardness and porosity of this absorbent. The most significant advantage of calcium hydroxide lime over other agents is that it is produced without sodium and potassium hydroxide which are strong bases.

干预措施: Sevoflurane (Drug)

结局指标

主要结局

Duration of Use Under Clinical Anesthesia of Each Absorbent

时间窗: Assesed for the duration of each absorbent's use, up to 5 days

Average lifespan of Spiralith® (minutes), Drägersorb® 800 Plus (minutes), Drägersorb® Free (minutes)

次要结局

  • Cost of Each Absorbent's Use(Assesed for the duration of each absorbent's use, up to 5 days)