Clinical Efficacy Evaluation of Supraglottic Airway Device SafeLM™ in Paediatric Population

National University of Malaysia2 个研究点 分布在 1 个国家目标入组 60 人开始时间: 2025年12月2日最近更新:

试验速览

阶段
不适用
状态
尚未招募
发起方
入组人数
60
试验地点
2
主要终点
Oropharyngeal leak pressure (OPLP) in cmH2O

研究概览

简要总结

The goal of this interventional study is to evaluate the clinical efficacy of SafeLM™, a video laryngeal mask airway device in paediatric population. The main questions it aims to answer are:

  1. To determine the success rate, ease of device insertion, time for insertion, and how well the device seals the airway by measuring the oropharyngeal leak pressure (OPLP) of SafeLM™.
  2. To determine the incidence of complication associated with SafeLM™ usage.
  3. To study the association of glottic view with success rate and rate of complication.

Participants perioperative clinical data will be collected for data analysis.

详细描述

Paediatric patients who are scheduled for surgery where supraglottic airway is deemed appropriate will be recruited with written consent. Following preoxygenation, application of standard monitors, and induction of general anaesthesia, an appropriate size of SafeLM™ will be inserted according to manufacturer's recommendation. Cuff will be deflated and flattened prior, with the back of laryngeal mask lubricated. The device will be advanced by sliding the laryngeal mask down the palatopharyngeal curve along the midline in the mouth until the end of the mask reach the hypopharynx cavity and resistance is met, with the process visualised in real time through the video camera. If there is malalignment, correction manoeuvre such as jaw trust, head reposition, or mask redirection will be done. Once the distal end of mask reaches the hypopharyngeal cavity, the cuff is inflated with the recommended volume of air according to the manufacturer recommendation, until adequate cuff seal is achieved. Then, the patient is put on either spontaneous or pressure-controlled mechanical ventilation. Successful placement will be confirmed. If insertion difficulty is encountered and the laryngeal mask is required to be taken out from the oral cavity, insertion can be reattempt with the same or different size of SafeLM™ after adequate preoxygenation, and it will be considered as second attempt. If it remains unsuccessful on the 3rd attempt, SafeLM™ insertion will be abandoned, subsequent management will be decided by the attending anaesthetist. Glottic exposure grades at the end of each attempt is recorded. The time for insertion is measured for each attempt, using a stopwatch, by obtaining the duration in between the moment when the anaesthetist starts holding the SafeLM™ device, until the appearance of 2 complete continuous capnography waveform on the monitor. Following successful SafeLM™ insertion, the camera can be removed, and orogastric tube will be inserted via the gastric port only if there is clinical indication. Subsequently, oropharyngeal leak pressure (OPLP) is measured and recorded by dialing the fresh gas flow rate to 3 L/min and putting the adjusting pressure limiting (APL) valve at 40 cmH2O, in manual ventilation mode. The equilibrium pressure that the pressure gauze pointer settles down at is the OPLP. The anaesthesia is maintained using TIVA or volatile agent as per the anaesthetist's discretion, keeping the MAC at least 1.0 or BIS level at 40-60, whichever is applicable. All the patients will receive IV dexamethasone 0.2 mg/kg and IV granisetron 0.02 mg/kg as post operative nausea vomiting (PONV) prophylaxis. At the end of anaesthesia, anaesthetic agent administration is terminated. If non-depolarising neuromuscular blockade is given during anaesthesia, an appropriate dose of reversal agent will be given according to the depth of neuromuscular blockade based on either clinical signs or nerve stimulation examination, whichever applicable. After the adequate neuromuscular blockade reversal is confirmed, the laryngeal mask can be removed at the awake state or deep plane of anaesthesia. Patient will be monitored in the post-anaesthesia care unit for at least 30 minutes prior sending back to the designated ward for continuation of care.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other
盲法
None

入排标准

年龄范围
0 Years 至 18 Years(Child, Adult)
性别
All
接受健康志愿者

入选标准

  • Patients who are scheduled for a surgery under general anaesthesia where supraglottic airway is deemed appropriate.
  • ASA I or II
  • Age up to 18 years old
  • Body weight of 10 kg and above

排除标准

  • Presence of risk of gastric regurgitation or aspiration
  • Recent upper respiratory tract infection within 2 weeks
  • Uncontrolled respiratory co-morbidity such as bronchial asthma, chronic lung disease, restrictive lung disease
  • Presence of congenital airway abnormalities
  • Patient or next of kins refusal in participating this study

结局指标

主要结局

Oropharyngeal leak pressure (OPLP) in cmH2O

时间窗: From the time of successful device placement confirmation to the time of determination of OPLP before commencement of surgery

Oropharyngeal leak pressure (OPLP) is measured and recorded by dialling the fresh gas flow rate to 3 L/min and putting the adjusting pressure limiting (APL) valve at 40 cmH2O, in manual ventilation mode. The equilibrium pressure that the pressure gauze pointer settles down at is the OPLP. The measurement is done after successful device placement confirmation, once OPLP value is obtained, positive pressure ventilation of spontaneous ventilation will resume.

Success rate of device placement in percentage

时间窗: Perioperative

Percentage of patient experiencing successful insertion of SafeLM™ The criteria of successful insertion: (i) good chest rise during positive pressure ventilation (ii) no significant air leakage (no audible leak on ausculatation, and the ventilator bellow not collapsing (iii) two or more continuous capnography waveforms (iv) able to achieve a minimum expiratory tidal volume of 6ml/kg Percetage calculation: number of participants fulfillling successful insertion criteria within 3 insertion attempts divided by total number of participants, multiplied by 100 percents

Glottic exposure grades

时间窗: At the time of completion of device placement procedure, when the distal end of the cuff reaches the hypopharyngeal cavity, resistance is met, and the cuff is inflated with appropriate volume of air as per manufacturer recommendation.

SaCoVLM™ Glottic exposure grades visualised on the videoscope are recorded for each patient Grade 1: Visualization of the lateral part of the right aryepiglottic fold and part of the laryngeal inlet, and the ventilation was good Grade 2: Visualization of the bilateral aryepiglottic fold and part of laryngeal inlet, and the ventilation was good Grade 3: Visualization of all laryngeal inlet and posterior glottis Grade 4: Visualization of the whole glottis

次要结局

  • The ease of SafeLM™ insertion(Perioperative)
  • Anaesthetists' overall satisfactory score(Perioperative)
  • Rate of post-extubation bronchospasm(Perioperative)
  • Insertion time in seconds (s)(Periprocedural)
  • Rate of post-extubation gastric aspiration event(Perioperative)
  • Rate of post-extubation sore throat event(Perioperative)
  • Rate of intraoperative laryngospasm(Intraoperative)
  • Rate of intraoperative desaturation event(Intraoperative)
  • Rate of intraoperative gastric aspiration event(Intraoperative)
  • Rate of intraoperative bronchospasm(Intraoperative)
  • Rate of post-extubation finding of blood stain(Perioperative)
  • Rate of post-extubation laryngospasm(Perioperative)