Eye Muscle Needle Electromyogram (EMu) Study

Completed
Conditions
Strabismus
Registration Number
NCT03895632
Lead Sponsor
Liverpool University Hospitals NHS Foundation Trust
Brief Summary

In some cases, strabismus, or squint, is treated with injections of Botulinum toxin (BTX) to temporarily relax the eye muscle responsible for causing the eye to be pulled to one side. Due to the deep location of the muscles, electromyogram (EMG) needle guidance is used to help ensure the toxin is delivered accurately. EMG needle guidance involves listening to the EMG signal from the tip of the delivery needle - when the needle is in the right place a sound akin to "rain on a tin roof" is heard and the BTX can be injected.

From previous research, clearly clinicians want to improve the technology of this procedure and increase the treatment efficacy and repeatability. After all, the current procedure is imprecise and subjective for what is a small target, with the eye muscles being around 2.5 mm to 4.7 mm in diameter. There may also be an increase in the likelihood of side effects such as droopy eyelids when the toxin spreads beyond the target muscle.

There is a strong need for new knowledge to improve EMG needle guidance in this context. It is hypothesised that the EMG signal could be cleaned and mined for information in real time during the procedure, thus providing the clinician with better information to guide the positioning of the needle before injecting. This pilot study will deliver the knowledge essential to indicate the feasibility of doing this. If feasible, this study will inform a full study with the aim of benefitting patients by allowing new technology to be developed to improve the treatment of strabismus with BTX.

Detailed Description

Not available

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
25
Inclusion Criteria

i. Adults undergoing planned, routine EMG needle guided BTX injection into either lateral or medial rectus muscle for the treatment of strabismus.

ii. Patients being treated by Mr Yagan.

Read More
Exclusion Criteria
  • Patients with Chronic progressive external ophthalmoplegia (CPEO) are excluded; a highly attenuated EMG signal is expected.
Read More

Study & Design

Study Type
OBSERVATIONAL
Study Design
Not specified
Primary Outcome Measures
NameTimeMethod
α (Slope) Parameter of the Electromyogram (EMG) Signal Power Spectral Density (PSD) Plot as Needle Approaches Then Enters Muscle.Approximately 5 minutes, i.e. immediately prior to needle insertion until Botulinum toxin (BTX) delivery and needle withdrawal

The continuous change in the EMG signal as the needle is (a) distal to, (b) approaches, and (c) enters, the target extraocular muscle.

Welch's periodogram estimator (Hamming window, 256-point FFT, 50% overlap) was applied to each of the three signal segments described above to obtain an estimate of the PSD of the signal. A robust linear model was fit to estimate the α (slope) parameter of each PSD plot.

Secondary Outcome Measures
NameTimeMethod
Clinical Score of EMG ChangeApproximately 5 minutes, i.e. immediately prior to needle insertion until Botulinum toxin (BTX) delivery and needle withdrawal

Clinician's score of the step change from baseline EMG to 'active' EMG level heard immediately prior to delivery of Botulinum toxin (BTX) from 1-5. Value from 1 = no significant step change heard, to 5 = a very clear step change heard.

Clinical Score of Needle Placement AccuracyApproximately 5 minutes, i.e. immediately prior to needle insertion until Botulinum toxin (BTX) delivery and needle withdrawal

Clinician's score of their confidence in accurate needle placement with respect to the target muscle. Given as a percentage, a higher score indicates more confidence in accurate needle placement.

Clinical Score of EMG Quality.Approximately 5 minutes, i.e. immediately prior to needle insertion until Botulinum toxin (BTX) delivery and needle withdrawal

Clinician's score of overall quality of the EMG signal heard during the procedure. A higher score indicates better quality of EMG signal, and is quantified as 1 to 5, where: 1 indicates signal not present and 5 indicates excellent quality EMG signal.

Post-treatment Angle of Deviation2 weeks

Change in participant's angle of deviation following treatment, measured in prism dioptres.