Validation of Point Partial User Needs With Partial Finger Amputees

Not Applicable
Completed
Conditions
Amputation; Traumatic, Hand
Registration Number
NCT05012657
Lead Sponsor
Point Designs
Brief Summary

The objective of the device feasibility study will be to validate the user needs of the Point Partial system. This study will be a single group intervention model where one group of 5 partial finger amputees will be asked to perform several tasks. Successful completion of a task results in a fulfilled user need. Failure to complete a task results in an unfulfilled user need.

Detailed Description

Not available

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
5
Inclusion Criteria
  • Partial finger amputees with partial absence of index and/or middle fingers and presence of thumb. Additional finger loss is acceptable if all other criterion are met.
  • Fluent in English
  • Individuals aged 18 or greater
  • Patients may present with either fused or mobile MCP joints.
Exclusion Criteria
  • Significant cognitive deficits as determined upon clinical evaluation
  • Significant neurological deficits as determined upon clinical evaluation
  • Significant physical deficits of the residual limb impacting full participation in the study as determined upon clinical evaluation
  • Uncontrolled pain or phantom pain impacting full participation in the study as determined upon OT evaluation
  • Serious uncontrolled medical problems as judged by the project therapist.

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Primary Outcome Measures
NameTimeMethod
VAL-1: System Comfort Validation TestDay 1 (8 hours)

User wears Point Partial(s) for a full day (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be comfortable for a full day's use. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

VAL-2: Unilateral Function Validation TestDay 1 (10 minutes)

User flexes and extends the Point Partial(s) without using contralateral hand (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall allow the user to flex and extend finger without their second hand. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

VAL-3.1: System Robustness Validation Test With 25 lb BagDay 1 (10 minutes)

User lifts a 25 lb. bag with Point Partial(s) (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

VAL-3.2: Robustness Validation Test With HammerDay 1 (10 minutes)

User grips a hammer with Point Partial(s) and drives a nail into a piece of wood (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

VAL-4: Object Release Validation TestDay 1 (10 minutes)

User releases Point Partial(s) while performing an active grasp by pressing the dorsal button (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be releasable while performing active grasps. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

VAL-5: Position Function Validation TestDay 1 (10 minutes)

User positions Point Partial(s) into each of the locking positions (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a smooth ratcheting mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

VAL-6: Extension Function Validation TestDay 1 (10 minutes)

User activates spring-back mechanism by using (1) contralateral hand, (2) tabletop, and (3) side or thigh throughout day without failed extension (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a reliable auto spring-back mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

VAL-7: Grip Functions Validation TestDay 1 (10 minutes)

User grips a 5 lb. cylindrical smooth object using Point Partial(s) for 10 seconds (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have high grip surface friction. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

VAL-8: Fine Motor Function Validation TestDay 1 (10 minutes)

User picks up 1 coin from a smooth tabletop (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall have a pronounced fingernail. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

VAL-9: Overall Function & Robustness Validation TestDay 1 (8 hours)

User performs VAL-1 through VAL-8 (Pass/Fail). This validation method was used to confirm the following user need of the Point Partial has been met: Point Partial shall be robustly attached to mounting system. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Partials were mounted to the socket. Then, each participant attempted the validation test.

Secondary Outcome Measures
NameTimeMethod

Trial Locations

Locations (2)

University of Colorado Denver | Anschutz Medical Campus

🇺🇸

Aurora, Colorado, United States

Point Designs

🇺🇸

Lafayette, Colorado, United States