Evaluation of the Performance of Contour Next® and Contour Plus Elite® Blood Glucose Monitoring Systems (BGMS) in Arterial Blood Samples From Hospitalized Adults

Ascensia Diabetes Care1 个研究点 分布在 1 个国家目标入组 120 人开始时间: 2022年11月29日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
120
试验地点
1
主要终点
Number of Results From the Contour Next BGMSs Reference Values Within ±12.5% of Reference Values

研究概览

简要总结

The purpose of the study is to extend the intended use of two BGMSs to include testing of arterial blood by Health Care Professionals (HCP) in a clinical setting.

详细描述

This trial will evaluate the performance of both Contour Next BGMS and Contour Plus Elite BGMS using arterial blood from adult patients hospitalized in a Critical Care Unit (medical and surgical intensive care units (ICUs)). The investigational BGMS will be tested by a Point-of-Care (POC) operator in a clinical setting using residual arterial blood samples from adults who underwent prescribed arterial blood tests that were deemed necessary due to their medical conditions.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 100 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Patients who are at least 18 years old.
  • Residual Arterial blood samples collected from subjects hospitalized in a Critical Care Unit (medical and surgical intensive care units (ICUs)).
  • Sample blood volume must be sufficient to complete investigational testing procedures clinical laboratory testing.

排除标准

  • Residual arterial blood samples collected from subjects previously enrolled. and evaluated for this study.

结局指标

主要结局

Number of Results From the Contour Next BGMSs Reference Values Within ±12.5% of Reference Values

时间窗: 1 day

At least 95% of all blood glucose results shall fall within ±12.5% of reference values (laboratory method) for glucose concentration ≥100 mg/dL(5.55 mmol/L )and within ±12 mg/dL(±0.67 mmol/L) at glucose concentrations \<100 mg/dL(5.55 mmol/L).

Number of Results From the Contour Plus Elite BGMSs Reference Values Within ±12.5% of Reference Values

时间窗: 1 day

At least 98% of values should be within ±12.5% of reference values (laboratory method) for glucose concentration ≥75 mg/dL(4.16 mmol/L) and within ±15 mg/dL(±0.83 mmol/L) at glucose concentrations \< 75 mg/dL(4.16 mmol/L).