MedPath

Controlled Acute Hypoxia Study - Abbreviated Sensor Line

Completed
Conditions
Hypoxia
Healthy
Registration Number
NCT02216344
Lead Sponsor
Medtronic - MITG
Brief Summary

The primary objective of the study is to validate the proposed claims for the OxiCable device, for pulse rate and saturation accuracy in a diverse subject population over a specified saturation range.

Detailed Description

The general procedure of invasive controlled hypoxia clinical studies is to directly compare test pulse oximeter measurements to saturation measurements made by a multi-wavelength CO-oximeter taken from arterial blood samples from healthy human subjects. Following arterial cannulation, hypoxia is induced by reducing the percent of inspired oxygen the subject breathes while simultaneous measurements are recorded from the test pulse oximeters and arterial blood samples at targeted levels of saturation.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
13
Inclusion Criteria
  • Male or female of any race
  • 18-50 years old, inclusive
  • Females: negative urine pregnancy test on the day of study participation (prior to exposure to hypoxia)
  • Completed within the last year: physical exam by a licensed physician, physician assistant (PA), or advanced practice nurse; including a 12 lead ECG, a medical history, and blood test (complete blood count and sickle cell trait/disease screening)
  • Meets specific demographic requirements for the monitoring device under study
  • Willing and able to provide written informed consent
  • Able to participate for the duration of the evaluation
Exclusion Criteria
  • Under 18 years or over 50 years of age
  • Pregnant and/or lactating women
  • Hypertension: on three consecutive readings, systolic pressure greater than 145 mm Hg or diastolic pressure greater than 90 mm Hg
  • Premature ventricular contractions (PVCs) that are symptomatic or occur at a rate of more than four per minute
  • History of seizures (except childhood febrile seizures) or epilepsy
  • History of unexplained syncope
  • Daily or more frequent use of anxiolytic drugs (benzodiazepines) for treatment of anxiety disorder
  • Recent history of frequent migraine headaches: average of two or more per month over the last year
  • Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test site needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
  • History of acute altitude sickness at or below moderate elevation (up to 5,000-10,000 feet) defined as three or more of the following symptoms: moderate to severe headache, general malaise, dizziness/lightheadedness, nausea/vomiting, fatigue/weakness, shortness of breath, nosebleed, persistent rapid pulse, or peripheral edema
  • History of significant respiratory disease such as severe asthma or emphysema
  • Sickle cell disease or trait
  • Clinically significant abnormal finding on medical history, physical examination, clinical laboratory test or ECG. Clinical significance will be assessed by the principal investigator or study physician as designated.
  • History of stroke, transient ischemic attack or carotid artery disease
  • History of myocardial ischemia, angina, myocardial infarction, congestive heart failure or cardiomyopathy
  • History of chronic renal impairment
  • Severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors
  • Unwillingness or inability to remove colored nail polish from test digit(s)
  • Unwillingness or inability to give informed consent
  • Prior or known allergies to: lidocaine (or similar pharmacological agents, e.g., Novocain) or heparin
  • Recent arterial cannulation (i.e., less than 30 days prior to study date)
  • Six or more arterial cannulations of each (right & left) radial artery
  • History of clinically significant complications from previous arterial cannulation
  • Current use of blood thinners: prescription or daily aspirin use

Study & Design

Study Type
OBSERVATIONAL
Study Design
Not specified
Primary Outcome Measures
NameTimeMethod
Pulse oximetry measurements of saturation from the test device will be compared to CO-Oximetry measurements of arterial oxygen saturation to demonstrate that the test sensor meets the oxygen saturation accuracy specifications for SpO2.up to 6 months
Secondary Outcome Measures
NameTimeMethod

Trial Locations

Locations (1)

Boulder Clinical Laboratory

🇺🇸

Boulder, Colorado, United States

© Copyright 2025. All Rights Reserved by MedPath