Rayner Toric IOL Rotational Stability
- Conditions
- CataractCorneal Astigmatism
- Interventions
- Procedure: Cataract surgeryDevice: Rayner T-flex Toric IOLs
- Registration Number
- NCT02393313
- Lead Sponsor
- Tel-Aviv Sourasky Medical Center
- Brief Summary
This study evaluates the Rayner T-flex Toric IOLs for spherical and astigmatic correction, and rotational stability in cataract surgery. 50 patients with cataract and corneal astigmatism \>= 2 diopters will undergo cataract surgery, with implantation of Rayner toric intraocular lens. Post-operative examination will include IOL axis measurements and refraction.
- Detailed Description
Background During cataract surgery, an intraocular lens is implanted in the bag. A monofocal lens allows spherical correction only. 15% to 29% of patients with cataracts have more than 1.50 diopters of astigmatism according to current estimations.1 Astigmatic correction can be performed during cataract surgery: the main clear corneal incision can be done at the steep axis, limbal relaxing incisions (LRI) or by opposite clear corneal incision (OCCI) can be performed. These techniques are not fully predictable and stable over time.2 Optimal corneal astigmatic correction can be done by implantation of a toric intra-ocular lens(IOL), with predictable and stable results over time. The toric IOL must be positioned in the correct axis. Rotation of the lens will reduce the astigmatic correction. Approximately one degree of off-axis rotation results in a loss of up to 3.3% of the cylinder power. 3 It is crucial for the intraocular lens to keep its correct axis for optimal astigmatic correction.
A number of toric intraocular lens are approved for implantation such as - Acrysof toric (Alcon), Tecnis toric IOL (Abbott medical), STAAR toric IOL (STAAR surgical company). This study will examine T-flex (Rayner) IOL. The rotational and refractive stability of this IOL has been tested on a relatively small number of patients. 4, 5
Objective Examination of Rayner T-flex Toric IOLs (573T and 623T; Rayner Intraocular Lenses Ltd,East Sussex, United Kingdom) for spherical and astigmatic correction, and rotational stability.
Methods Prospective study on 50 eyes. Patients will be recruited during examination in pre-operative clinic. This examinations include visual acuity, slit-lamp biomicroscopy, applanation tonometry, fundus examination after instillation of tropicamide 0.5% and phenylephrine 10%, IOL calculation with IOLMaster(Zeiss), which also measures corneal astigmatism. After acquiring informed consent, further examinations will be done including corneal topography with Galilei Dual Scheimpflug Analyzer ( Zeimer Ophthalmology), and Tomey keratometer, auto-refractometry and subjective refraction.
Since corneal astigmatism is often binocular, patients with second eye corneal astigmatism, will be offered a toric IOL, even if unsuitable for inclusion in this study. All operations will be done by a senior surgeon, experienced in toric IOL implantations. Rayner T-flex Toric IOLs (573T / 623T; Rayner Intraocular Lenses Ltd,East Sussex, United Kingdom) will be implanted. IOL calculation will be done using Ranyer's online calculator (WWW.TORICIOL.RAYNER.COM).
Post-operative examination will include IOL axis measurements, IOL photography and refraction after a week, month and three months.
Recruitment & Eligibility
- Status
- UNKNOWN
- Sex
- All
- Target Recruitment
- 50
- Corneal astigmatism ≥ 2.00 measured by IOLMaster
- Pterygium (can be included 6 months after successful surgical excision)
- Corneal pathologies including corneal scars causing corneal irregularity
- Severe dry eye
- Traumatic cataract
- Pregnant women
- Severe macular disease or macular atrophy
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description Cataract surgery toric intraocular lens Cataract surgery Rayner T-flex Toric IOLs (573T / 623T; Rayner Intraocular Lenses Ltd,East Sussex, United Kingdom) will be implanted in the lens capsule Cataract surgery toric intraocular lens Rayner T-flex Toric IOLs Rayner T-flex Toric IOLs (573T / 623T; Rayner Intraocular Lenses Ltd,East Sussex, United Kingdom) will be implanted in the lens capsule
- Primary Outcome Measures
Name Time Method IOL off-axis rotation 3 months after surgery IOL axis will be measured by slit lamp biomicroscopy and compared to the planned axis
- Secondary Outcome Measures
Name Time Method Subjective refraction 3 months after surgery Uncorrected distance visual acuity 3 months after surgery Corrected distance visual acuity 3 months after surgery
Trial Locations
- Locations (1)
Departent of Ophthalmlogy, Tel Aviv Sourasky Medical Center
🇮🇱Tel Aviv, Israel