Evaluation of the PERFormances of a Bioelectrical Impedancemetry Device (Lucie™), for the Measurement of Hydration Parameters in Patients with Renal Failure: a Multicentre Comparative Study
试验速览
- 阶段
- 不适用
- 状态
- 尚未招募
- 发起方
- 入组人数
- 150
- 主要终点
- Hydratation parameters
研究概览
简要总结
The main objective of the observatory study is to evaluate the performance of the Lucie™ medical device for measuring water balance before and after dialysis (water overload, dry weight) in patients with chronic renal failure. We will compare the results and performance of the device with the reference device in nephrology: the BCM®. This comparison is based on bioimpedance. Bioimpedance is an effective technology for monitoring hydration if used regularly, and will give you greater autonomy in your home monitoring. It competes with other technologies that we have found to be less effective. This is what clinical research is seeking to demonstrate: only a clinical study can verify the reliability of this hypothesis. The aim is to help Lucie technology mature so that it can detect early signs of fluid overload, and ultimately prevent complications or hospitalisation.
The process is as follow:
- After removing their shoes and socks, the patient will be invited to step onto the scale;
- Patient will insert a personal card into the card reader and place his right hand loosely on the LucieTM handle, keeping a natural posture;
- The device will take measurements for 45 seconds under the nurse's supervision;
- Patient will go down Lucie and will answer a short form;
- The device is cleaned after each use For optimum results, we recommand patients to keep their feet dry and their hands clean (without cream), and we recommend them to remove their jewellery, to keep their smartphones away from the device (to avoid interference), and to wear a minimum of nylon-free clothing, with empty pockets.
In practical terms, patients are familiar with monitoring their fluid overload using the Fresinus BCM® bathroom scale. They already use it as part of their clinical routine when entering and leaving the dialysis room. This BCM® is used at the start of the dialysis session, before the session begins, by a nurse trained by the manufacturer in its use: he or she knows how to place the electrodes and read the results for status and fluid overload. The nurse also takes patients' usual vital signs (blood pressure, temperature) and notes any incidents during the dialysis session (low blood pressure, cramps, headaches, nausea, etc.) in line with standard practice.
With Lucie™
详细描述
The goal of this observational study is to evaluate the performance of the Lucie™ medical device for the measurement of pre-dialytic hyperhydration by bioimpedance (water volume in L) in groups of patients with renal failure, compared with the reference device, the BCM. The main question its aimed to answer is
The primary hypothesis will be to determine the degree of agreement between the BMC measure of pre-dialytic hyperhydration and Lucie's water balance (volume (L)).
Participants will do the following tasks :
- Patients weigh themselves on entering the dialysis room to find out their total weight and have it checked by the nurse (routine care) with a scale that complies with weighing standard NF EN 45501:2015 and Directive 2014/31/EU;
- Patients weigh themselves with Lucie™ and complete a bio-impedance measurement cycle in 45 seconds (interventional research);
- A nurse measures the blood pressure on the cuff (CE medical mark), standing and lying down, and determines the ultrafiltration for the dialysis session to be entered on the machine (routine care). She records the pre-dialysis information;
- The nurse performs a bio-impedance measurement with the BCM™ reference device from Fresenius Medical Care and records measurements of hyperhydration - or overhydration (OH) -, extracellular water (ECW), total body water (TBW), the ECW/TBW ratio, the E/I ratio (deep breathing test), the OH/ECW ratio (interventional research);
- Patients begin their dialysis session (routine care);
- The nurse records blood pressure, standing and lying down, after the dialysis session and records post-dialysis information and lists any undesirable events during the dialysis session (e.g. cramps, low blood pressure, headaches, etc.);
- Patients are weighed on leaving the dialysis room to check that they are achieving the prescribed target weight (routine care);
- Patients weigh themselves with Lucie™ and complete a cycle of bioimpedance measurements in 45 seconds under the control of a nurse (interventional research).
- Patients weigh themselves a second time with the Lucie™ device and again complete one cycle of bio-impedance measurement in 45 seconds without nurse supervision (interventional research, the aim of which here is to measure inter-operator reliability).
- No bio-impedance measurement by the BCM™ device after the dialysis session given the difficulty of devoting an extra 30 minutes to patients to carry out a second measurement;
- Lucie™ device surfaces are cleaned according to the procedure provided.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Screening
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Patients age : 18 years min
- •Weight minor than 125 kg
- •must be able to read
- •able to get on and off a bathroom scale, with or without assistance;
- •Patient able to stand in balance for 45 seconds with support, alone and barefoot.
排除标准
- •persons with vulnerabilities : pregnant women, nursing mothers, persons deprived of their liberty, adults under legal protection, persons in emergency situations (art. L. 1121-5 to 8 and L.1122-1-2 of the Public Health Code)
- •patient with an open wound on the extremities (palms and soles of the feet in contact with the device);
- •Patient with an electronic cardiac implant;
- •An amputee, with or without a prosthesis;
- •Patients with paralysis, even partial paralysis;
- •Patients with a 3-month mean natraemia of less than 130 mmol/L;
- •Patients fitted with a metal prosthesis;
- •Patients with behavioural problems that prevent them from taking part in the study;
- •Patient already included in another interventional research study.
结局指标
主要结局
Hydratation parameters
时间窗: Through study completion (2 months)
Extra Cellular Water (ECW), Total Body Water (TBW), Intra Cellular Water (ICW) ; the agreement of the variation before and after dialysis with Lucie™ of the hydration parameters (ECW, ICW, ECW/ICW, OH/ECW, TBW) and the actual ultrafiltration of the session displayed on the generator (UF). This measure will help to calculate Bioimpedance measure.
Concordance of Hydration parameters (%)
时间窗: Through study completion (2 months)
Concordance of hydration parameters (OH/ECW (%) (Over Hydration) to Extra Cellular Water), ECW/ICW (%) (Extra Cellular Water to Intra Cellular Water) between the 2 medical devices (Lucie™ and BCM® reference device), as well as sensitivity, specificity, positive predictive value, negative predictive value by Lucie™ medical device.
次要结局
- Weight Measures(Through study completion (2 months))
- Skin/eat Temperature Agreement(Through study completion (2 months))
- Satisfactiory Questionnaire(Through study publication (4 months))
- Medico-economic Impact(From the end of inclusion to one year after CER publication, assessed to 16 months)
- Organisational Impact(From the end of inclusion to one year after CER publication, assessed to 16 months)