Comparison Between Flow Sensor and Colourimetric Capnometer in Verifying the Correct Positioning of the Endotracheal Tube in Term and Preterm Infants
试验速览
- 阶段
- 不适用
- 入组人数
- 105
- 试验地点
- 1
- 主要终点
- Time taken by the flow-sensor to indicate the correct or not correct positioning of the tracheal tube, compared with time taken by the colorimetric capnometer.
研究概览
简要总结
This is a multicentre clinical trial with medical device. As currently recommended by international neonatal resuscitation guidelines, the most reliable method to verify the correct positioning of the endotracheal tube, in association with clinical signs, is the end-tidal capnometry obtained either by infrared spectroscopy or colorimetric method. The aim of the present study is to evaluate whether the flow sensor of a standard mechanical ventilator can discriminate with similar or faster times the correct positioning of the tube after a tracheal intubation attempt compared to the colorimetric capnometer, in newborns undergoing this procedure during hospitalization in the neonatal intensive care unit.
Given the importance of defining a rapid and effective method to prevent possible adverse events of incorrect endotracheal intubation, this study aims to verify whether the ventilator flow sensor can allow a reliable assessment of the correct positioning of the endotracheal tube, with timing and success rates equivalent to or better than the colorimetric capnometer.
详细描述
Specifically, the neonatologist performing the procedure will follow the standard procedure for tracheal intubation, which includes using a colourimetric capnometer to confirm the correct tube positioning. However, he/she will remain blind and not aware of the flow sensor signal, while a study investigator will monitor that.
The investigators will measure the percentage of responses of correct positioning of the endotracheal tube as well as the number of insufflations and the time elapsed before obtaining a signal of correct intubation by simultaneously comparing both methods and checking any differences based on the characteristics of the sample analyzed (e.g. premature infants versus term infants, elective versus emergency intubations).
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Diagnostic
- 盲法
- None
盲法说明
Participants are newborns. The care provider will be the attending physician performing the tracheal intubation procedure as per local protocols, including the colourimetric capnometry analysis. However, he/she will remain blind and not aware of the flow sensor data until the end of the procedure.
入排标准
- 年龄范围
- 20 Minutes 至 3 Months(Child)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Term and preterm infants requiring tracheal intubation within the neonatal intensive care unit, both in elective and emergency conditions.
排除标准
- •intubation at birth
- •intubation during cardiopulmonary resuscitation
- •absence of informed consent by parents
研究组 & 干预措施
all patients undergoing tracheal intubation procedure
In current practice, the endotracheal intubation procedure includes the systematic use of the capnometer, which is immediately connected to the tube after its insertion by direct laringoscopy. Confirmation of the correct tube positioning is established by color change of the capnometer from violet to yellow. Then, the colorimetric capnometer is usually removed before connecting the endotracheal tube to the flow-sensor of the ventilator.
For the purpose of our study, after the tube insertion the care provider will connect in line both the capnometer and the ventilator flow-sensor to the tracheal tube.
干预措施: confirmation of endotracheal tube positioning (Device)
结局指标
主要结局
Time taken by the flow-sensor to indicate the correct or not correct positioning of the tracheal tube, compared with time taken by the colorimetric capnometer.
时间窗: Immediately after the tracheal intubation attempt
To compare the time (expressed in seconds) taken by the flow sensor and the colorimetric capnometer, respectively, to indicate the correct or not correct positioning of the tube in newborns undergoing tracheal intubation procedure during neonatal intensive care unit admission. The aim is to evaluate if the flow sensor of the mechanical ventilator can discriminate with similar or faster times the correct positioning of the tube during the tracheal intubation procedure compared to the colorimetric capnometer.
Number of mechanical breaths necessary to indicate the correct or not correct positioning of the tracheal tube by evaluating the flow-sensor curves and the color change of the colorimetric capnometer.
时间窗: Immediately after the tracheal intubation attempt
To compare the minimal number of mechanical breaths provided by the ventilator to indicate the correct or not correct positioning of the tube in newborns undergoing tracheal intubation procedure, either evaluating the flow-sensor curves and the color change of the colorimetric capnometer, respectively. The aim is to evaluate if the flow sensor of the mechanical ventilator can indicate with lower or equal number of mechanical breaths the correct positioning of the tube during tracheal intubation compared to the colorimetric capnometer.
次要结局
未报告次要终点
研究者
Paolo Biban
Director, Neonatal Intensive Care Unit, Dept. Neonatal and Pediatric Critical Care
Azienda Ospedaliera Universitaria Integrata Verona
