跳至主要内容
临床试验/NCT05830578
NCT05830578尚未招募不适用

Clinical Validation of the Electrocardiogram Measurement and Atrial Fibrillation Detection Function of a Wearable Smartwatch

Taipei Medical University Hospital0 个研究点目标入组 602 人开始时间: 2023年12月1日最近更新:
适应症

试验速览

阶段
不适用
状态
尚未招募
入组人数
602
主要终点
Sensitivity and specificity of the ECG App algorithm

研究概览

简要总结

Atrial fibrillation (AFib) is a common type of cardiac arrhythmia in clinical practice, affecting millions of people worldwide. Early detection and treatment of atrial fibrillation are crucial in preventing serious complications such as stroke and heart failure. In recent years, with the flourishing development of wearable devices and mobile technology, electrocardiogram (ECG) measurement applications embedded in smartwatches have gradually become a non-invasive and convenient method for heart rate monitoring. However, the accuracy of these devices has not yet been fully determined. This study aims to verify the ECG measurement and atrial fibrillation detection function of the ASUS Blood Pressure Monitor/Oximeter/ECG Monitor. The accuracy of the ECG application in detecting atrial fibrillation and measuring ECG will be evaluated by comparison with standard 12-lead ECGs.

详细描述

This study plans to recruit 602 adults over the age of 22. All participants will undergo heart rate measurements using both the smartwatch ECG application and the 12-lead ECG. The heart rate measurement using the ECG application will be operated by the participants themselves under the guidance of the testing personnel. The 12-lead ECG will be operated by trained medical professionals. The results of the heart rate measurement from both devices will be recorded synchronously, and the consistency of heart rate interpretation and ECG waveforms between the smartwatch ECG application and the 12-lead ECG measurement will be compared to verify the accuracy of the ECG application.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
22 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Willing and able to participate in the study procedures as described in the consent form
  • Individuals who are 22 years of age and older at time of screening
  • Able to communicate effectively with and follow instructions from the study staff
  • Subjects enrolled into Cohort 1 must have no known medical history of AFib and in normal sinus rhythm at the time of screening.
  • Subjects enrolled into Cohort 2 must have a known diagnosis of persistent or permanent AFib and be in AFib at the time of screening.

排除标准

  • Physical disability that precludes safe and adequate testing
  • Mental impairment as determined by the Investigator
  • Pregnant women at the time of the screening visit.
  • Subjects with any Medical History, Physical exam, vital sign or any other study procedure finding/assessment that in the opinion of the investigator could compromise subject safety during study participation or interfere with the study integrity and/or the accurate assessment of the study objectives. This includes patients with known untreated medical conditions that are considered clinically significant by the Investigator, such as but not limited to significant anemia, important electrolyte imbalance and untreated or uncontrolled thyroid disease. Physical exam limitations include but not limited to open wound(s) on the wrist and forearm where the subject will be wearing the watch.
  • Any history of wrist surgery with scarring in the area of the sensor location on the wrist where the subject will be wearing the watch.
  • Vital signs measurement, medical history, or physical exam finding that makes the subject inappropriate for participation according to the Investigator.
  • Tattoos or moles in the area of the sensor location on the wrist where the subject will be wearing the watch.
  • Skin conditions on either wrist that would preclude subject from wearing watch on either wrist. Severe symptomatic (or active) overly dry/injured skin, skin disorders, or allergic skin reactions such as eczema, rosacea, impetigo, dermatomyositis or allergic contact dermatitis on wrist and locations where the electrodes will be placed (e.g. chest, forearms, stomach), as determined by the investigator.
  • Known allergy or significant sensitivity to medical adhesives, isopropyl alcohol, or ECG patch.
  • Clinically significant hand tremors as judged by the Investigator.
  • Subjects with implanted cardiac devices such as a Pacemaker or an automated Implantable Cardioverter - Defibrillator.

结局指标

主要结局

Sensitivity and specificity of the ECG App algorithm

时间窗: One day visit

Sensitivity and specificity of the ECG App algorithm in detecting AFib compared with physician-adjudicated 12-lead ECG will be calculated.

次要结局

未报告次要终点

研究者

申办方类型
Other
责任方
Sponsor

相似试验

Clinical Validation of the Electrocardiogram... | 临床试验