临床试验/NCT07530731
NCT07530731已完成不适用

Comparison of NixOS-based Neuro-Ophtalmic Testing Platform Versus Windows and Ubuntu-based Platforms

Bulbitech1 个研究点 分布在 1 个国家目标入组 36 人开始时间: 2026年4月1日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
36
试验地点
1
主要终点
No statistical difference between acuracy of horizontal pro-saccades when comparing the three platforms

研究概览

简要总结

The purpose of the study is to verify that BulbiCAM®, a video-oculography medical device, gives similar results regardless of the Operating System (OS) used. The study compares three platforms: Windows, Linux Ubuntu, and Linux NixOS.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Crossover
主要目的
Other
盲法
None

入排标准

年龄范围
18 Years 至 75 Years(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Healthy adult age 18-75 years inclusive
  • Willing and able to provide informed consent
  • Able to understand and follow instructions (Norwegian or English)
  • Able to complete testing procedures

排除标准

  • Ophthalmic:
  • Visual disturbance and blindness of any cause, except for corrected refractive errors
  • Visual acuity worse than 1.0 logMAR in the tested eye (unable to focus on test stimuli)
  • History of eye surgery in the last 6 months
  • Visible abnormalities of the eye (subconjunctival hemorrhages, inflammation, deformed pupils)
  • Significant eyelid ptosis that would interfere with testing
  • Pupils unable to respond normally to dilation/contraction due to nerve damage, mechanical damage, or medication
  • Neurological/Medical:
  • Physical or psychiatric disease which may disturb the measuring procedure
  • Unable to tolerate testing or maintain attention for ~80 minutes with rest periods
  • Unable to understand and follow instructions
  • Photosensitive epilepsy
  • Known alcoholic or drug dependency
  • Any condition that, in the investigator's judgment, would make the subject unsuitable for the study

研究组 & 干预措施

Ubuntu - Windows - NixOS

Experimental

order of tested platforms

干预措施: administration of 15 neuro-ophtalmological tests per platform by video-oculography (Device)

Windows - Ubuntu - NixOS

Experimental

order of tested platforms

干预措施: administration of 15 neuro-ophtalmological tests per platform by video-oculography (Device)

Windiws - NixOS - Ubuntu

Experimental

order of tested platforms

干预措施: administration of 15 neuro-ophtalmological tests per platform by video-oculography (Device)

NixOS - Ubuntu - Windows

Experimental

order of tested platforms

干预措施: administration of 15 neuro-ophtalmological tests per platform by video-oculography (Device)

NixOS - Windows - Ubuntu

Experimental

order of tested platforms

干预措施: administration of 15 neuro-ophtalmological tests per platform by video-oculography (Device)

Ubuntu-NixOS-Windows

Experimental

order of tested platforms

干预措施: administration of 15 neuro-ophtalmological tests per platform by video-oculography (Device)

结局指标

主要结局

No statistical difference between acuracy of horizontal pro-saccades when comparing the three platforms

时间窗: single visit at enrollment

unit: %

No statistical difference between peak velocity of horizontal pro-saccades when comparing the three platforms

时间窗: Single visit at enrollment

unit: degree/s

No statistical difference between acuracy of multidirectional pro-saccades when comparing the three platforms

时间窗: Single visit at enrollment

Unit: %

No statistical difference between peak velocity of multidirectional pro-saccades when comparing the three platforms

时间窗: Single visit at enrollment

unit: degree/s

No statistical difference between contrast sensitivity results when comparing the three platforms

时间窗: Single visit at enrollment

unit: Pelli-Robson

No statistical difference between variable strobe test results when comparing the three platforms

时间窗: Single visit at enrollment

unit: Pelli-Robson

No statistical difference between strobe contrast sensitivity results when comparing the three platforms

时间窗: Single visit at enrollment

unit: Pelli-Robson

No statistical difference between RAPD test result when comparing the three platforms

时间窗: Single visit at enrollment

no unit

No statistical difference between ptosis MRD1 and 2 when comparing the three platforms

时间窗: Single visit at enrollment

unit mm

No statistical difference between duration of fixation when comparing the three platforms

时间窗: Single visit at enrollment

Unit: ms

No statistical difference between the amplitude of intrusive saccades when comparing the three platforms

时间窗: Single visit at enrollment

Unit: degree

No statistical difference between the peak velocity of intrusive saccades when comparing the three platforms

时间窗: Single visit at enrollment

unit: degree/s

No statistical difference between the mean fixation time when comparing the three platforms

时间窗: Single visit at enrollment

unit: ms

No statistical difference between mean pupil size when comparing the three platforms

时间窗: Single visit at enrollment

unit: mm

No statistical difference between anti-saccade amplitudes when comparing the three platforms

时间窗: Single visit at enrollment

unit: degree

No statistical difference between anti-saccades peak velocity when comparing the three platforms

时间窗: Single visit at enrollment

unit: degree/s

No statistical difference between memory saccades accuracy when comparing the three platforms

时间窗: Single visit at enrollment

unit: %

No statistical difference between number of errors in memory saccades task when comparing the three platforms

时间窗: Single visit at enrollment

no unit

No statistical difference between memory saccades latencies when comparing Ubuntu and NixOS platforms

时间窗: Single visit at enrollment

unit: ms

No statistical difference between anti saccades latencies when comparing Ubuntu and NixOS platforms

时间窗: Single visit at enrollment

unit: ms

No statistical difference between saccadic reaction time in visual field test when comparing Ubuntu and NixOS platforms

时间窗: Single visit at enrollment

unit: ms

No statistical difference between pupil area when comparing Ubuntu and NixOS platforms

时间窗: Single visit at enrollment

unit: mm\*s

No statistical difference between pupil diameters when comparing Ubuntu and NixOS platforms

时间窗: Single visit at enrollment

unit: mm

No statistical difference between pupil contraction latencies when comparing Ubuntu and NixOS platforms

时间窗: Single visit at enrollment

unit: ms

No statistical difference between peak velocity of pupil dilation when comparing Ubuntu and NixOS platforms

时间窗: Single visit at enrollment

unit: mm/s

No statistical difference between multidirectional pro-saccade latencies when comparing Ubuntu and NixOS platforms

时间窗: Single visit at enrollment

unit: ms

No statistical difference between horizontal pro-saccades latencies when comparing Ubuntu and NixOS platforms

时间窗: Single visit at enrollment

unit: ms

No statistical difference between smooth pursuit gain when comparing Ubuntu and NixOS platforms

时间窗: Single visit at enrollment

unit: %

No statistical difference between the calibration ratio when comparing the three platforms

时间窗: Single visit at enrollment

no unit

No statistical difference between pursuit acuity results when comparing Windows versus NixOS platforms

时间窗: Single visit at enrollment

unit: LogMAR

次要结局

  • No statistical difference between pupil areas when comparing Windows versus Ubuntu using the corrective formulas for signal processing enhancement, modified stimulus and timing calibration(Single visit at enrollment)
  • No statistical difference between pupil diameters when comparing Windows versus Ubuntu using the corrective formulas for signal processing enhancement, modified stimulus and timing calibration(Single visit at enrollment)
  • No statistical difference between latency of pupil contraction when comparing Windows versus Ubuntu using the corrective formulas for signal processing enhancement, modified stimulus and timing calibration(Single visit at enrollment)
  • No statistical difference between smooth pursuit gain when comparing Windows versus Ubuntu using the corrective formulas for timing calibration(Single visit at enrollment)
  • No statistical difference between horizontal pro-saccade latency when comparing Windows versus Ubuntu using the corrective formulas for timing calibration(Single visit at enrollment)
  • No statistical difference between multidirectional pro-saccade latencies when comparing Windows versus Ubuntu using the corrective formulas for timing calibration.(Single visit at enrollment)
  • No statistical difference between saccadic reaction times in visual field test when comparing Windows versus Ubuntu using the corrective formulas for timing calibration.(Single visit at enrollment)
  • No statistical difference between anti-saccade latencies when comparing Windows versus Ubuntu using the corrective formulas for timing calibration(Single visit at enrollment)
  • No statistical difference between memory saccades latencies when comparing Windows versus Ubuntu using the corrective formulas for timing calibration(Single visit at enrollment)
  • No statistical difference between peak velocity of pupil dilation when comparing Windows versus Ubuntu using the corrective formulas for signal processing enhancement and modified stimulus(Single visit at enrollment)
  • No statistical difference between pupil area when comparing Windows versus NixOS using the corrective formulas for signal processing enhancement, modified stimulus and timing calibration(Single visit at enrollment)
  • No statistical difference between pupil diameters when comparing Windows versus NixOS using the corrective formulas for signal processing enhancement, modified stimulus and(Single visit at enrollment)
  • No statistical difference between latencies of pupil contraction when comparing Windows versus NixOS using the corrective formulas for signal processing enhancement, modified stimulus and timing calibration(Single visit at enrollment)
  • No statistical difference between smooth pursuit gain when comparing Windows versus NixOS using the corrective formulas for timing calibration(Single visit at enrollment)
  • No statistical difference between horizontal pro-saccades latencies when comparing Windows versus NixOS using the corrective formulas for timing calibration(Single visit at enrollment)
  • No statistical difference between multidirectional pro-saccades latencies when comparing Windows versus NixOS using the corrective formulas for timing calibration(Single visit at enrollment)
  • No statistical difference between saccadic reaction time fo visual field test when comparing Windows versus NixOS using the corrective formulas for timing calibration(Single visit at enrollment)
  • No statistical difference between anti-saccades latencies when comparing Windows versus NixOS using the corrective formulas for timing calibration(Single visit at enrollment)
  • No statistical difference between memory saccades latencies when comparing Windows versus NixOS using the corrective formulas for timing calibration(Single visit at enrollment)
  • No statistical difference between peak velocity of the pupil when comparing Windows versus NixOS using the corrective formulas for signal processing enhancement and modified stimulus(Single visit at enrollment)