Performance Validation of a Non-contact Breathing Frequency Monitor (Model BM07)

ResMed1 个研究点 分布在 1 个国家目标入组 24 人开始时间: 2010年10月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
ResMed
入组人数
24
试验地点
1
主要终点
Respiration Rate Accuracy - Mean Difference in Breaths Per Minute Between the BM07 5GHz Sensor and SomnoScreen Devices

研究概览

简要总结

The purpose of this study is to compare the accuracy of the BiancaMed BM07 breathing frequency indicator device with the respiration rate measured by a SomnoScreen RC Easy device on 20+ human volunteers. This is being carried out in support of a 510(k) submission.

详细描述

The purpose of this study is to compare the accuracy of the BiancaMed BM07 breathing frequency indicator device with the respiration rate measured by a SomnoScreen RC Easy device on 20+ human volunteers. This is being carried out in support of a 510(k) submission.

  • The BM07 is a non-contact device, using radio frequency (RF) range measurement to detect the chest wall motion, and thus the respiration rate. The BM07 is a CE Class IIa cleared device.

  • The SomnoScreen RC Easy is both a CE and FDA (K060708) cleared device, used as an overnight, portable sleep recorder. One of its inputs is a pair of chest respiration effort bands that can be used to record the breathing rate of the subject.

  • The human volunteers are selected to be a range of age, body mass index (BMI) and gender so as to cover a broad cross section of the population. BMI, gender and age are features that might affect the performance of the RF reflected signal:

  • High BMI or low BMI could affect the skin RF reflectivity.

  • High BMI may mask chest wall movement during breathing.

  • The thinner skin often associated with increasing age may affect RF reflectivity.

  • Female breast tissue may affect RF reflectivity or mask chest wall movement.

The target performance is an agreement between the BM07 and the Somnoscreen of +/- 5 breaths per minute. This target has been selected based upon a literature review, in particular:

  • Lim et al: Respiratory Rate Measurement in Adults - How Reliable is it?
  • Droitcour PhD Thesis: Non-Contact Measurement of Heart and Respiration Rates with a Single-Chip Microwave Doppler Radar.

研究设计

研究类型
Observational
观察模型
Case Only
时间视角
Cross Sectional

入排标准

年龄范围
10 Years 至 —(Child, Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • volunteer

排除标准

  • 未提供

结局指标

主要结局

Respiration Rate Accuracy - Mean Difference in Breaths Per Minute Between the BM07 5GHz Sensor and SomnoScreen Devices

时间窗: Recording duration of 2 minutes; measurement made after 45 seconds settling time, for at least 60 seconds

The respiration rate measured contemporaneously by the BM07 5GHz sensor and the reference device, SomnoScreen, are compared on a second-by-second basis, over the recording duration. The difference between the two respiration rate measurements (the delta) for all 24 subjects' recordings, in all positions, are plotted on a normal distribution chart, giving 10,542 data points. The distribution chart reveals the mean difference between the two devices' measurements and the Standard Deviation (SD). The measure of success is if the 95% (ie 2SD) of the difference in the respiration rate indicated by the two devices (the delta) is less than 5 breaths per minute. This measure is derived from the peer review paper by Lim et al, cited in the References section

次要结局

未报告次要终点