Agreement and Precision Study of the Nidek Mirante OCT Compared to the Optovue RTVue XR Avanti OCT and SLO Image Comparison of the Nidek Mirante and the OPTOS P200DTx

Nidek Co. LTD.2 个研究点 分布在 1 个国家目标入组 170 人开始时间: 2020年1月25日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
170
试验地点
2
主要终点
Agreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT

研究概览

简要总结

This study is a prospective comparative, randomized, single center study to gather agreement and precision of the Nidek Mirante OCT with SLO and Anterior Segment Imaging capabilities in comparison to the Optovue RTVue XR Avanti OCT and Optos P200DTx in normal subjects, subjects with glaucoma, subjects with retinal disease and subjects with corneal disease.

详细描述

This is a prospective comparative clinical study to be conducted at one clinical site located in the United States. Three Nidek Mirante devices, three Optovue RTVue XR Avanti devices and one OPTOS P200DTx will be used. For agreement and precision analysis of OCT scans and agreement analysis of image quality of OCT ACA scans, each Nidek Mirante device will be paired with one of the Optovue RTVue XR Avanti devices. One device operator will be assigned to each device pair to create three distinct operator/device configurations.

For agreement analysis of SLO image quality, one OPTOS P200DTx will be included in one of three operator/device configurations.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
22 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Normal Group:
  • Male or female subjects from 22 years of age or older who have full legal capacity to volunteer on the date the informed consent is signed;
  • Subjects who can follow the instructions by the clinical staff at the clinical site, and can attend examinations on the scheduled examination date;
  • Subjects who agree to participate in the study;
  • Subjects with normal eye examinations (without pathology other than cataract) in both eyes on the date of the study visit as observed with a +90 Diopter Fundus Lens;
  • Subjects with current best-spectacle-corrected visual acuity (BSCVA) of 20/40 or better in both eyes on the date the study visit.

排除标准

  • Normal Group:
  • Subjects unable to tolerate ophthalmic imaging;
  • Subjects with ocular media not sufficiently clear to obtain acceptable OCT images in either eye;
  • Subjects with any current ocular pathology other than cataract in either eye, as determined by self-report and confirmed by the investigator assessment and/or confirmed by the investigator assessment at the study visit;
  • Subjects with (or history of) leukemia, dementia or multiple sclerosis;
  • Subject has a condition or will be in a situation which the investigator feels may put the subject at significant risk, may confound the study results, or may interfere significantly with the subject's participation in the study.
  • Inclusion Criteria - Glaucoma Group:
  • Male or female subjects from 22 years of age or older who have full legal capacity to volunteer on the date the informed consent was signed;
  • Subjects who can follow the instructions by the clinical staff at the clinical site, and can attend examinations on the scheduled examination date;
  • Subjects who agree to participate in the study;
  • Subjects who have been diagnosed with glaucoma in the glaucoma study eye(s), with optic nerve damage as evidenced by either of the following optic disc or RNFL structural abnormalities observed via fundus exam during the study visit:
  • Diffuse thinning, focal narrowing, or notching of the optic disc rim, especially at the inferior or superior poles with or without disc hemorrhage; or
  • Optic disc neural rim asymmetry of the two eyes consistent with loss of neural tissue;
  • Subjects with a current BSCVA of 20/40 or better in the glaucoma study eye(s) on the date of the visit;
  • History of a reliable Humphrey Field Analyzer (HFA) visual field (24-2 SITA Standard, white on white) measured on the day of study visit 1, or within the previous six (6) months from study visit 1 with reliable results, defined as fixation losses < 20%, or false positives < 33%, or false negatives < 33% in the glaucoma study eye(s) that is consistent with glaucomatous optic nerve damage in glaucoma study eye(s) based on at least one of the following findings:
  • On pattern deviation (PD), there exists a cluster of 3 or more points in an expected location of the visual field depressed below the 5% level, at least 1 of which is depressed below the 1% level; or
  • Glaucoma hemi-field test "outside normal limits."
  • Exclusion Criteria - Glaucoma Group:
  • Subjects unable to tolerate ophthalmic imaging;
  • Subjects with ocular media not sufficiently clear to obtain acceptable OCT images in the glaucoma study eye(s);
  • Subjects with retinal disease in the glaucoma study eye(s), as determined by self-report and confirmed by the investigator assessment and/or confirmed by the investigator assessment at the study visit;
  • Subjects with (or history of) leukemia, dementia or multiple sclerosis;
  • Subject has a condition or will be in a situation which the investigator feels may put the subject at significant risk, may confound the study results, or may interfere significantly with the subject's participation in the study.
  • Inclusion Criteria - Retinal Disease Group:
  • Male or female subjects 22 years of age or older who have full legal capacity to volunteer on the date the informed consent is signed;
  • Subjects who can follow the instructions by the clinical staff at the clinical site, and can attend examinations on the scheduled examination date;
  • Subjects who agree to participate in the study;
  • Subjects with a current BSCVA of 20/400 or better in the retinal disease study eye(s) at the study visit;
  • Subjects diagnosed with retinal pathology including but not limited to: Non Exudative Macular Degeneration (dry AMD), Diabetic Macular Edema, Macular Hole, Epiretinal Membrane, Cystoid Macular Edema, or other retinal disease in the study eye(s) as confirmed within the past six (6) months, who exhibit structural lesions in the study eye.
  • Exclusion Criteria - Retinal Disease Group
  • Subjects unable to tolerate ophthalmic imaging;
  • Subjects with ocular media not sufficiently clear to obtain acceptable OCT images in the study eye;
  • Subjects with glaucoma in the retinal disease study eye(s), as determined by self-report and confirmed by the investigator assessment and/or confirmed by the investigator assessment at the study visit;
  • Subjects with (or history of) leukemia, dementia or multiple sclerosis;
  • Subject has a condition or will be in a situation which the investigator feels may put the subject at significant risk, may confound the study results, or may interfere significantly with the subject's participation in the study.
  • Inclusion Criteria - Corneal Group
  • Male or female subjects 22 years of age or older who have full legal capacity to volunteer on the date the informed consent is signed;
  • Subjects who can follow the instructions by the clinical staff at the clinical site, and can attend examinations on the scheduled examination date;
  • Subjects who agree to participate in the study;
  • Subjects with a current BSCVA of 20/400 or better in the study eye(s) at the study visit;
  • Subjects with corneal pathologies including but not limited to: Post status LASIK surgery, Keratoconus, or other corneal dystrophies or degenerations in the study eye(s) as confirmed within the past six (6) months.
  • Exclusion Criteria - Corneal Group
  • Subjects unable to tolerate ophthalmic imaging;
  • Subjects with ocular media not sufficiently clear to obtain acceptable OCT images in the study eye;
  • Subjects with (or history of) leukemia, dementia or multiple sclerosis;
  • Subject has a condition or will be in a situation which the investigator feels may put the subject at significant risk, may confound the study results, or may interfere significantly with the subject's participation in the study.

结局指标

主要结局

Agreement of Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT

时间窗: 1 day

Agreement of the measured Macular thickness (μm) between the test device and the predicate device for the OCT function of NIDEK Mirante.

Agreement of Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT

时间窗: 1 day

Agreement of the measured Retinal nerve fiber layer thickness around Optic nerve head between the test device and the predicate device for the OCT function of NIDEK Mirante.

Agreement of Corneal Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT

时间窗: 1 day

Agreement of the measured Corneal thickness between the test device and the predicate device for the OCT function of NIDEK Mirante.

Agreement of SLO Image Quality (Likert Scale) for Nidek Mirante and OPTOS P200DTx

时间窗: 1 day

Agreement of the SLO Image quality between the test device and the predicate device for the SLO function of NIDEK Mirante. The image quality of is graded Clinical Utility and Overall Utility for each image patterns (SLO Color Fundus, SLO B-FAF and SLO G-FAF) on a Likert scale. The grading used the 0-3 point scale, and higher scores mean a better image quality.

Agreement of the ACA Image Quality (Likert Scale) for Nidek Mirante and Optovue RTVue XR Avanti OCT

时间窗: 1 day

Agreement of the ACA Image quality between the test device and the predicate device for the OCT function of NIDEK Mirante. The image quality of is graded Clinical Utility and Overall Utility on a Likert scale. The grading used the 0-3 point scale, and higher scores mean a better image quality.

Precision of the Macular Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT

时间窗: 1 day

Precision (repeatability) of the measured Full retinal thickness at Macula for the OCT function of the NIDEK TONOREF III.

Precision of the Retinal Nerve Fiber Layer Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT

时间窗: 1 day

Precision (repeatability) of the measured Retinal nerve fiber layer thickness around Optic nerve head at Macula for the OCT function of the NIDEK TONOREF III.

Precision of the Corneal Thickness (μm) Measurement for Nidek Mirante and Optovue RTVue XR Avanti OCT

时间窗: 1 day

Precision (repeatability) of the measured Thickness of the Cornea for the OCT function of the NIDEK TONOREF III.

次要结局

  • Adverse Events(1 day)

研究者

发起方
Nidek Co. LTD.