Smartwatches for Detection of Atrial Fibrillation (AFib) in Secondary Prevention of Cryptogenic Stroke - WATCH AFib A Prospective, Intraindividual-controlled, Multicenter Clinical Study

Technical University of Munich2 个研究点 分布在 1 个国家目标入组 400 人开始时间: 2024年10月15日最近更新:
适应症

试验速览

阶段
不适用
状态
招募中
发起方
入组人数
400
试验地点
2
主要终点
Sensitivity and Specificity of AFib Detection

研究概览

简要总结

Scientific Background: In secondary prevention of ischemic stroke, detection of atrial fibrillation (AFib) and subsequent anticoagulation therapy reduce the risk of recurrent stroke by approximately 60%. Prolonged electrocardiogram (ECG) monitoring up to 6 months significantly increases detection of AFib in cryptogenic stroke. Wearables like smartwatches have recently been shown to adequately detect AFib in the general population. Thus, prolonged ECG monitoring after cryptogenic ischemic stroke or transient ischemic attack (TIA) using a smartwatch could lead to a reduction of recurrent stroke by prompting adequate anticoagulation therapy and may constitute a cost-effective, non-invasive, and broadly-available alternative to the current standard of care.

Hypothesis: The investigators hypothesize that AFib detection via smartwatch in patients with cryptogenic TIA or ischemic stroke is accurate compared to an implantable event recorder.

Methods: The investigators introduce a prospective, intraindividual-controlled, multicentre clinical study in patients with cryptogenic ischemic stroke or TIA. In addition to an implanted event recorder as indicated by clinical standard, included patients receive a smartwatch for detection of AFib. ECG-data from smartwatches will be continuously monitored by two independent cardiologists. As soon as AFib is confirmed, a doctoral appointment is set to evaluate start of anticoagulation. The follow-up period will be six months. The study consists of four study visits: a baseline visit, two phone visits at one and three months, and an end of trial visit at six months.

Primary Objective: To compare smartwatch and event recorder based analysis for sensitivity and specificity of AFib detection per patient after six months

详细描述

Ischemic stroke is worldwide the second most frequent cause of death and disability and has a lifetime risk of approximately 25%. In the secondary prevention of ischemic stroke, detection of atrial fibrillation (AFib) and subsequent anticoagulation therapy reduce the risk of recurrent stroke by approximately 60%. Cryptogenic stroke is defined as ischemic stroke for which no probable cause is found despite a full standard evaluation and comprises 25% of all ischemic strokes. Prolonged electrocardiogram (ECG) monitoring for 30 days to 6 months significantly increases the detection of occult paroxysmal AFib in cryptogenic stroke, which is present in 9 - 16%. Thus, prolonged ECG monitoring is likely to lead to a reduction of recurrent stroke by prompting adequate anticoagulation therapy.

Still, prolonged AFib screening after stroke is currently suboptimal due to a limitation of resources, loss to follow-up, invasiveness of procedures, and costs. Wearables like smartwatches have recently been shown to adequately detect AFib in the general population. Sensitivity and specificity of AFib detection compared to ECG-based diagnosis is high (93-98% and 90-98%, respectively). Compared to implantable Event Recorders, smartwatch based AFib diagnosis in the general population exhibits a sensitivity of 97.5% for AFib episodes >1h and a sensitivity of 100% for AFib detection per patient.

The investigators hypothesize that AFib detection via smartwatch in patients suffering from cryptogenic transient ischemic attack (TIA) or ischemic stroke is accurate for AFib detection compared to an implantable Event Recorder and might therefore be a non-invasive, cost-effective, widely available alternative, which could potentially change the current standard of post-stroke care.

The primary objective is to assess whether AFib detection via smartwatch in patients suffering from cryptogenic TIA/ ischemic stroke is accurate in comparison to implantable Event Recorders. This will be assessed by sensitivity and specificity of AFib detection per patient after six months. The investigators aim on a comparison of smartwatch based, continuous, automated, cardiologist supervised rhythm analysis of photoplethysmography (PPG)- signal and patient activated 1-lead ECG with Event Recorder based, continuous, automated ECG rhythm analysis.

This is a prospective, intraindividual-controlled, multicenter clinical study.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
40 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Implanted Event Recorder with telemedicinal function (e.g., Biotronik BioMon 3m ProMRI HomeMon or similar)
  • Cryptogenic stroke (within the last six month) after full standard evaluation:
  • Stroke detected by computed tomography (CT) or magnetic resonance imaging (MRI) that is not lacunar (lacunar is defined as a subcortical infarct in the distribution of the small, penetrating cerebral arteries whose largest dimension is ≤1.5 cm on CT or ≤2.0 cm on MRI diffusion images)
  • Absence of extracranial or intracranial atherosclerosis causing ≥50 percent luminal stenosis of the artery supplying the area of ischemia
  • No major-risk cardioembolic source of embolism (i.e., no permanent or paroxysmal atrial fibrillation, sustained atrial flutter, intracardiac thrombus, prosthetic cardiac valve, atrial myxoma or other cardiac tumors, high-grade mitral valve stenosis, recent (within four weeks) myocardial infarction, left ventricular ejection fraction <30 percent, valvular vegetations, or infective endocarditis)
  • No other specific cause of stroke identified (e.g., arteritis, dissection, migraine, vasospasm, drug abuse)
  • No paroxysmal atrial fibrillation in 72h of in-hospital ECG-monitoring, including at least one Holter- ECG for 24 hours.
  • Cryptogenic TIA with definite cortical syndrome (aphasia, neglect or homonymous hemianopia) (within the last six month) after full standard evaluation (see above)
  • Age: ≥40 yrs.
  • At least one of the following risk factors:
  • CHA2DS2VASc score ≥4
  • Atrial runs
  • Left atrial size > 45mm
  • Left atrial appendage flow ≤ 0.2m/s
  • No contraindication for anticoagulant therapy after acute phase of stroke
  • Written informed consent by patient or authorized caregiver

排除标准

  • Patient is not able to perform 1-lead ECG recording with smartwatch
  • Patient possesses no smartphone (iOS version ≥10.0 or Android)
  • Implanted pacemaker or cardioverter defibrillator (ICD)
  • Pregnancy and breastfeeding period

结局指标

主要结局

Sensitivity and Specificity of AFib Detection

时间窗: From baseline visit (V0) to the last follow-up at visit (V3) at six months

To estimate and compare sensitivity and specificity for AFib detection per patient after six months (Comparison of smartwatch based, continuous, automated, cardiologist supervised rhythm analysis of photoplethysmography (PPG)- signal and patient activated 1-lead ECG with Event Recorder based, continuous, automated ECG rhythm analysis).

次要结局

  • Sensitivity and specificity for the detection of any AFib episode(From baseline visit (V0) to the last follow-up at visit (V3) at six months)
  • Sensitivity for the detection of AFib episodes >1 hour(From baseline visit (V0) to the last follow-up at visit (V3) at six months)
  • Specificity for episodes of sinus rhythms >1 hour(From baseline visit (V0) to the last follow-up at visit (V3) at six months)
  • Stroke and TIA recurrence within six months(From baseline visit (V0) to the last follow-up at visit (V3) at six months)
  • Subgroup analysis: accuracy in severely affected patients (i.e., National Institute of Health Stroke Scale (NIHSS) ≥8)(From baseline visit (V0) to the last follow-up at visit (V3) at six months)
  • Exploratory endpoint: count of AFib diagnoses(From baseline visit (V0) to the last follow-up at visit (V3) at six months)
  • To estimate and compare the positive and negative predictive values for AFib detection per patient(From baseline visit (V0) to the last follow-up at visit (V3) at six months)
  • Sensitivity and specificity for AFib detection of the automated PPG-signal rhythm analysis(From baseline visit (V0) to the last follow-up at visit (V3) at six months)
  • Exploratory endpoint: Acceptance and practicability of smartwatches for AFib detection(From baseline visit (V0) to the last follow-up at visit (V3) at six months)
  • Exploratory endpoint: AFib burden/ patient(From baseline visit (V0) to the last follow-up at visit (V3) at six months)
  • Exploratory endpoint: relevance of AFib risk factors(From baseline visit (V0) to the last follow-up at visit (V3) at six months)
  • Exploratory endpoint: AFib detection rates after one month and after six months and time to confirmed AFib diagnosis(From baseline visit (V0) to the last follow-up at visit (V3) at six months)
  • Sensitivity and specificity of AFib detection per recorded/ per analyzable time (i.e., intervals in which the watch is actually worn/records an analyzable signal; per-protocol analysis)(From baseline visit (V0) to the last follow-up at visit (V3) at six months)