The Clinical Performance Validation of Electronic Thermometer and Infrared Thermometers
- Conditions
- FebrileHealthy
- Interventions
- Device: Infrared Forehead Thermometer (GE-TF03)Device: Infrared Ear Thermometer (GE-TE06)Device: Electronic Thermometer (MT-B321FB)Device: Electronic Thermometer (MT-B231)
- Registration Number
- NCT06353490
- Lead Sponsor
- Geon Corporation
- Brief Summary
Geon Corporation develops ear thermometers and forehead thermometers that use infrared measurement technology to measure human body temperature. This study verify the accuracy and repeatability of investigational devices. Efficacy verification is carried out by comparing with the referent medical devices.
The main objectives of this study are:
1. To verify that the investigational devices and the reference medical devices have a reasonable deviation value.
2. To verify the consistency of repeated measurements of investigational devices.
- Detailed Description
This Study carried out in accordance with the ISO 80601-2-56: 2017, which included definitions for the age groups and sample sizes, the definition of fever, the flowchart for relevant procedures, and statistical methods for clinical performance.
All subjects were measured by the reference device (electronic thermometer MT-B231) and three investigational devices under test evaluation (infrared forehead thermometer GE-TF03, infrared ear thermometer GE-TE06, and electronic thermometer MT-B321FB). For each subject, the reference device measured one time and each investigational device measured three times respectively.
Recruitment & Eligibility
- Status
- ACTIVE_NOT_RECRUITING
- Sex
- All
- Target Recruitment
- 150
- Newborn to 80 years old adult
- Male or Female
- The subject or parent/guardian can talk to the researcher, understand the content of the study after researcher's explanation and agree to sign the subject's consent form, and is willing to accept the trial procedure.
- Currently have signs or symptoms of infection or inflammation at the temperature measurement site (axillary, forehead, ear canal)
- The forehead cannot be fully exposed to the environment for more than 15 minutes
- Using cooling blankets, fans, or placing ice cubes on the forehead
- There is a catheter placed in the ear
- Use of antipyretics (such as aspirin, ibuprofen), barbiturates, antipsychotics in the past 2 hours
- Received thyroxine, corticosteroids or immunotherapy in the past 7 days
- Subject has been diagnosed with cognitive impairment or has a history of dementia, mental illness, mental illness, pervasive developmental disorder, illicit drug use or alcoholism, etc.
- Pregnant women
- The researcher considers that the subject is not suitable for participating in this trial, such as taking any drugs that may change body temperature
- Participated in clinical trials of investigational drugs within the last 30 days
- The subject is unwilling to sign the subject's consent form
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- PARALLEL
- Arm && Interventions
Group Intervention Description Subjects: 0 up to 1 year Electronic Thermometer (MT-B231) Collecting temperatures using the "Geon" Digital, Infrared Ear and Infrared Forehead thermometer Subjects: Older than 1 year and younger than 5 years Infrared Forehead Thermometer (GE-TF03) Collecting temperatures using the "Geon" Digital, Infrared Ear and Infrared Forehead thermometer Subjects: 0 up to 1 year Infrared Forehead Thermometer (GE-TF03) Collecting temperatures using the "Geon" Digital, Infrared Ear and Infrared Forehead thermometer Subjects: 0 up to 1 year Electronic Thermometer (MT-B321FB) Collecting temperatures using the "Geon" Digital, Infrared Ear and Infrared Forehead thermometer Subjects: Older than 1 year and younger than 5 years Infrared Ear Thermometer (GE-TE06) Collecting temperatures using the "Geon" Digital, Infrared Ear and Infrared Forehead thermometer Subjects: 0 up to 1 year Infrared Ear Thermometer (GE-TE06) Collecting temperatures using the "Geon" Digital, Infrared Ear and Infrared Forehead thermometer Subjects: Older than 1 year and younger than 5 years Electronic Thermometer (MT-B321FB) Collecting temperatures using the "Geon" Digital, Infrared Ear and Infrared Forehead thermometer Subjects: Older than 1 year and younger than 5 years Electronic Thermometer (MT-B231) Collecting temperatures using the "Geon" Digital, Infrared Ear and Infrared Forehead thermometer Subjects: Older than 5 years Infrared Forehead Thermometer (GE-TF03) Collecting temperatures using the "Geon" Digital, Infrared Ear and Infrared Forehead thermometer Subjects: Older than 5 years Infrared Ear Thermometer (GE-TE06) Collecting temperatures using the "Geon" Digital, Infrared Ear and Infrared Forehead thermometer Subjects: Older than 5 years Electronic Thermometer (MT-B321FB) Collecting temperatures using the "Geon" Digital, Infrared Ear and Infrared Forehead thermometer Subjects: Older than 5 years Electronic Thermometer (MT-B231) Collecting temperatures using the "Geon" Digital, Infrared Ear and Infrared Forehead thermometer
- Primary Outcome Measures
Name Time Method Limits of Agreement Through study completion, an average of 2 years The magnitude of a potential disagreement between investigational thermometer and reference thermometer.
Clinical Bias Through study completion, an average of 2 years Mean Difference between investigational thermometer and reference thermometer.
Clinical Repeatability Through study completion, an average of 2 years Pooled standard deviation of triplicate measurements over the entire population of subject
- Secondary Outcome Measures
Name Time Method
Trial Locations
- Locations (2)
HsinChu Municipal MacKay Memorial Hospital
🇨🇳Hsinchu, Taiwan
Hsinchu MacKay Memorial Hospital
🇨🇳Hsinchu, Taiwan