Accuracy of pulse-oximetry under INTELLiVENT-ASV - Accuracy of SpO2 under i-ASV
Jichi Medical University School of Medicine0 个研究点目标入组 100 人开始时间: 2017年6月9日最近更新:
试验速览
- 阶段
- 未知
- 状态
- 已完成
- 发起方
- 入组人数
- 100
研究概览
简要总结
Mean SaO2 values, Nihon Kohden SpO2, and Masimo SpO2 were 95.7%, 96.4%, and 96.9%, respectively. The Nihon Kohden SpO2 were less biased than Masimo measurements; their precision was not significantly different. Nihon Kohden and Masimo SpO2 were not significantly different in the SaO2 < 94% group (P = 0.083). In the 94% <= SaO2 < 98% and SaO2 >= 98% groups, there were significant differences between the Nihon Kohden and Masimo SpO2 (P < 0.0001; P = 0.006; respectively).
研究设计
- 研究类型
- Observational
入排标准
- 年龄范围
- 20years-old 至 ot applicable(—)
- 性别
- All
入选标准
- 未提供
排除标准
- •1. patients who can not measure pulse oximeter with fingers. 2. in case of an unreliable pulsatile SpO2 curve 3. patients who are under 20 years old. 4. patients who can not obtain the consent. 5. patients who undergo extracorporeal membrane oxygenation. 6 patients if carboxyhemoglobin (COHb) levels were > 3% or methemoglobin (MetHb) > 1.5%
研究者
相似试验
已完成
不适用
Induced Hypoxia Study for Validation of SpO2 AccuracyHypoxiaNCT00788983Nonin Medical, Inc12
已完成
不适用
SpO2 Accuracy Validation of Pulse Oximetry Systems During Motion and Non-Motion ConditionsHypoxiaNCT02475668Nonin Medical, Inc13
已完成
不适用
SpO2 Accuracy Validation of the OxySoft Sensor Via Reference CO-Oximetry Motion StudyHypoxiaDesaturation of BloodNCT04559763Medtronic - MITG16
已完成
不适用
Vital USA SpO2 Accuracy Comparison to Arterial Blood CO-OximetryArterial DiseaseNCT04285411Vital USA, Inc.10
已完成
不适用
SpO2 Accuracy Low Saturation Validation of the OxySoftN SensorHypoxiaDesaturation of BloodNCT04811963Medtronic - MITG10