研究概览

简要总结

Mean SaO2 values, Nihon Kohden SpO2, and Masimo SpO2 were 95.7%, 96.4%, and 96.9%, respectively. The Nihon Kohden SpO2 were less biased than Masimo measurements; their precision was not significantly different. Nihon Kohden and Masimo SpO2 were not significantly different in the SaO2 < 94% group (P = 0.083). In the 94% <= SaO2 < 98% and SaO2 >= 98% groups, there were significant differences between the Nihon Kohden and Masimo SpO2 (P < 0.0001; P = 0.006; respectively).

研究设计

研究类型
Observational

入排标准

年龄范围
20years-old 至 ot applicable(—)
性别
All

入选标准

  • 未提供

排除标准

  • 1. patients who can not measure pulse oximeter with fingers. 2. in case of an unreliable pulsatile SpO2 curve 3. patients who are under 20 years old. 4. patients who can not obtain the consent. 5. patients who undergo extracorporeal membrane oxygenation. 6 patients if carboxyhemoglobin (COHb) levels were > 3% or methemoglobin (MetHb) > 1.5%