Determination of SpO2 and PR Accuracy Specifications at Rest Accuracy of Pulse Oximeters With Profound Hypoxia Pulse Oximeter Accuracy Evaluation Protocol

Nihon Kohden1 个研究点 分布在 1 个国家目标入组 13 人开始时间: 2021年3月24日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
Nihon Kohden
入组人数
13
试验地点
1
主要终点
Accuracy of Sensor Oxygen Saturation by Arms Calculation

研究概览

简要总结

The aim of this study is to determine the accuracy of devices called pulse oximeters, which measure blood oxygen by shining light through fingers, ears or other skin, without requiring blood sampling. Study will be used with patients at rest.

详细描述

This study is intended to evaluate performance of a new design pulse oximeter (test device) manufactured by Nihon Kohden Corporation sufficiently to support performance claims for an FDA 510K submission or ISO technical file. Specifically, SpO2 and pulse rate accuracy will be assessed for a Nihon Kohden OLV-4202 pulse oximeter with adult patients under a controlled setting of varying levels of inhaled oxygen concentration levels for patients at rest.

研究设计

研究类型
Observational
观察模型
Case Only
时间视角
Prospective

入排标准

年龄范围
18 Years 至 50 Years(Adult)
性别
All
接受健康志愿者

入选标准

  • Both male and female subjects who can give written informed consent
  • Healthy subjects capable of undergoing controlled hypoxemia to the levels outlined in the desaturation profile
  • Meeting the demographic requirements

排除标准

  • Age below 18 or over 50
  • Pregnant women
  • Significant arrhythmia
  • Blood pressure above 150 systolic or 90 diastolic
  • Carboxyhemoglobin levels over 3%
  • Subjects whom the investigator consider ineligible for the study

结局指标

主要结局

Accuracy of Sensor Oxygen Saturation by Arms Calculation

时间窗: 30 seconds

Accuracy will be determined by comparing the noninvasive blood oxygen saturation measurement of the pulse oximeter (SpO2i) to that obtained from a blood sample (SRi) and calculating the arithmetic root mean square (Arms) error value as follows: Arms =Square root (sum of n samples of ((SpO2i - SRi) squared) / n)

次要结局

  • Accuracy of Sensor Pulse Rate by Arms Calculation(30 seconds)

研究者

发起方
Nihon Kohden
申办方类型
Industry
责任方
Sponsor