Evaluation of Clarity Ultrasound Guidance for Prostate Therapy

Institute of Cancer Research, United Kingdom1 个研究点 分布在 1 个国家目标入组 61 人开始时间: 2015年5月15日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
61
试验地点
1
主要终点
Prostate position (measured in millimeters) measured using Clarity ultrasound and implanted fiducial markers

研究概览

简要总结

Radiotherapy (RT) is an important treatment for prostate cancer with over 10,000 men receiving RT per year in the UK. Prostate RT aims to deliver radiation dose to the prostate to kill cancer cells whilst minimizing the dose given to surrounding normal tissues, such as the bladder or the rectum. Radiation is delivered in a number of daily treatments which are called 'fractions'. The position of the prostate varies each day and during the radiation delivery. To accurately aim the radiation the prostate needs to be located before daily treatment. Clarity is an ultrasound image guidance system that locates the prostate and monitors its position during treatment. Ultrasound does not give a radiation dose, is non-invasive, provides fast imaging and can easily visualise soft tissues. Before Clarity is implemented the investigators wish to ensure that it can accurately locate and monitor the prostate by comparing Clarity with marker-based techniques in 24 patients. Patients receive markers as part of other trials which are currently recruiting at The Royal Marsden (for example, patients in the DELINEATE trial and the PACE trials receive implanted gold markers and patients in the PROSPARE trial have received electromagnetic markers). Patients who have received or will be receiving markers as part of other studies will be asked to also to take part in this study. This means that there will be no additional intervention for patients as a result of this study. Clarity will be considered a safe and effective imaging system if the mean and standard deviation of the differences between Clarity measurements and marker measurements, of prostate position, are less than 1mm. It is possible that markers will influence the Clarity measurements. To find out if this happens, the mean and standard deviation of Clarity measurements in a group of 24 patients without markers to the group with markers will be compared.

Primary Aim: To evaluate the accuracy of Clarity ultrasound image guidance system for prostate motion estimation during treatment (intra-fraction), by comparison with implanted marker-based image guidance techniques.

Primary Hypothesis: Clarity can be safely used to deliver radiotherapy using similar (within 1mm) treatment margins to those that would be employed if fiducial markers based image guidance techniques were used.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
Male
接受健康志愿者

入选标准

  • Two patient groups and two patient volunteer groups will be recruited:
  • Groups 1: Patient volunteers who have received radiotherapy for prostate cancer which included a planning CT scan, or who are currently receiving radiotherapy.
  • Groups 2 and 3:
  • Patients receiving radiotherapy for prostate cancer as part of a clinical trial or standard care which includes gold FM or EM insertion
  • Patients who have previously received RT for prostate cancer which included EM insertion.
  • Group 4: Patients who are receiving radiotherapy for prostate cancer and who are not receiving FMs. Attempts will be made to match Group 4 patients BMI and time receiving hormone therapy to patients in group 2 and 3 (see below, section 6.1 recruitment.

排除标准

  • Groups 1 to 4: Patients who have received prostatectomy.
  • Groups 2 and 4: Patients receiving less than 6 RT fractions.
  • Group 4 only: Patients who have received FMs or EMs.

结局指标

主要结局

Prostate position (measured in millimeters) measured using Clarity ultrasound and implanted fiducial markers

时间窗: Outcome measures will be assessed for each patient during radiotherapy, an average of five weeks.

Mean (and standard deviation) of the differences in prostate position determined using Clarity and that determined using fiducial marker based image guidance during radiotherapy delivery.

次要结局

未报告次要终点

研究者

发起方
Institute of Cancer Research, United Kingdom
申办方类型
Other