Effect Of Dialysate Flow Rate On Mass Transfer Coefficient - Area (KoA) In Dialyzers With Different Membrane Packing Densities

Vantive Health LLC2 个研究点 分布在 1 个国家目标入组 12 人开始时间: 2008年2月1日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
12
试验地点
2
主要终点
Effect of Increased Dialysate Flow Rate on KoA for Phosphorus Between 4 Dialyzers With Different Membrane Packing Densities.

研究概览

简要总结

The purposes of this study are to determine if the performance of a dialyzer depends on how tightly the hollow fiber membranes are packed in the housing of the dialyzer (the membrane packing density) and if that dependence is a function of the dialysate flow rate. The study will examine how efficiently three different sized molecules pass through a dialyzer membrane at different dialysate flow rates in dialyzers with different membrane packing densities. Transfer of urea, phosphorus and beta-2-microglobulin from blood to dialysate will be measured during clinical hemodialysis using four different dialyzers, each used at three different dialysate flow rates. The data derived from these measurements may provide insight into the importance of membrane packing density as a design parameter for hemodialyzers and if changing the membrane packing density might provide equivalent performance at a lower dialysate flow rate.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Crossover
主要目的
Treatment
盲法
None

入排标准

年龄范围
19 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Stable hemodialysis patients dialyzing through a native fistula or Gore-Tex graft. The access must be capable of delivering a stable blood flow of 400 ml/min.
  • Age older than 18 years.
  • Fluid removal requirement less than 3 liters per treatment.

排除标准

  • Noncompliance with dialysis regimen.
  • Hematocrit less than 28%.
  • Active infection

结局指标

主要结局

Effect of Increased Dialysate Flow Rate on KoA for Phosphorus Between 4 Dialyzers With Different Membrane Packing Densities.

时间窗: During the third treatment with each dialyzer (one time during each trial period week)

Effect of Increased Dialysate Flow Rate on Whole Blood Phosphorus Clearance Between 4 Dialyzers With Different Membrane Packing Densities.

时间窗: During the third treatment with each dialyzer (one time during each trial period week)

Effect of Increased Dialysate Flow Rate on Whole Blood B2-microglobulin Clearance Between 4 Dialyzers With Different Membrane Packing Densities.

时间窗: During the third treatment with each dialyzer (one time during each trial period week)

Effect of Increased Dialysate Flow Rate on KoA for Urea Between 4 Dialyzers With Different Membrane Packing Densities.

时间窗: During the third treatment with each dialyzer (one time during each trial period week)

KoA is a constant that describes the efficiency of a dialyzer in removing urea. KoA is determined by surface area of the membrane, the thickness of the membrane and pore size.

Effect of Increased Dialysate Flow Rate on Whole Blood Urea Clearance Between 4 Dialyzers With Different Membrane Packing Densities.

时间窗: During the third treatment with each dialyzer (one time during each trial period week)

Effect of Increased Dialysate Flow Rate on KoA for b2-microglobulin Between 4 Dialyzers With Different Membrane Packing Densities.

时间窗: During the third treatment with each dialyzer (one time during each trial period week)