Investigation of the Utility of the Audible Alert in Current Generation Medtronic ICDs

Englewood Hospital and Medical Center1 个研究点 分布在 1 个国家目标入组 100 人开始时间: 2012年6月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
100
试验地点
1
主要终点
Audibility

研究概览

简要总结

The purpose of this study is to determine whether the alert tones emitted by the latest generation of Medtronic implantable cardioverter-defibrillators (ICDs) are audible to patients.

ICD patients can have difficulties hearing the beeps, which are designed to notify patients of possible lead or generator malfunction. In fact, the only published study of this issue, conducted at Englewood Hospital and Medical Center, revealed that only 50% of patients with older Medtronic devices were able to hear alerts. Since then, modifications to the alert have been made by Medtronic. This study will establish the "real-world" value of these changes and also provide useful in vivo information regarding the validity of the in vitro techniques which have been used thus far to design the Patient Alert feature.

详细描述

There will be a single study visit. In a quiet examination room, the Patient Alert tones will be demonstrated to the subjects. Subsequently, this will be repeated with a wall air conditioner turned on. The subjects will then be asked to complete a three-part questionnaire regarding their perceptions of the Patient Alert tones.

Data analysis will include overall ability to hear tones, as well as subgroup analysis according to gender, age, and BMI.

Simultaneously, in order to calculate the loudness of the alert tones, digital recordings will be obtained using two microphones located near the subject. These digital recordings will be analyzed, and a loudness value will be calculated. The loudness value is a calculated index, devised by Medtronic, which incorporates frequency and intensity. Data analysis will include a correlation between loudness value and patient characteristics, and between loudness value and ability of subjects to perceive the tones.

研究设计

研究类型
Observational
观察模型
Case Only
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Previous insertion of a Medtronic Generation 2 ICD (Consulta, Secura, Concerto II, Virtuoso II, Maximo II devices) not fewer than 7 days prior to study entry
  • Able to provide informed consent

排除标准

  • Known deafness
  • Recent (fewer than 6 weeks) ICD pocket surgery
  • Non-Generation 2 Medtronic ICD
  • Pocket hematoma
  • Pocket infection

结局指标

主要结局

Audibility

时间窗: participants will be tested during a single, routine clinic visit - the average duration for each patient will be 30 minutes

how frequently patients are able to detect the alert tones

次要结局

  • Loudness value(1 month)

研究者

申办方类型
Other
责任方
Sponsor