跳至主要内容
临床试验/NCT04458155
NCT04458155已完成不适用

Sample Comparison With Siemens® Point-of-care Device for Cardiac Troponin I Assay by Using a Heparin Coated Transfer Device vs. Non-heparin Coated Transfer Device at the Emergency Department (Validation Study 2.0)

VieCuri Medical Centre1 个研究点 分布在 1 个国家目标入组 70 人开始时间: 2020年6月18日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
70
试验地点
1
主要终点
Sample comparison

研究概览

简要总结

This study is a prospective, diagnostic, cohort study within the standard care of acute coronary syndrome (ACS) patients. It compares the analytical performance of Siemens® point-of-care high sensitive troponin I testing in venous, plasma and capillary sample types. The investigators hypothesize that there is a good correlation between the Siemens® POC HS cTnI assay results for the three sample types and that the bias between different POC sample types reduces from ~10% to ≤ 5% when using heparinized transfer device for the capillary sample.

详细描述

In September 2019, the validation study 1.0 started, in which the cTnI result of the Siemens POC device on three sample types are compared. Interim analysis of the sample comparison was performed by regression analysis using Passing and Bablock, and calculating the Pearson correlation coefficient.

The Li-hep Plasma vs Li-hep venous blood show a very good correlation of 1.00-1.03 with an R of >0.99, so the results between these sample types can be used interchangeably. For the capillary sample vs the Li-hep sample (both blood and plasma) the slope is 8-12% higher. With a 8-12% higher response, the capillary test results may not be interchangeably used with the other 2 sample types.

It is remarkable that capillary samples give a higher response, since it was anticipated that the result may be slightly lower due to the possible dilution by interstitial fluid. The investigators hypothesize that the presence of the Li-heparin anti-coagulant in the venous draw lead to a slight reduction of the apparent cTnI concentration. By using a heparin coated transfer device for the capillary samples instead of an uncoated transfer device, this hypothesis will be tested.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Retrospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Age 18 years or older.
  • Referred to cardiac ED with chest pain suspected of ACS; inclusion at arrival (T=0) or one hour after arrival (T=1).
  • Subacute STEMI or NSTEMI patients admitted at the CCU who have an indication for coronary angiography but do not need rescue/emergency PCI.
  • STEMI patients who already underwent rescue/emergency PCI; inclusion post PCI.

排除标准

  • Out of hospital cardiac arrest.
  • Patients with sudden onset tachycardia and a frequency of 110 bpm or higher (supraventricular or ventricular).
  • Patients who are hemodynamically unstable or in which an acute non-coronary diagnosis is suspected, e.g. pulmonary embolism, thoracic aortic dissection etc.
  • Patients recently already admitted for the same set of symptoms at a previous healthcare institution before being transferred to the participating clinical site.
  • Patients not willing or not able to provide informed consent due to their medical condition as judged by the physician.

结局指标

主要结局

Sample comparison

时间窗: 30 days

The primary objective is to compare the analytical performance of Siemens® point-of-care high sensitive troponin I testing in different sample types by using the coefficient of variation. This comparison will comprise Siemens® POC capillary vs. Siemens® POC venipuncture and vs. Siemens® POC plasma (Sample comparison). The primary objective will be achieved by taking: * Two capillary samples \> one sample with a heparin coated transfer device (drawn from a finger of the right hand) and the other one with a non-heparin coated transfer device. (drawn from a finger of the left hand) * One extra venous blood sample \> for POC venous and POC plasma analysis. Trained clinical staff will collect the different samples. All samples will be collected once, at one time point with a maximum of 10 minutes in between.

次要结局

  • Bland-Altman method(30 days)
  • Overview baseline characteristics.(30 days)
  • Linear regression and Pearson's correlation.(30 days)
  • Modified HEART score(30 days)
  • Modified HEART score comparison(30 days)
  • Major adverse cardiac event (MACE)(30 days)

研究者

发起方
VieCuri Medical Centre
申办方类型
Other
责任方
Principal Investigator
主要研究者

Braim Rahel

Dr. B.M. Rahel, Principal investigator, Cardiologist

VieCuri Medical Centre

研究点 (1)

Loading locations...