Moni-Patc Temperature Monitoring Patch System Clinical Accuracy Validation
试验速览
- 阶段
- 不适用
- 状态
- 尚未招募
- 发起方
- 入组人数
- 30
- 主要终点
- Agreement Between Moni-Patch™ Temperature Monitoring Patch System and Esophageal (Teso) and Bladder (Tbld) Core Temperature Measurements
研究概览
简要总结
Moni-Patc Temperature Monitoring Patch System, a non-invasive, continuous temperature monitoring device designed to estimate core body temperature by applying a Sensor to the neck.This clinical accuracy validation study aims to verify the accuracy and repeatability of the Moni-Patch compared with an FDA-cleared continuous esophageal and bladder temperature monitoring device used as the clinical reference.Main ObjectivesTo verify that the Moni-Patch and the reference device(esophageal and bladder) demonstrate acceptable agreement, as defined by Bland-Altman Analyses (bias within ±0.4°C and limits of agreement between -1.0°C and 1.0°C).To verify the consistency and repeatability of temperature measurements obtained from the Moni-Patch during continuous monitoring.
详细描述
This single-center, open-label accuracy validation compares continuous core temperature measured by the Moni-Patch applied on the neck with esophageal (T_eso) and bladder (T_bld) temperatures during elective surgeries under general anesthesia. Approximately 30 adult participants will be monitored for ≥60 minutes intraoperatively. Agreement will be assessed using Bland-Altman methods (bias and 95% limits of agreement). The device is non-invasive and Bluetooth-enabled; readings are captured via the receiver and routed to the OR multi-parameter monitor and EMR. Safety monitoring focuses on device-related skin reactions and device deficiencies; no changes to anesthetic care are mandated by the protocol.
研究设计
- 研究类型
- Observational
- 观察模型
- Case Only
- 时间视角
- Prospective
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Age ≥18 years
- •Subject is undergoing an elective surgical procedure expected to require at least 60min of general anesthesia and during which routine esophageal (Teso) and bladder (Tbld) temperature monitoring is expected.
- •Subjects are expected to remain hospitalized at least overnight after surgery.
- •Subject specifically agrees to temperature Sensor placement on the neck prior to surgery.
- •Must be able to attach the Sensor away from the operative field (neck) with uninterrupted wireless connection between the Sensor and the Receiver (as close as possible).
- •Use environment must meet both of the following parameters:
- •Temperature 15-40⁰ C/59-104⁰ F
- •Humidity 30-75% RH (non-condensing)
- •For non-English speaking subjects, a translator must be present at the time of consent.
排除标准
- •Subjects unable to provide their own consent. Consent from family or LOR is not acceptable.
- •Subjects undergoing surgical procedures who have conditions of the esophagus or bladder that contraindicate placement of these monitoring probes/sensor.
- •Subjects whose surgery does not involve the monitoring of esophageal or bladder temperature.
- •Subjects undergoing surgery that prohibits placement of an esophageal temperature monitoring probe/sensor (e.g. neck surgery).
- •Subjects with medical conditions such as inflammation, scratches, redness, rash at the Sensor placement site.
- •Subjects for whom the sensor cannot be securely adhered to the skin without floating or detaching, or accurately positioned over the carotid artery.
- •Not for use in MRI or CT facilities.
- •Use environment does not meet one or more parameters.
结局指标
主要结局
Agreement Between Moni-Patch™ Temperature Monitoring Patch System and Esophageal (Teso) and Bladder (Tbld) Core Temperature Measurements
时间窗: Continuous intra-operative monitoring for ≥60 minutes, beginning after ≥30 minutes of Moni-Patch sensor acclimation prior to operating room entry.
This outcome assesses agreement between the Moni-Patch™ Temperature Monitoring Patch System and clinical reference core temperature measurements obtained using FDA-cleared esophageal temperature probes (Teso) and FDA-cleared bladder temperature probes (Tbld) during general anesthesia. Paired temperature values are collected simultaneously from: Moni-Patch™ Sensor (neck placement) Esophageal core temperature probe (Teso) Bladder core temperature probe (Tbld) Agreement is evaluated using Bland-Altman analysis, including: Mean temperature difference (bias; °C) 95% limits of agreement (LOA; °C) Name of Measurement: Core temperature difference (Moni-Patch - Teso; Moni-Patch - Tbld) Measurement Tools: Moni-Patch™ Temperature Monitoring Patch System FDA-cleared esophageal core temperature monitor (Teso) FDA-cleared bladder core temperature monitor (Tbld) Unit of Measure: Degrees Celsius (°C)