MedPath

Factors in Accuracy Studies Influencing Measured CGM Performance.

Not Applicable
Completed
Conditions
Diabetes Mellitus, Type 1
Interventions
Device: Venous reference measurement (YSI)
Device: Arterialized-venous reference (YSI)
Device: Capillary reference (SMBG)
Registration Number
NCT02843503
Lead Sponsor
Academisch Medisch Centrum - Universiteit van Amsterdam (AMC-UvA)
Brief Summary

Primary objective is to assess the impact of different reference glucose measurement methods on the measured performance of a CGM system, aiming at further standardization for the assessment of CGM systems. Secondary objective is to describe the relationship between glucose concentration of venous, capillary and arterialized-venous measurement samples.

Study design: Open-label trial. Study population: Nineteen patients with type 1 diabetes aged 18 years and older will be included.

Intervention: The study will use the Dexcom G5 Mobile CGM. The participant will visit the clinical research center (CRC) twice during a 6 day period. These visits will have a duration of 8 hours. Blood will be drawn for the determination of glucose concentrations using venous, arterialized-venous and capillary sampling techniques. The patient will receive his usual breakfast and an increased insulin bolus will be administered to correct the breakfast glucose excursion with the aim of inducing a period of minor hypoglycemia. Blood sampling will continue until the end of the admission. The sensor will be worn until the end of the second 8-hour CRC session, which will be planned on day 6 or earlier depending on patient preference. The sensor will be removed and their sensor and blood glucose meter data downloaded from the devices. In the case of sensor failure before the first CRC session, patients will be instructed to insert a new sensor per the manufacturer's instruction and to notify study staff of the event. In case of failure of the sensor after the first CRC session but before the second CRC session, patients will return to the CRC for sensor removal and for data download.

Detailed Description

Not available

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
21
Inclusion Criteria
  • diagnosed with type 1 DM at least 6 months according to the WHO definition
  • Body Mass Index (BMI) <35 kg/m²
  • willing and able to wear a CGM device for the duration of the study and undergo all study procedures
  • HbA1c <10%
  • Signed informed consent form prior to study entry
Exclusion Criteria
  • patient is pregnant, or breast feeding during the period of the study.
  • patient is using a medication that significantly impacts glucose metabolism (oral steroids) except if stable for at least the last three months and expected to remain stable for the study duration.
  • patient requires regular use of acetaminophen (paracetamol) while participating in the study due to known negative impact on CGM performance.
  • patient is actively enrolled in another clinical trial or took part in a study within 30 days.
  • known adrenal gland problem, panhypopitutarism, gastroparesis, migraine, epilepsy or ischemic heart disease or other cardiovascular event in the year previous to study participation .
  • inability of the patient to comply with all study procedures.
  • inability of the patient to understand the patient information.
  • patient donated blood in the last 3 months.
  • has severe medical or psychological condition(s) or chronic conditions/infections that in the opinion of the investigator would compromise the patient's safety or successful participation in the study.
  • is assessed by the investigators to have difficult intravenous (IV) access .

Study & Design

Study Type
INTERVENTIONAL
Study Design
PARALLEL
Arm && Interventions
GroupInterventionDescription
Venous reference samples YSIVenous reference measurement (YSI)CGM monitoring performance per venous reference measurement (YSI)
Arterialized-venous reference YSIArterialized-venous reference (YSI)CGM monitoring performance per arterialized venous reference measurement (YSI)
Venous reference samples YSICapillary reference (SMBG)CGM monitoring performance per venous reference measurement (YSI)
Primary Outcome Measures
NameTimeMethod
Mean Absolute Relative Difference (MARD)up to day 6 of use

MARD will be assessed per average of the 8 hour arterialization and regular venous study visit.

Secondary Outcome Measures
NameTimeMethod
Accuracy of sensors (MARD) per hypoglycemic rangeup to day 6 of use

Accuracy of the sensor in the hypoglycaemic (defined as a blood glucose value ≤70 mg/dL) area as an average of the first six days of use.

Accuracy (MARD) of sensors per hyperglycemic rangeup to day 6 of use

Accuracy of the sensor in the hyperglycaemic (defined as a blood glucose value \>180 mg/dL) area as an average of the first six days of use.

Glucose content (mg/dL) of venous reference samples.Up day 6 of study during both intervention visits

Glucose content of venous reference samples will be measured per YSI.

Glucose content (mg/dL) of venous-arterialized reference samples.up day 6 of study during both intervention visits

Glucose content of venous-arterialized reference samples will be measured per YSI.

Glucose content (mg/dL) of capillary reference samples.up day 6 of study during both intervention visits

Glucose content of capillary reference samples will be measured per SMBG.

Trial Locations

Locations (1)

Academic Medical Center

🇳🇱

Amsterdam, Noord-Holland, Netherlands

© Copyright 2025. All Rights Reserved by MedPath