Accuracy of Linear Measurements of Ultra-low Dose Cone Beam Computed Tomography and Digital Panoramic Radiography Performed on Mandibular Anaesthetic Landmarks Versus Real Mandibular Measurements

Cairo University2 个研究点 分布在 1 个国家目标入组 10 人开始时间: 2018年7月10日最近更新:
适应症
干预措施

试验速览

阶段
不适用
入组人数
10
试验地点
2
主要终点
assessment of diagnostic accuracy of linear measurements

研究概览

简要总结

measurements performed directly on dry mandibles will be compared with measurements obtained from CBCT( ultra-low dose) scans and Digital panoramic radiography.

详细描述

Mandibular anaesthetic landmarks will be identified on each dry mandible and will be marked using gutta percha.

The ten human dry mandibles will be submitted for:

  1. Digital Panoramic Radiography. Panoramic radiographs will be obtained using the SOREDEX CRANEX™ 3Dx unit. The exposure parameters will be 8Ma, 63kVp and 16.4 sec exposure time.
  2. Cone Beam Computed Tomography (CBCT). SOREDEX CRANEX™ 3Dx unit will be used in this study. Images will be acquired at a single rotation. SOREDEX® MiniDose parameters will be used. The exposure parameters will be 3.2 Ma, 90kVp and 2.3 sec exposure time to take the advantage of 3D data in dose sensitive cases like children, or reduce the radiation dose for the patient.

The real linear measurements will be measured in millimeter on dry mandibles using digital caliper and will be compared with measurements obtained from both techniques.

Images will be evaluated by two experienced radiologists of ten years of experience. Each one will evaluate the images separately twice with a period of two weeks in-between the two reading sessions. The researcher will measure out the real linear measurements on the dry mandibles using digital caliper.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Diagnostic
盲法
None

入排标准

年龄范围
10 Years 至 70 Years(Child, Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • age 10 to 70 years
  • no gender nor race predilection
  • Either dentulous or edentulous are acceptable.

排除标准

  • fractures
  • pathology

研究组 & 干预措施

column

Other

ultra-low dose CBCT versus digital panoramic radsiography

干预措施: CBCT (Other)

结局指标

主要结局

assessment of diagnostic accuracy of linear measurements

时间窗: within 2 years

measurements made on dry mandibles will be compared with measurements obtained from CBCT scans \& digital panoramic radiography