Accuracy of non-invasive cardiac output measurement in awake patients
- Conditions
- Postoperative patients transferred to the intensive care unit under intubation
- Registration Number
- JPRN-UMIN000044712
- Lead Sponsor
- agoya University
- Brief Summary
The paired esCCO and TDCO measurements were compared. The respective bias (standard deviation) values were 0.13(0.60) L/min before extubation, and -0.48(0.78) L/min after extubation. The percentage errors were 25.1% before extubation and 29.6% after extubation. Thus, the accuracy of the noninvasive esCCO measurement is clinically acceptable as an alternative to TDCO measurement under different respiratory conditions of mechanical ventilation and spontaneous respiration.
- Detailed Description
Not available
Recruitment & Eligibility
- Status
- Complete: follow-up complete
- Sex
- All
- Target Recruitment
- 40
Not provided
1. Emergency surgery
Study & Design
- Study Type
- Observational
- Study Design
- Not specified
- Primary Outcome Measures
Name Time Method Accuracy of cardiac output measurements during sedation and awake in the same patient
- Secondary Outcome Measures