Accuracy of non-invasive cardiac output measurement in awake patients

Not Applicable
Conditions
Postoperative patients transferred to the intensive care unit under intubation
Registration Number
JPRN-UMIN000044712
Lead Sponsor
agoya University
Brief Summary

The paired esCCO and TDCO measurements were compared. The respective bias (standard deviation) values were 0.13(0.60) L/min before extubation, and -0.48(0.78) L/min after extubation. The percentage errors were 25.1% before extubation and 29.6% after extubation. Thus, the accuracy of the noninvasive esCCO measurement is clinically acceptable as an alternative to TDCO measurement under different respiratory conditions of mechanical ventilation and spontaneous respiration.

Detailed Description

Not available

Recruitment & Eligibility

Status
Complete: follow-up complete
Sex
All
Target Recruitment
40
Inclusion Criteria

Not provided

Exclusion Criteria

1. Emergency surgery

Study & Design

Study Type
Observational
Study Design
Not specified
Primary Outcome Measures
NameTimeMethod
Accuracy of cardiac output measurements during sedation and awake in the same patient
Secondary Outcome Measures