Evaluation of Temperature Detecting (Infrared) Cameras and Non-contact (Infrared) Thermometers (NCITs) for Measuring Body Temperatures
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 1,125
- 主要终点
- Evaluating accuracy of infrared cameras and thermometers for skin temperature measurement
研究概览
简要总结
Fever is one of the key symptoms exhibited by contagious patients infected with the Ebola and other viruses. Thermographs and non-contact infrared thermometers (NCITs) are commonly used as early detection tools for screening and isolating sick individuals in healthcare settings and transit centers such as airports. The objective of this clinical study is to evaluate the temperature measurement accuracy of thermographs and NCITs. The study will be accomplished by establishing a protocol based on best practices for screening published in the IEC 80601-2-59:2008 Medical electrical equipment -- Part 2-59: Particular requirements for basic safety and essential performance of screening thermographs for human febrile temperature screening standard, performing initial measurements in patients to optimize clinical procedures, then performing quantitative clinical measurements in febrile and afebrile human subjects to compare thermographs - used in accordance with recognized best practices for measurement - with NCITs for fever screening. Since thermographs and NCITs represent the only currently viable mass screening approaches for infectious disease pandemics, like the recent Ebola virus disease (EVD) outbreak in West Africa, the study will evaluate these thermal modalities as medical countermeasures for fever-related pandemics and therefore improve response to these diseases.
研究设计
- 研究类型
- Observational
- 观察模型
- Other
- 时间视角
- Cross Sectional
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •Any human subject above the age of 18 that can sit for about 40 minutes with permitted breaks as needed and follow the study instructions can be included.
排除标准
- •Human subjects should be free of recent (1 week prior) or current disease affecting the ICR and forehead skin. They should not use topical cosmetics, creams or medications on the face on the day of participation.
结局指标
主要结局
Evaluating accuracy of infrared cameras and thermometers for skin temperature measurement
时间窗: Through study completion, about 2 year
The temperatures of each participant were measured with two infrared (IR) cameras and multiple IR thermometers at different locations on facial skin. The measured temperatures were compared with oral temperature to evaluate the accuracy of IR cameras and IR thermometers.
次要结局
未报告次要终点
研究者
Quanzeng Wang
Biomedical Engineer
Food and Drug Administration (FDA)